经颅超声超分辨成像中血脑屏障破坏的可能性
本研究旨在量化三维经颅超声超分辨成像中血脑屏障破坏(BBBD)与机械指数(MI)及压力分布的关系。小鼠实验显示,换能器表面MI为0.78时出现显著BBBD(p<0.05),而低于0.78时可在无显著染料外渗的情况下完成成像;局部MI在0.44-0.53范围可能产生血脑屏障生物效应。
要点
- 换能器表面MI为0.78时出现显著血脑屏障破坏(p<0.05),低于0.78时可在无显著染料外渗下完成成像。
- 换能器表面MI为0.64时脑内局部MI平均0.18±0.09,未见显著血脑屏障破坏。
- 产生血脑屏障生物效应的局部MI在0.44-0.53范围。
核心结论
- 三维经颅超声超分辨成像在换能器表面MI为0.78时可导致血脑屏障破坏,而低于0.78时可在无显著染料外渗的情况下完成成像;产生血脑屏障生物效应的局部MI在0.44-0.53范围。
研究设计与样本
- 小鼠研究,采用常规三维超分辨序列成像,通过荧光成像量化染料外渗以评估血脑屏障破坏,机械指数范围为0-0.78(水中测量),并进行水听器测量和三维模拟估算脑内局部压力分布。
诊断标准与关键数值
- 换能器表面MI为0.64时,脑内局部MI平均0.18±0.09,范围0.03-0.44,未见显著血脑屏障破坏。
- 换能器表面MI为0.78时,出现显著血脑屏障破坏(p<0.05),脑内局部MI范围0.04-0.53,平均0.22±0.11。
- 产生血脑屏障生物效应的局部MI在0.44-0.53范围。
- 换能器表面MI为0.78时,水听器测量显示4.8%的总体脑体积压力高于0.44 MI(扫描的100 mm²二维平面中4.8 mm²),模拟显示11.10%(777.5 mm³体积中74.6 mm³)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Super-resolution (SR) ultrasound imaging dramatically improves the resolution of microvascular images beyond the diffraction limit. It is well-established that ultrasound with contrast agents, such as those used in SR, can open the blood-brain barrier (BBB) under certain conditions. This study aimed to quantify potential BBB disruption (BBBD) during 3-D transcranial ultrasound SR as a function of mechanical index (MI) and pressure distribution. Mice were imaged using conventional 3-D SR sequences. BBBD was quantified through fluorescence imaging of dye extravasation for mechanical indices in the range of 0-0.78 (measured in water). The results suggest that when 3-D SR images are acquired at 0.78 MI, BBBD occurs; however, imaging was achievable without significant dye extravasation below 0.78. In addition, hydrophone measurements and 3-D simulations were performed to estimate local pressure distributions in the brain. For a transducer surface MI of 0.64, estimates of the local MI within the brain averaged 0.18 ± 0.09, ranging from 0.03 to 0.44, with no significant BBBD observed. However, for a transducer surface MI of 0.78, significant BBBD was observed (p < 0.05), where the values in the brain range from 0.04 to 0.53, averaging 0.22 ± 0.11. This suggests that the local MI that generates BBB bio-effects is within the range of 0.44-0.53 MI. For the 0.78 MI case, 4.8% of the total brain volume had a pressure above 0.44 MI according to hydrophone measurements (4.8 mm2 of the 100 mm2 2-D plane scanned) and 11.10% according to simulations (74.6 mm3 of the 777.5 mm3 volume).
引用
- Lee HR, Jones RM, Durham PG, et al. The Potential for Blood-Brain Barrier Disruption During Transcranial Ultrasound Super-Resolution Imaging. Ultrasound Med Biol. 2025. PMID: 40107891. doi:10.1016/j.ultrasmedbio.2025.02.006
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。