深度学习去噪算法在全身体光子计数CT成像中降低辐射剂量的评估:一项尸体研究
研究采用24具尸体在光子计数CT上以100%、50%、25%、10% mAs四种剂量水平行全身扫描,比较QIR 2级与深度学习去噪算法重建的图像质量,共192组数据集。客观分析显示去噪后噪声更低,25% mAs去噪与100% mAs原始数据集在噪声和CNR上无显著差异;主观评价中50% mAs去噪优于100% mAs原始,25% mAs去噪与100% mAs原始无显著差异。结论认为该算法可在维持图像质量的同时将辐射剂量降低达75%。
要点
- 25% mAs去噪数据集与100% mAs原始数据集在噪声和CNR上无显著差异。
- 50% mAs去噪数据集主观评分优于100% mAs原始数据集。
- 去噪算法未显著改变肝脏实质CT值。
核心结论
- 深度学习去噪算法可在光子计数CT成像中维持图像质量,同时实现高达75%的辐射剂量降低。
研究设计与样本
- 尸体研究,24具尸体,在光子计数CT(NAEOTOM Alpha,Siemens Healthineers)上以100%、50%、25%、10% mAs四个剂量水平行全身扫描,分别用QIR 2级和深度学习算法(ClariCT.AI,ClariPi Inc.)重建,共192组数据集;以肝脏实质固定ROI测量CT值稳定性、图像噪声和CNR进行客观评价,并由两名放射科医生独立评价整体图像清晰度、锐利度和对比度。
诊断标准与关键数值
- 客观评价指标为肝脏实质ROI的CT值稳定性、图像噪声和CNR;主观评价指标为整体图像清晰度、锐利度和对比度。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: Photon Counting CT (PCCT) offers advanced imaging capabilities with potential for substantial radiation dose reduction; however, achieving this without compromising image quality remains a challenge due to increased noise at lower doses. This study aims to evaluate the effectiveness of a deep learning (DL)-based denoising algorithm in maintaining diagnostic image quality in whole-body PCCT imaging at reduced radiation levels, using real intraindividual cadaveric scans.
MATERIALS AND METHODS: Twenty-four cadaveric human bodies underwent whole-body CT scans on a PCCT scanner (NAEOTOM Alpha, Siemens Healthineers) at four different dose levels (100%, 50%, 25%, and 10% mAs). Each scan was reconstructed using both QIR level 2 and a DL algorithm (ClariCT.AI, ClariPi Inc.), resulting in 192 datasets. Objective image quality was assessed by measuring CT value stability, image noise, and contrast-to-noise ratio (CNR) across consistent regions of interest (ROIs) in the liver parenchyma. Two radiologists independently evaluated subjective image quality based on overall image clarity, sharpness, and contrast. Inter-rater agreement was determined using Spearman's correlation coefficient, and statistical analysis included mixed-effects modeling to assess objective and subjective image quality.
RESULTS: Objective analysis showed that the DL denoising algorithm did not significantly alter CT values (p ≥ 0.9975). Noise levels were consistently lower in denoised datasets compared to the Original (p < 0.0001). No significant differences were observed between the 25% mAs denoised and the 100% mAs original datasets in terms of noise and CNR (p ≥ 0.7870). Subjective analysis revealed strong inter-rater agreement (r ≥ 0.78), with the 50% mAs denoised datasets rated superior to the 100% mAs original datasets (p < 0.0001) and no significant differences detected between the 25% mAs denoised and 100% mAs original datasets (p ≥ 0.9436).
CONCLUSION: The DL denoising algorithm maintains image quality in PCCT imaging while enabling up to a 75% reduction in radiation dose. This approach offers a promising method for reducing radiation exposure in clinical PCCT without compromising diagnostic quality.