常规与低剂量CT检查中肝脏病灶检测的大小与对比度阈值:100例患者2300个合成病灶的阅片者研究

Size and Contrast Thresholds for Liver Lesion Detection in Regular and Low-dose CT Examinations: A Reader Study of 2300 Synthetic Lesions Across 100 Patients.
PubMed 文献 · Acad Radiol · 2025-03-22

该回顾性阅片者研究在100例腹部CT数据集(A、B两家厂商各50例,常规剂量与低剂量肾结石方案各半)中数字插入23种直径与对比度组合的圆柱形肝脏病灶,由7名放射科医师用四选项强制选择法评估可检测性。常规剂量下15、10、5mm病灶的检测阈值分别约为-30、-35、-70 HU(A厂商)和-25、-35、-65 HU(B厂商);低剂量下15mm和10mm病灶阈值增高至-40~-50 HU,5mm病灶未达检测阈值。

要点

  • 常规剂量肿瘤CT中,肝脏病灶可靠检测所需对比度随直径减小而增大:15mm约-30 HU、10mm约-35 HU、5mm约-70 HU。
  • 低剂量肾结石方案下,15mm和10mm病灶分别需约-40 HU和-50 HU的对比度,5mm病灶未达检测阈值。
  • 低剂量扫描图像噪声在B厂商显著升高(25.9±4.9 HU vs 30.9±3.1 HU,P<.001),A厂商无显著差异(P=.06)。

核心结论

  • 肿瘤学CT检查中,直径15、10、5mm的肝脏病灶可靠检测分别需要-30、-35、-70 HU的对比度;低剂量检查中15mm和10mm病灶需-40和-50 HU的更大对比度,而阅片者不能可靠检测5mm病灶。

研究设计与样本

  • 回顾性收集100例腹部CT数据集,A厂商和B厂商各50例,每家厂商半数数据集为常规剂量肿瘤学检查、半数为低剂量肾结石方案采集,数字插入23种直径与对比度组合的圆柱形肝脏病灶,7名放射科医师行四选项强制选择阅片实验并测量肝内图像噪声。

诊断标准与关键数值

  • 常规剂量检测阈值:A厂商15mm、10mm、5mm病灶分别为-30、-35、-70 HU,B厂商分别为-25、-35、-65 HU;低剂量检测阈值:A厂商15mm和10mm病灶分别为-40和-45 HU,B厂商分别为-40和-50 HU,5mm病灶未达检测阈值;图像噪声常规剂量21.5±2.3 HU vs 低剂量22.2±2.0 HU(A厂商,P=.06),常规剂量25.9±4.9 HU vs 低剂量30.9±3.1 HU(B厂商,P<.001)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: To determine the size and contrast required for liver lesion detection in regular and low-dose computed tomography (CT) examinations.

MATERIALS AND METHODS: 100 abdominal CT datasets were retrospectively collected, with 50 originating from vendor A and 50 from vendor B. Half the datasets from each scanner were regular-dose oncologic examinations, the other half were acquired using a low-dose kidney stone protocol. Cylindrical liver lesions with 23 different combinations of diameter and contrast to the surrounding liver were digitally inserted. Seven radiologists assessed lesion detectability in a four-alternative forced choice reading experiment, and image noise was measured within the liver.

RESULTS: Lesion detection thresholds at regular dose were at -30, -35, and -70 Hounsfield unit (HU) lesion contrast (vendor A) and -25, -35, and -65 HU (vendor B) for lesions with 15-, 10-, and 5-mm diameter, respectively. At low dose, thresholds were -40 and -45 HU (vendor A) and -40 and -50 HU (vendor B) for 15- and 10-mm lesions, while 5-mm lesions did not reach the detection threshold. Noise levels were 21.5±2.3 HU at regular dose vs 22.2±2.0 HU at low dose for vendor A (P=.06) and 25.9±4.9 HU vs 30.9±3.1 HU for vendor B (P<.001).

CONCLUSION: In oncologic CT examinations, liver lesions with diameters of 15-, 10-, and 5-mm require contrasts of -30, -35, and -70 HU, respectively for reliable detection. In low-dose examinations, greater contrasts of -40 and -50 HU are required for lesions measuring 15- and 10-mm, while readers do not reliably detect 5-mm lesions.

来源:Acad Radiol,查看原文。版权归原作者所有。