低剂量儿童心胸CT中定位像与团注跟踪的辐射剂量占比
该研究评估低剂量儿童心胸CT中定位像和团注跟踪的辐射剂量占比:在852例婴儿中,20 mA组定位像剂量占比2.5%显著低于35 mA组的6.5%,且解剖标志均可见;在1015例≤2岁儿童中,团注跟踪剂量占比与监测扫描次数相关性最强(R=0.93)。
要点
- 低管电流(20 mA)定位像剂量占比2.5%,显著低于35 mA组的6.5%,解剖标志仍均可见。
- 定位像剂量占比与over z-axis覆盖比例高度相关(35 mA组R=0.736,20 mA组R=0.660)。
- 团注跟踪剂量占比与监测扫描次数呈最强正相关(R=0.93),年龄越小占比可能越高。
核心结论
- 低剂量儿童心胸CT中,低管电流设置可大幅降低定位像剂量占比并保持图像质量;应关注减少over z-axis覆盖比例;监测扫描次数是增加团注跟踪辐射剂量占比的最重要因素,尤其在低龄儿童中。
研究设计与样本
- 研究纳入852例年龄<1岁婴儿行低剂量儿童心胸CT,采用80 kV后前位定位像,管电流35 mA或20 mA;采用倾向性评分匹配比较两组剂量占比等;另对1015例≤2岁儿童行团注跟踪,评估监测扫描次数等因素对剂量占比的影响。
诊断标准与关键数值
- 定位像剂量占比:20 mA组2.5%(n=94)vs 35 mA组6.5%(n=94),P<0.001。
- 图像噪声:20 mA组2.1 Hounsfield units vs 35 mA组1.0 Hounsfield units,P<0.001;所有病例解剖标志均可见。
- 定位像剂量占比与over z-axis覆盖比例相关性:35 mA组R=0.736(P<0.001),20 mA组R=0.660(P<0.001)。
- 团注跟踪剂量-length product占比与监测扫描次数相关性最强(R=0.93,P<0.001);与年龄、诊断扫描剂量-length product、水等效面积呈弱负相关(R=-0.46~-0.50,P<0.001)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Optimization of localizer radiograph and bolus tracking doses is necessary, as their dose proportion may increase with a decreasing diagnostic scan dose in pediatric cardiothoracic computed tomography (CT).
OBJECTIVE: To evaluate the radiation dose proportions of the localizer radiograph and bolus tracking in low-dose pediatric cardiothoracic CT.
MATERIALS AND METHODS: For low-dose pediatric cardiothoracic CT, a posteroanterior localizer radiograph was acquired with 80 kV, and 35 mA or 20 mA in 852 infants (age<1 year). Propensity score matching was employed in comparing the 35 mA and 20 mA groups on dose proportion, over z-axis proportion, visibility of anatomic landmarks, and image noise. The over z-axis coverage proportion was correlated with the dose proportion of the localizer radiograph in both groups. Additionally, bolus tracking was performed in 1,015 children (≤2 years). The effects of the number of monitoring scan, dose-length product of the diagnostic scan, age, and water-equivalent area of the scanned patient body on the radiation dose proportion of bolus tracking were evaluated.
RESULTS: The dose proportion of the localizer radiograph was significantly lower in the 20 mA group (2.5%, n = 94) than in the 35 mA group (6.5%, n = 94) (P < 0.001). While image noise was higher in the 20 mA group (2.1 Hounsfield units versus 1.0 Hounsfield units of the 35 mA-group, P < 0.001), all the anatomic landmarks remained visible in all cases. The over z-axis coverage proportion demonstrated high correlations with the dose proportion for both groups (R = 0.736, P < 0.001 for the 35 mA group and R = 0.660, P < 0.001 for the 20 mA group). The bolus tracking dose-length product proportion demonstrated the strongest positive correlation with the number of monitoring scans (R = 0.93, P < 0.001), while age, diagnostic scan dose-length product, and water-equivalent area showed weak negative correlations (R-values = -0.46~-0.50, P-values < 0.001).
CONCLUSIONS: In low-dose pediatric cardiothoracic CT, the dose proportion of the localizer radiograph can be substantially reduced with a low tube current setting while maintaining image quality. Additionally, minimization of the over z-axis coverage proportion merits attention. The number of monitoring scans is the most significant factor for increasing the radiation dose proportion of bolus tracking, especially in young ages.
引用
- Goo HW, Goo SY. Radiation dose proportions of localizer radiograph and bolus tracking in low-dose pediatric cardiothoracic computed tomography. Pediatr Radiol. 2025. PMID: 39888399. doi:10.1007/s00247-025-06163-z