光子计数探测器CT的X线剂量对肺癌筛查队列中结节特性的影响:一项前瞻性研究

The Effect of X-ray Dose Photon-Counting Detector Computed Tomography on Nodule Properties in a Lung Cancer Screening Cohort: A Prospective Study.
PubMed 文献 · Invest Radiol · 2025-10-01

前瞻性纳入101例参与者,在同次检查中采集低剂量与胸片剂量光子计数探测器CT,评估172枚大于5 mm结节的检测、直径、体积与密度。X线剂量扫描的半自动检测总敏感度为0.94,实性结节体积测量对图像质量变化更稳健,Lung-RADS分类准确度良好,但结节密度有被低估的趋势。

要点

  • X线剂量扫描半自动检测结节总敏感度0.94,实性结节>0.95,亚实性结节>0.86。
  • 半自动体积测量不受剂量影响,直径测量在实性结节中被低估(P=0.01)。
  • 两位阅片者均认为X线剂量扫描上结节密度偏低(R1 P<0.001,R2 P=0.006)。

核心结论

  • 光子计数探测器CT的X线剂量扫描上半自动结节检测敏感度高,半自动体积测量比直径测量对图像质量变化更稳健,Lung-RADS分类准确,但结节密度有被低估的趋势。

研究设计与样本

  • 前瞻性研究,2023年2月至7月纳入101例参与者,同次检查行低剂量与胸片剂量非增强光子计数探测器CT,排除未转诊肺癌筛查或结节随访者及直径小于5 mm的结节,以低剂量为参考标准。

关键影像表现/征象

  • X线剂量扫描上半自动检测未检出的结节显著更小。
  • 两位阅片者均评价X线剂量扫描上结节密度更低(R1 P<0.001,R2 P=0.006)。
  • 两位阅片者测量的结节直径在两种剂量间无显著差异(R1 P=0.141,R2 P=0.554)。

诊断标准与关键数值

  • X线剂量扫描半自动检测结节总敏感度0.94,实性结节>0.95,亚实性结节>0.86。
  • 实性结节半自动直径在X线剂量扫描上被低估(P=0.01),结节体积无显著影响(P=0.775)。
  • Lung-RADS v2022分类一致性Cohen κ=0.73(半自动)、0.62(R1)、0.76(R2),准确度0.82、0.78、0.85。
  • 半自动体积测量未观察到Lung-RADS分类变化。
  • X线剂量扫描平均CTDIvol为0.11±0.03 mGy,低剂量为0.65±0.15 mGy(P<0.001)。

临床应用与阅片要点

  • 在X线剂量光子计数探测器CT上可进行准确的半自动和手动结节测量。
  • 半自动体积测量比直径测量对图像质量变化更稳健。
  • X线剂量扫描上结节密度有被低估的趋势,阅片时需结合密度评估。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The aim of the study was to evaluate the effect of photon-counting detector (PCD-)CT dose reduction to x-ray equivalent levels on nodule detection, diameter, volume, and density compared to a low-dose reference standard using semiautomated and manual methods.

MATERIALS AND METHODS: Between February and July 2023, 101 prospectively enrolled participants underwent noncontrast same-study low- and chest x-ray-dose CT scans using PCD-CT. Patients who were not referred for lung cancer screening or nodule follow-up, as well as those with nodules smaller than 5 mm in diameter, were excluded. Nodule detection and measurement of nodule diameters and volumes was semiautomatically performed for low- and x-ray-dose scans using computer-aided diagnosis software. Additionally, 2 blinded readers manually measured largest nodule diameters and examined nodule density. Nodules were classified using Lung-RADS v2022. Image quality was assessed with subjective and objective measures.

RESULTS: Mean CTDIvol for x-ray dose scans was 0.11 ± 0.03 mGy, compared to 0.65 ± 0.15 mGy for low-dose images ( P < 0.001). One hundred seventy-two nodules larger than 5 mm were detected in 53 of the 101 participants (32 male, 61.6 ± 12.5 years; 21 female, 60.3 ± 12.5 years). The semiautomated method had high overall sensitivity for nodule detection (0.94) on x-ray dose scans, with a higher sensitivity for solid nodules (>0.95) and lower for subsolid nodules (>0.86). Nodules not detected on x-ray dose scans were significantly smaller. Semiautomated measurements underestimated nodule diameter for solid nodules on x-ray dose scans ( P = 0.01), but no significant effect for nodule volume was found ( P = 0.775). Readers rated nodule density less dense on x-ray dose scans (R1: P < 0.001, R2: P = 0.006). There was no significant difference in nodule diameter for both readers between scan doses (R1: P = 0.141; R2: P = 0.554). There were good to excellent correlations between semiautomated and reader nodule diameters. Agreement and accuracy between low-dose and x-ray dose Lung-RADS classifications across methods were good (Cohens' к = 0.73, 0.62, 0.76 for semiautomated method, R1 and R2; resp. Accuracy: 0.82, 0.78, 0.85). No Lung-RADS classification changes were observed with semiautomated volumetric measurements of nodules.

CONCLUSIONS: Semiautomated nodule detection is highly sensitive in PCD-CT x-ray dose scans. Semiautomated nodule volume measurement is more robust to image quality changes than nodule diameter. Accurate semiautomated and manual nodule measurements are feasible on x-ray dose scans, but nodule density was in tendency underestimated. Nodule classification using Lung-RADS was shown to be accurate on x-ray dose scans.

来源:Invest Radiol,查看原文。版权归原作者所有。