文献速递胸部骨肌技术与安全CT

基于深度学习的迭代重建是否影响低剂量胸部CT骨密度测量的准确性

Does the deep learning-based iterative reconstruction affect the measuring accuracy of bone mineral density in low-dose chest CT?
PubMed 文献 · Br J Radiol · 2025-01-01

研究在体模和患者中比较滤波反投影、混合迭代重建与深度学习迭代重建(AIIR)对低剂量胸部CT图像质量和骨密度测量准确性的影响。AIIR显著降低噪声、提高分辨率和信噪比,各重建组骨密度值无显著差异,AIIR图像骨密度相对误差更小。

要点

  • 深度学习迭代重建(AIIR)不改变低剂量胸部CT的骨密度测量值,各重建组间无显著差异。
  • AIIR高强度的噪声最低、信噪比和对比噪声比最高,且不影响骨密度准确性。
  • 低剂量胸部CT采用AIIR时骨密度相对误差更小,可准确测量骨密度。

核心结论

  • 基于深度学习的迭代重建不影响低剂量胸部CT骨密度测量的准确性,同时降低图像噪声并改善空间分辨率。

研究设计与样本

  • 研究分别进行体模研究和患者研究,采用相同的低剂量扫描方案。
  • 图像重建方式包括滤波反投影、混合迭代重建(KARL®,等级3、5、7)和深度学习迭代重建(人工智能迭代重建AIIR,低、中、高强度)。
  • 体模研究评估噪声功率谱和任务传递函数,并使用欧洲脊柱体模评估骨密度准确性和相对误差。
  • 由2名经验丰富的放射科医师进行主观评价,骨密度采用定量CT测量。

关键影像表现/征象

  • AIIR在体模图像上显著降低噪声并提高分辨率。
  • 所有AIIR强度均实现更低的图像噪声和更高的信噪比、对比噪声比(均P < 0.05)。
  • AIIR高强度图像噪声最低、信噪比和对比噪声比最高(P < 0.05)。

诊断标准与关键数值

  • 各重建组图像骨密度值无显著差异(均P > 0.05)。
  • AIIR图像测量的骨密度相对误差更小。
  • AIIR算法强度增加未显著影响骨密度值(均P > 0.05)。

临床应用与阅片要点

  • 低剂量胸部CT采用深度学习迭代重建时,骨密度值可准确测量,同时图像噪声降低、空间分辨率改善。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To investigate the impacts of a deep learning-based iterative reconstruction algorithm on image quality and measuring accuracy of bone mineral density (BMD) in low-dose chest CT.

METHODS: Phantom and patient studies were separately conducted in this study. The same low-dose protocol was used for phantoms and patients. All images were reconstructed with filtered back projection, hybrid iterative reconstruction (HIR) (KARL®, level of 3,5,7), and deep learning-based iterative reconstruction (artificial intelligence iterative reconstruction [AIIR], low, medium, and high strength). The noise power spectrum (NPS) and the task-based transfer function (TTF) were evaluated using phantom. The accuracy and the relative error (RE) of BMD were evaluated using a European spine phantom. The subjective evaluation was performed by 2 experienced radiologists. BMD was measured using quantitative CT (QCT). Image noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), BMD values, and subjective scores were compared with Wilcoxon signed-rank test. The Cohen's kappa test was used to evaluate the inter-reader and inter-group agreement.

RESULTS: AIIR reduced noise and improved resolution on phantom images significantly. There were no significant differences among BMD values in all groups of images (all P > 0.05). RE of BMD measured using AIIR images was smaller. In objective evaluation, all strengths of AIIR achieved less image noise and higher SNR and CNR (all P < 0.05). AIIR-H showed the lowest noise and highest SNR and CNR (P < 0.05). The increase in AIIR algorithm strengths did not affect BMD values significantly (all P > 0.05).

CONCLUSION: The deep learning-based iterative reconstruction did not affect the accuracy of BMD measurement in low-dose chest CT while reducing image noise and improving spatial resolution.

ADVANCES IN KNOWLEDGE: The BMD values could be measured accurately in low-dose chest CT with deep learning-based iterative reconstruction while reducing image noise and improving spatial resolution.

来源:Br J Radiol,查看原文。版权归原作者所有。