文献速递心血管技术与安全CT

光子计数探测器CT对心血管植入式电子装置导线的成像

Imaging of Cardiovascular Implantable Electronic Device Leads With Photon-Counting Detector Computed Tomography.
PubMed 文献 · Invest Radiol · 2025-09-01

本研究在非搏动心脏体模中评估5种心血管植入式电子装置导线,采用双源光子计数探测器CT进行心电图门控光谱标准分辨率(准直144×0.4 mm)和非光谱超高分辨率(准直120×0.2 mm)螺旋采集,比较不同重建参数及迭代金属伪影去除算法对导线显示的影响,并形成优化扫描与重建方案。

要点

  • 超高分辨率与标准分辨率扫描在相似重建参数下金属伪影无显著差异。
  • 迭代金属伪影去除算法中「神经线圈」设置表现最佳。
  • 0.2 mm/Bv56重建利于显示螺旋等细微结构,0.6 mm/Bv40利于观察导线整体及邻近解剖结构。

核心结论

  • 超高分辨率和标准分辨率光子计数探测器CT联合迭代金属伪影去除算法可实现心血管植入式电子装置导线的优质成像,能显示细小细节且不损害邻近结构可见性;推荐采用超高分辨率扫描加两种重建(0.2 mm/Bv56和0.6 mm/Bv40,使用迭代金属伪影去除算法)的专用采集与重建方案。

研究设计与样本

  • 体模研究,在非搏动心脏体模中评估5种心血管植入式电子装置导线,使用双源光子计数探测器CT行心电图门控光谱标准分辨率(准直144×0.4 mm)和非光谱超高分辨率(准直120×0.2 mm)螺旋采集,每条导线每种采集模式各扫描一次。

关键影像表现/征象

  • 金属伪影随卷积核变锐和虚拟单能重建keV水平升高而增加。
  • 金属伪影不依赖于层厚。
  • 使用相似重建参数时,超高分辨率与标准分辨率扫描之间金属伪影无显著差异。
  • 最薄层厚和最锐卷积核(0.2 mm、Bv56)可更好勾画螺旋形状等细微结构。
  • 较厚层厚和较软卷积核(0.6 mm、Bv40)更适于观察导线整体外观和邻近解剖结构。

诊断标准与关键数值

  • 图像质量主观评估采用5分制。
  • 重建参数包括层厚0.2 mm、0.4 mm、0.6 mm,卷积核Bv40、Bv44、Bv48、Bv56,虚拟单能水平60-140 keV(步长10 keV),以及不同迭代金属伪影去除算法设置。
  • 迭代金属伪影去除算法在所有卷积核、层厚和keV水平下均有效减少伪影,其中「神经线圈」设置表现最佳。

临床应用与阅片要点

  • 作者认为,包含超高分辨率扫描加两种重建(0.2 mm/Bv56和0.6 mm/Bv40,使用迭代金属伪影去除算法)的专用采集与重建方案似乎是光子计数探测器CT心血管植入式电子装置导线成像的最佳方案。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Computed tomography (CT) imaging of cardiovascular implantable electronic device (CIED) leads is currently hampered by large metal artifacts. Recently, photon-counting detector CT (PCD-CT) has been clinically introduced, offering high-resolution imaging with thin slice thicknesses and improved contrast-to-noise ratios. Suspected complications of CIED such as perforation, infection and venous obstruction could potentially be imaged with PCD-CT if metal artifacts were effectively reduced through adapted scan protocols and iterative metal artifact reduction algorithms (IMAR). The study evaluated the impact of various scan and reconstruction parameters, including different IMAR settings, on CIED lead visualization with PCD-CT in order to develop an optimized scan and reconstruction protocol for imaging leads.

MATERIALS AND METHODS: Five different CIED leads were evaluated in a nonbeating heart phantom using a dual-source PCD-CT with electrocardiography-gated spectral standard resolution (collimation 144 × 0.4 mm) and nonspectral ultra-high resolution (UHR) mode (collimation 120 × 0.2 mm) spiral data acquisition. One scan was performed for each lead and each acquisition mode. Images were reconstructed with different slice thicknesses (0.2 mm, 0.4 mm, 0.6 mm), convolution kernels (Bv40, Bv44, Bv48, and Bv56), virtual monoenergetic energy levels (60-140 keV in steps of 10 keV), without and with different IMAR settings. The extent of metal artifacts was objectively evaluated using 4 different parameters. Additionally, 3 observers subjectively assessed image quality using a 5-point scale.

RESULTS: Metal artifacts increased with sharper kernels and higher keV levels in virtual monoenergetic reconstructions. The artifacts were not dependent on slice thickness. No significant differences in metal artifacts were observed between UHR and standard-resolution scans when using similar reconstruction parameters. IMAR effectively reduced artifacts across all kernels, slice thicknesses, and keV levels, with the "neuro coils" setting showing the best performance. Subjective analysis of image quality revealed that thinnest slices and sharpest kernels (0.2 mm, Bv56) allowed for better delineation of fine structures, such as the shape of helices, while reconstructions with thicker slices and softer kernels (0.6 mm, Bv40) were preferred for visualizing general lead appearance and adjacent anatomical structures.

CONCLUSIONS: Ultra-high and standard resolution PCD-CT with IMAR enables good-quality imaging of CIED leads, showing even small details without compromising the visibility of nearby structures. A dedicated acquisition and reconstruction protocol comprising an UHR scan with 2 reconstructions (0.2 mm/Bv56 and 0.6 mm/Bv40, using IMAR) appears optimal for PCD-CT imaging of CIED leads.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Schalla S, Wildberger JE, Ramaekers MJFG, et al. Imaging of Cardiovascular Implantable Electronic Device Leads With Photon-Counting Detector Computed Tomography. Invest Radiol. 2025. PMID: 40138655. doi:10.1097/RLI.0000000000001177

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。