基于1.5-T MRI的R2*法测量肝脏铁浓度的可重复性:一项Meta分析
该系统评价纳入21项研究共1147名参与者,评估1.5-T MRI基于R2*法测量肝脏铁浓度在不同扫描仪、扫描参数及后处理技术间的可重复性。初始回波时间≤1.0 ms且回波间距≤1.4 ms的7项研究(435名参与者)两两比较R2*值无显著差异(p>0.05)。R2*与肝脏铁浓度的校准关系为肝脏铁浓度(mg/g)=0.042+2.85×10⁻²×R2*(s⁻¹)(R²=0.79)。
要点
- 初始TE≤1.0 ms且TE间距≤1.4 ms时,1.5-T MRI的R2*法测量肝脏铁浓度在不同扫描仪、参数及后处理间可重复。
- R2*与肝脏铁浓度的校准公式为肝脏铁浓度(mg/g)=0.042+2.85×10⁻²×R2*(s⁻¹),R²=0.79。
- 初始TE≤1.0 ms且TE间距≤1.4 ms的7项研究(435名参与者)两两比较R2*值无显著差异(p>0.05)。
核心结论
- 基于1.5-T MRI的R2*法在估算肝脏铁浓度时,在不同MRI扫描仪、扫描参数和后处理技术间具有可重复性。
研究设计与样本
- 系统检索PubMed、Embase、Medline、Cochrane Library和Web of Science数据库,纳入使用1.5-T MRI的R2*法测量肝脏铁浓度的研究。
- 共纳入21项研究、1147名参与者,提取原始数据。
- 以扫描设备、扫描参数和后处理技术作为随机效应,在校正所测肝脏铁浓度的影响后,判断不同研究获得的R2*值是否有显著差异。
- 通过单变量回归估计R2*值与肝脏铁浓度最佳拟合的校准曲线。
诊断标准与关键数值
- 初始TE≤1.0 ms且TE间距≤1.4 ms的7项研究(435名参与者)两两比较所测R2*值无显著差异(p>0.05)。
- R2*与肝脏铁浓度的校准关系为:肝脏铁浓度(mg/g)=0.042+2.85×10⁻²×R2*(s⁻¹)(R²=0.79)。
临床应用与阅片要点
- 采用1.5-T MRI的R2*法量化肝脏铁浓度时,扫描序列的初始TE≤1.0 ms且TE间距≤1.4 ms有助于获得可重复的测量结果。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To explore the reproducibility of the 1.5-T MR imaging (MRI)-based R2* method in measuring the liver iron concentration (LIC) across different MRI scanners, scan parameters, and postprocessing techniques.
MATERIALS AND METHODS: We performed a systematic search of the PubMed, Embase, Medline, Cochrane Library, and Web of Science databases and identified studies that used the 1.5-T MRI-based R2* method to measure the LIC. The original data were extracted from the selected studies. Reproducibility was assessed in terms of the ability to replicate the measured LIC with the 1.5-T MRI-based R2* method across different scanning instruments, scan parameters, and postprocessing techniques. We used the scanner equipment, scan parameters, and postprocessing technique as random effects to determine whether the R2* values obtained in the different studies were significantly different after the effects of the measured LIC were excluded. The calibration curve that best fit the R2* value to the LIC was estimated via univariate regression.
RESULTS: Twenty-one studies (1147 participants) were included. The seven studies (435 participants) in which the initial TE was ≤ 1.0 ms and the TE spacing was ≤ 1.4 ms did not have significant differences in the measured R2* value in pairwise comparisons (p > 0.05). Calibration of R2* to LIC was as follows: LIC (mg/g) = 0.042 + 2.85 × 10-2 R2* (s-1) (R2 = 0.79).
CONCLUSION: The 1.5-T MRI-based R2* method in estimating LIC was reproducible across different MRI scanners, scan parameters, and postprocessing techniques.
KEY POINTS: Question The R2* method is widely used to quantify LIC, but the reproducibility of R2*-based LIC estimation remains unknown. Findings In scan sequences with an initial echo time (TE) ≤ 1.0 ms and a TE spacing ≤ 1.4 ms, the R2*-based LIC quantification at 1.5 T achieved good reproducibility. Clinical relevance The 1.5-T MRI-based R2* method enables reproducible quantification of the LIC; its reproducibility spares patients from unnecessary trauma and economic burdens while promoting the extensive clinical dissemination of R2*-based LIC quantification methods.