静脉CT对比剂对腰椎松质骨骨密度内部校准技术的影响
本研究在34名受试者共93次腰椎临床CT扫描中,比较基于体模与内部校准技术测定的松质骨骨密度在平扫与增强扫描间的差异。对比剂使体模校准骨密度在动脉期和门静脉期分别显著升高8.6±14.7 mg/cm3和14.4±16.3 mg/cm3;以皮下脂肪/空气或肌肉/空气为内部参考时,内部校准骨密度的升高与体模校准升高无显著差异(p>0.1)。结论认为内部校准无法减轻体模校准中观察到的对比剂相关骨密度升高。
要点
- 对比剂使体模校准的腰椎松质骨骨密度在动脉期升高8.6±14.7 mg/cm3、门静脉期升高14.4±16.3 mg/cm3(p<0.01)。
- 以皮下脂肪/空气或肌肉/空气为内部参考时,内部校准骨密度升高与体模校准升高无显著差异(p>0.1)。
- 对比剂对肌肉和皮下脂肪CT值的影响在动脉期<5HU、门静脉期<10HU。
核心结论
- 虽然对比剂给药后内部参考材料的CT值升高,但内部校准无法减轻体模校准中观察到的对比剂相关骨密度升高。
研究设计与样本
- 在34名受试者共93次腰椎临床CT扫描(平扫及增强扫描)中,比较基于体模与内部校准技术测定的松质骨骨密度。
关键影像表现/征象
- 动脉期主动脉血液CT值升高210±99HU,下腔静脉血液升高47±35HU。
- 门静脉期主动脉血液CT值升高110±49HU,下腔静脉血液升高73±28HU。
- 对比剂对肌肉和皮下脂肪CT值的影响在动脉期<5HU、门静脉期<10HU。
诊断标准与关键数值
- 体模校准骨密度在动脉期升高8.6±14.7 mg/cm3,门静脉期升高14.4±16.3 mg/cm3(p<0.01)。
- 以皮下脂肪/空气或肌肉/空气为内部参考材料时,内部校准骨密度升高在动脉期<12 mg/cm3、门静脉期<17 mg/cm3,与体模校准升高无显著差异(p>0.1)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Contrast agents are frequently administered in computed tomography (CT) scans used for opportunistic screening of osteoporosis. The objective of this study is to compare the impact of contrast-related bone mineral density (BMD) increase between phantom-based and internal CT calibration techniques.
MATERIALS AND METHODS: Phantom-based and internal CT calibration techniques were used to determine trabecular BMD in 93 existing clinical CT scans of the lumbar spine of 34 subjects, scanned before and after administration of contrast agents. BMD differences between native (Nat) and contrast-enhanced scans in arterial (Art) and portal venous (PV) phases were compared among calibration methods. Three pairs of internal reference materials were investigated: blood/air, subcutaneous adipose tissue (SAT)/air and muscle tissue/air.
RESULTS: Contrast agents increased CT values of blood in the aorta and the inferior vena cava in the Art phase by 210 ± 99HU and 47 ± 35HU, respectively and in the PV phase by 110 ± 49HU and 73 ± 28HU, respectively. Effects on CT values of muscle and SAT were < 5HU in the Art phase and < 10HU in the PV phase. BMD obtained with phantom calibration increased significantly (p < 0.01) in Art and PV phases by 8.6 ± 14.7 mg/cm3 and 14.4 ± 16.3 mg/cm3, respectively. If SAT/air or muscle/air were used as internal reference materials, increases in internally calibrated BMD (<12 mg/cm3 in AT phase and < 17 mg/cm3 in PV phase) did not significantly differ (p > 0.1 for Art and PV phases) from increases of phantom-calibrated BMD.
CONCLUSIONS: Although CT values of internal reference materials increased following the administration of contrast, internal calibration was unable to mitigate the contrast-related BMD increases observed in phantom-based calibration.
引用
- Bartenschlager S, Cavallaro A, Pogarell T, et al. Impact of intravenous CT contrast agents on internal calibration techniques to determine trabecular BMD of the lumbar spine. Eur J Radiol. 2025. PMID: 39823659. doi:10.1016/j.ejrad.2025.111923