文献速递骨肌MR

通过后处理优化肩部合成MRI并与常规MRI比较

Optimization of shoulder synthetic MRI through post-processing and comparison with conventional MRI.
PubMed 文献 · Eur J Radiol · 2025-03-22

研究纳入20名健康志愿者行肩部MRI,评估合成MRI与常规MRI的效用,并通过后处理优化形态学图像参数。结果显示短回波时间和长重复时间合成的图像肌腱-肌肉对比度高;脂肪抑制T2加权图像上常规图像质量最高(p < 0.001),优化图像肩袖可见性与常规图像相当(p = 0.031-1.0);MDME序列T2值在肌肉处与MESE一致(p = 0.81),但在软骨处约长8.3 ms(p < 0.001)。

要点

  • 短回波时间与长重复时间后处理合成的图像可提高肌腱-肌肉对比度。
  • 脂肪抑制T2加权图像上常规图像质量最高(p < 0.001),优化图像肩袖可见性与常规相当(p = 0.031-1.0)。
  • MDME与MESE序列T2值在肌肉处一致(p = 0.81),软骨处MDME约长8.3 ms(p < 0.001)。

核心结论

  • 合成MRI使用适当后处理参数可提供可接受的图像质量,5分钟内同时获取多种形态学图像和定量图对肩部检查有前景。

研究设计与样本

  • 20名健康志愿者行肩部MRI,比较常规图像、预设参数合成图像和优化图像三组,并由两名放射科医生分析。

关键影像表现/征象

  • 短回波时间和长重复时间合成的图像显示高肌腱-肌肉对比度。
  • 脂肪抑制T2加权图像上常规图像质量最高(p < 0.001),优化图像肩袖可见性与常规图像相当(p = 0.031-1.0)。
  • 质子密度加权图像和T1加权图像中各图像组间无显著差异(p > 0.05)。

诊断标准与关键数值

  • 脂肪抑制T2加权图像上常规图像质量最高(p < 0.001),优化图像肩袖可见性与常规图像相当(p = 0.031-1.0)。
  • MDME序列T2值在肌肉处与MESE一致(p = 0.81),在软骨处约长8.3 ms(p < 0.001)。

临床应用与阅片要点

  • 合成MRI可通过短回波时间和长重复时间的后处理参数优化肌腱-肌肉对比度。
  • 脂肪抑制T2加权图像上优化图像肩袖可见性与常规图像相当,但常规图像整体质量最高(p < 0.001)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To evaluate the utility of synthetic MRI of the shoulder compared with conventional MRI and to optimize the parameters of morphological images through post-processing. Additionally, we aimed to assess the agreement of T2 values between multi-echo spin-echo (MESE) and multi-dynamic multi-echo (MDME) sequences.

METHODS: Twenty healthy volunteers underwent shoulder MRI. The optimal post-processing parameters for the tendon-muscle contrast were examined using synthetic MRI, and two radiologists analyzed three image sets: conventional images, synthetic images using preset parameters, and optimized images. Qualitative analysis included assessment of the visibility of anatomical structures, overall image quality, and magic angle effect, whereas the quantitative analysis included measurement of the relative signal intensity and relative contrast. The T2 values of the articular cartilage and supraspinatus muscle were measured for each sequence.

RESULTS: Images synthesized with short echo times and long repetition times showed high tendon-muscle contrast. For fat-suppressed T2-weighted images, conventional images showed the highest image quality (p < 0.001), whereas the optimized images achieved comparable visibility of the rotator cuff (p = 0.031-1.0). No significant differences were observed among image sets in proton density-weighted images and T1-weighted images (p > 0.05). The T2 values of the MDME sequence were consistent with those of the MESE sequence at the muscle (p = 0.81), but were approximately 8.3 ms longer at the cartilage (p < 0.001).

CONCLUSIONS: Synthetic MRI provided acceptable image quality using appropriate post-processing parameters. The simultaneous acquisition of multiple morphological images and quantitative maps within five minutes holds promise for shoulder examination.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Nishizawa N, Yabuuchi H, Nishikawa K, et al. Optimization of shoulder synthetic MRI through post-processing and comparison with conventional MRI. Eur J Radiol. 2025. PMID: 40157115. doi:10.1016/j.ejrad.2025.112069

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。