超高清光子计数探测器CT冠状动脉斑块定量:与能量积分探测器CT的个体内比较
本研究对48例受试者的164个斑块进行个体内比较,评估超高清光子计数探测器CT与能量积分探测器CT在冠状动脉斑块定量中的差异。结果显示,光子计数探测器CT测得的总斑块体积和纤维斑块体积更低,低衰减斑块体积更高,钙化斑块体积无显著差异,且低衰减斑块测量的可重复性更好。
要点
- 超高清光子计数探测器CT测得的总斑块体积较能量积分探测器CT降低近三分之一。
- 低衰减斑块体积在光子计数探测器CT上更高,且观察者内和观察者间可重复性优于能量积分探测器CT。
- 钙化斑块体积在两种CT间无显著差异,总、钙化、纤维斑块体积在两种CT上均具有极好的一致性。
核心结论
- 与能量积分探测器CT相比,光子计数探测器CT超高清成像使分割的冠状动脉斑块体积降低近三分之一,并改善低衰减斑块测量的可重复性。
研究设计与样本
- 个体内比较研究,连续纳入48例临床指征行冠状动脉CT血管成像的受试者,在30天内接受超高清光子计数探测器CT检查;能量积分探测器CT采用192×0.6mm准直,光子计数探测器CT采用120×0.2mm准直;光子计数探测器CT使用与临床扫描等效或更低的CT剂量指数和等效对比剂用量。
诊断标准与关键数值
- 共对48例受试者的164个斑块在两种扫描上进行了分割;总斑块体积光子计数探测器CT为723.5 mm3(IQR 500.6-1184.7 mm3) vs 能量积分探测器CT为1084.7 mm3(IQR 710.7-1609.8 mm3),P < .001;纤维斑块体积光子计数探测器CT为325.4 mm3(IQR 151.7-519.2 mm3) vs 能量积分探测器CT为627.7 mm3(IQR 385.8-795.1 mm3),P < .001;低衰减斑块体积光子计数探测器CT为72.1 mm3(IQR 38.6-161.9 mm3) vs 能量积分探测器CT为58.1 mm3(IQR 23.4-102.3 mm3),P = .004;钙化斑块体积光子计数探测器CT为344.5 mm3(IQR 174.3-605.7 mm3) vs 能量积分探测器CT为342.1 mm3(IQR 180.4-607.5 mm3),P = .13。
临床应用与阅片要点
- 对于总、钙化、纤维斑块体积,两种CT的重复测量和观察者间一致性均极好(所有ICC > 0.90);低衰减斑块体积在光子计数探测器CT上观察者内ICC为0.84(95% CI: 0.57, 0.94)、观察者间ICC为0.92(95% CI: 0.81, 0.97),在能量积分探测器CT上观察者内ICC为0.62(95% CI: 0.11, 0.86)、观察者间ICC为0.47(95% CI: 0.01, 0.79)。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Background Other than enhancing the accuracy of stenosis measurements, the improved spatial resolution of photon-counting detector (PCD) CT may have an impact on quantitative plaque assessment at coronary CT angiography (CCTA). Purpose To evaluate the effect of PCD CT on coronary plaque quantification and characterization compared with that of energy-integrating detector (EID) CT. Materials and Methods Consecutive participants undergoing clinically indicated CCTA at EID CT (192 × 0.6-mm collimation) were enrolled to undergo ultrahigh-spatial-resolution (UHR) PCD CT (120 × 0.2-mm collimation) within 30 days. PCD CT was performed using equivalent or lower CT dose index and equivalent contrast media volume as the clinical scan. Total, calcified, fibrotic, and low-attenuation coronary plaque volumes were quantified and compared between scanners. Intra- and interreader reproducibility was assessed for both systems. Results A total of 164 plaques from 48 participants were segmented on both scans. Total plaque volume was lower at PCD CT compared with EID CT (723.5 mm3 [IQR, 500.6-1184.7 mm3] vs 1084.7 mm3 [IQR, 710.7-1609.8 mm3]; P < .001). UHR-based segmentations produced lower fibrotic (325.4 mm3 [IQR, 151.7-519.2 mm3] vs 627.7 mm3 [IQR, 385.8-795.1 mm3], respectively; P < .001) and higher low-attenuation plaque volumes (72.1 mm3 [IQR, 38.6-161.9 mm3] vs 58.1 mm3 [IQR, 23.4-102.3 mm3], respectively; P = .004) than EID CT-based measurements. Calcified plaque volumes did not differ significantly between PCD CT and EID CT (344.5 mm3 [IQR, 174.3-605.7 mm3] vs 342.1 mm3 [IQR, 180.4-607.5 mm3], respectively; P = .13). Total, calcified, and fibrotic plaque volumes demonstrated excellent agreement between repeated measurements and between readers for both PCD CT and EID CT (all intraclass correlation coefficients [ICCs] > 0.90). Whereas low-attenuation plaque volume had strong intrareader (ICC, 0.84; 95% CI: 0.57, 0.94) and interreader (ICC, 0.92; 95% CI: 0.81, 0.97) agreements for PCD CT, EID CT showed only moderate (ICC, 0.62; 95% CI: 0.11, 0.86) and poor (ICC, 0.47; 95% CI: 0.01, 0.79) intrareader and interreader reproducibility. Conclusion Compared with EID CT, PCD CT UHR imaging reduced segmented coronary plaque volume by nearly one-third and improved reproducibility of low-attenuation plaque measurements. © RSNA, 2025 Supplemental material is available for this article.