标准化胰腺MRI-T1测量方法:手动测量与带自动质量控制的半自动流程比较

Standardized pancreatic MRI-T1 measurement methods: comparison between manual measurement and a semi-automated pipeline with automatic quality control.
PubMed 文献 · Br J Radiol · 2025-01-01

多中心研究评估手动ROI法测量胰腺srT1的可重复性,并验证带自动质控模块的半自动化测量方法。半自动法与手动法偏倚低(3.8 ms或0.5%),一致性界限[-36.6, 44.1] ms;加入质控后内部验证集一致性改善,外部验证集不变,均在操作者间界限内。半自动法改善扫描-再扫描一致性并减少人工分析时间。

要点

  • 半自动法测量胰腺srT1与手动ROI法偏倚低(3.8 ms或0.5%),一致性界限为[-36.6, 44.1] ms。
  • 未加自动质控时两法一致性良好(内部验证集3.2 [-47.1, 53.5] ms,外部验证集-0.5 [-35.2, 34.2] ms)。
  • 加自动质控后内部验证集一致性改善(-0.04 [-33.4, 33.3] ms),外部验证集不变(-1.9 [-37.6, 33.7] ms)。

核心结论

  • 半自动测量标准化胰腺T1的方法有潜力减少分析时间,同时保持准确性并改善扫描-再扫描一致性。

研究设计与样本

  • 多中心研究;扫描-再扫描队列46例评估手动分析可重复性;纵向多中心研究466例共708次扫描分为训练、内部验证和外部验证队列;提出基于机器学习的半自动srT1测量方法并与手动分析比较。

诊断标准与关键数值

  • 手动ROI法与半自动法操作者间偏倚3.8 ms或0.5%,一致性界限[-36.6, 44.1] ms;未加质控时内部验证集3.2 [-47.1, 53.5] ms、外部验证集-0.5 [-35.2, 34.2] ms;加质控后内部验证集-0.04 [-33.4, 33.3] ms、外部验证集-1.9 [-37.6, 33.7] ms。

临床应用与阅片要点

  • 半自动法测量srT1与手动ROI分析一致性良好,与操作者间表现相当;加入自动质控可改善内部验证集手动与自动测量间的一致性,外部验证集一致性不变。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Scanner-referenced T1 (srT1) is a method for measuring pancreas T1 relaxation time. The purpose of this multi-centre study is 2-fold: (1) to evaluate the repeatability of manual ROI-based analysis of srT1, (2) to validate a semi-automated measurement method with an automatic quality control (QC) module to identify likely discrepancies between automated and manual measurements.

METHODS: Pancreatic MRI scans from a scan-rescan cohort (46 subjects) were used to evaluate the repeatability of manual analysis. Seven hundred and eight scans from a longitudinal multi-centre study of 466 subjects were divided into training, internal validation (IV), and external validation (EV) cohorts. A semi-automated method for measuring srT1 using machine learning is proposed and compared against manual analysis on the validation cohorts with and without automated QC.

RESULTS: Inter-operator agreement between manual ROI-based method and semi-automated method had low bias (3.8 ms or 0.5%) and limits of agreement [-36.6, 44.1] ms. There was good agreement between the 2 methods without automated QC (IV: 3.2 [-47.1, 53.5] ms, EV: -0.5 [-35.2, 34.2] ms). After QC, agreement on the IV set improved, was unchanged in the EV set, and the agreement in both was within inter-operator bounds (IV: -0.04 [-33.4, 33.3] ms, EV: -1.9 [-37.6, 33.7] ms). The semi-automated method improved scan-rescan agreement versus manual analysis (manual: 8.2 [-49.7, 66] ms, automated: 6.7 [-46.7, 60.1] ms).

CONCLUSIONS: The semi-automated method for characterization of standardized pancreatic T1 using MRI has the potential to decrease analysis time while maintaining accuracy and improving scan-rescan agreement.

ADVANCES IN KNOWLEDGE: We provide intra-operator, inter-operator, and scan-rescan agreement values for manual measurement of srT1, a standardized biomarker for measuring pancreas fibro-inflammation. Applying a semi-automated measurement method improves scan-rescan agreement and agrees well with manual measurements, while reducing human effort. Adding automated QC can improve agreement between manual and automated measurements.

SUMMARY STATEMENT: We describe a method for semi-automated, standardized measurement of pancreatic T1 (srT1), which includes automated quality control. Measurements show good agreement with manual ROI-based analysis, with comparable consistency to inter-operator performance.

来源:Br J Radiol,查看原文。版权归原作者所有。