亚急性支架变形是p64血流导向装置治疗后血管狭窄的潜在原因:生物学机制与后果的回顾与讨论
研究纳入48例远端ICA动脉瘤患者,均置入p64血流导向支架,于术后第14天行高剂量透视评估支架状态。58%的病例观察到支架变形,变形与随访中载瘤血管管径狭窄相关,35%的病例出现明显支架内狭窄,提示亚急性支架变形是血流导向治疗后血管狭窄的潜在机制。
要点
- p64支架置入后亚急性支架变形发生率为58%,随访中35%的病例出现明显支架内狭窄。
- 术后第14天高剂量透视可评估内皮化前的支架状态,有助于发现亚急性变形。
- 支架变形与随访中载瘤血管管径狭窄存在相关性。
核心结论
- p64支架血流导向治疗中,亚急性支架变形发生于不可忽视比例的病例,且随访中变形与后续支架内狭窄存在统计学显著关联。
研究设计与样本
- 单中心研究纳入48例31至71岁(8例男性)远端ICA动脉瘤患者,均以p64血流导向支架治疗,术后第14天行高剂量透视短期随访评估支架状态。
关键影像表现/征象
- 58%的病例观察到支架变形。
- 71%的病例术后出现轻度/中度单侧头痛。
- 1例观察到支架变形的患者出现数次急性对侧单侧肢体无力发作。
- 随访中建立支架变形与载瘤血管管径后续狭窄之间的相关性。
- 35%的病例报告明显支架内狭窄。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Flow-diverter stents are a potent and efficient tool in the instrumentarium of neurointerventional radiologists for the treatment of intracranial aneurysms. With their implementation, some adverse effects and complications such as hemorrhagic and ischemic ones, have been seen as a potential downfall of the method. In-stent stenosis is one such complication, which until now has not received enormous attention due to its seemingly benign characteristic. In our study we propose a different point of view on this matter and aim to establish a potential mechanism for its development: a subacute postprocedural stent deformation, due to segmental vessel constriction as a reaction to the implant.
MATERIALS AND METHODS: We enrolled 48 patients between the ages of 31 and 71 (8 men) with aneurysms on the distal portions of the ICA, all of whom were treated with the p64 flow-diverter stent, to assess the incidence of this phenomenon, as well as establish a correlation between it and subsequent clinical symptoms. A protocol for short-term follow-up, consisting only of a high-dose fluoroscopy image of the implant (conducted on the 14th postprocedural day) was implemented to assess the state of the implant before endothelization was to be expected.
RESULTS: Stent deformities were seen in 58% of cases. One patient with an observed stent deformity presented with several episodes of acute contralateral 1-sided weakness of the limbs. Seventy-one percent of those patients presented with a mild/moderate unilateral headache postprocedurally. A correlation between the deformity and a subsequent narrowing of the parent vessel diameter was established on follow-ups. Notable in-stent stenosis was reported in 35% of all cases.
CONCLUSIONS: Cases with subacute stent deformities are presented in a nonnegligible percent of flow-diversion therapies with the p64 stent. A statistically significant association between the observed deformation and subsequent in-stent stenosis was observed on follow-up.
引用
- Vladev G, Sirakov A, Matanov S, et al. Subacute Stent Deformities as an Underlying Reason for Vessel Stenosis after Flow Diversion with the p64 Stent: Review and Discussion of Biologic Mechanisms and Consequences. AJNR Am J Neuroradiol. 2025. PMID: 40113252. doi:10.3174/ajnr.A8564
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。