面向大尺寸器官的实时容积自由手超声成像:全脊柱成像研究
本研究提出一种实时容积自由手超声成像系统,采用增量成像与算法加速实现大尺寸器官扫描数据的实时处理与可视化,并用两种组织分割算法重建脊柱解剖。在6例脊柱侧凸患者和10名正常志愿者中验证,从肩到臀超过500 mm的脊柱超声成像在2分钟内完成,实时三维成像所示脊柱弯曲度与X线图像的相关系数为0.96。作者认为该系统可能具备临床应用于其他大尺寸器官的潜力。
要点
- 从肩到臀超过500 mm的脊柱容积超声成像可在2分钟内完成。
- 实时三维超声显示的脊柱弯曲度与X线图像相关系数为0.96。
- 系统已在多型号超声设备及不同类型探头上验证适应性。
核心结论
- 所提出的容积超声成像系统可能具备临床应用于其他大尺寸器官的潜力。
研究设计与样本
- 本研究为方法学研究,提出实时容积自由手超声成像系统,采用增量成像与算法加速及两种组织分割算法;在6例脊柱侧凸患者和10名正常志愿者中验证,并与X线图像对照。
关键影像表现/征象
- 通过近似骨结构所在深度并对不同深度超声图像进行分割,在三维空间中重建并可视化脊柱解剖。
诊断标准与关键数值
- 从肩到臀超过500 mm的脊柱超声成像在2分钟内完成。
- 实时显示的三维成像结果与对应X线图像在脊柱弯曲度上的相关系数为0.96。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Three-dimensional (3D) ultrasound imaging can overcome the limitations of conventional two-dimensional (2D) ultrasound imaging in structural observation and measurement. However, conducting volumetric ultrasound imaging for large-sized organs still faces difficulties including long acquisition time, inevitable patient movement, and 3D feature recognition. In this study, we proposed a real-time volumetric free-hand ultrasound imaging system optimized for the above issues and applied it to the clinical diagnosis of scoliosis.
METHODS: This study employed an incremental imaging method coupled with algorithmic acceleration to enable real-time processing and visualization of the large amounts of data generated when scanning large-sized organs. Furthermore, to deal with the difficulty of image feature recognition, we proposed two tissue segmentation algorithms to reconstruct and visualize the spinal anatomy in 3D space by approximating the depth at which the bone structures are located and segmenting the ultrasound images at different depths.
RESULTS: We validated the adaptability of our system by deploying it to multiple models of ultrasound equipment and conducting experiments using different types of ultrasound probes. We also conducted experiments on six scoliosis patients and 10 normal volunteers to evaluate the performance of our proposed method. Ultrasound imaging of a volunteer spine from shoulder to crotch (more than 500 mm) was performed in 2 minutes, and the 3D imaging results displayed in real-time were compared with the corresponding X-ray images with a correlation coefficient of 0.96 in spinal curvature.
CONCLUSION: Our proposed volumetric ultrasound imaging system might hold the potential to be clinically applied to other large-sized organs.
引用
- Li C, Shen E, Wang H, et al. Real-Time Volumetric Free-Hand Ultrasound Imaging for Large-Sized Organs: A Study of Imaging the Whole Spine. Ultrasound Med Biol. 2025. PMID: 39757051. doi:10.1016/j.ultrasmedbio.2024.12.015