基于深度学习降噪的运动补偿多激发胰腺扩散加权成像
前瞻性纳入22例患者,比较无运动补偿(M0)、凸优化扩散编码生成的运动补偿梯度(M1)及M1联合深度学习降噪(M1+DL)的2激发胰腺DWI;M1在运动伪影和信号均匀性上优于M0,DL降噪可逆转M1的噪声劣势并保持ADC一致性。
要点
- M1在运动伪影和信号均匀性上显著优于M0(P<0.001)。
- DL降噪使M1+DL在感知噪声上显著优于M0(P<0.001)。
- M0的ADC值在胰头、体、尾间存在差异(P=0.001),M1和M1+DL则均匀(P=0.24)。
核心结论
- CODE生成的运动补偿梯度改善了多激发胰腺DWI的专家阅片表现,并消除了全胰腺ADC变异。
- 深度学习降噪减轻了运动补偿带来的信号损失,同时保持ADC一致性。
- 整合两种技术可能提高多激发胰腺DWI的准确性和可靠性。
研究设计与样本
- 前瞻性、机构审查委员会批准的研究,纳入22例患者,于2022年8月22日至2023年5月17日在3.0 T扫描仪上行腹部MR检查。
- 在临床检查中加入2激发自旋回波平面DWI(b=0、800 s/mm²),分别不施加(M0)和施加(M1)CODE生成的一阶梯度矩置零。
- 对M1图像离线应用基于深度学习的降噪(M1+DL),并为所有3种方法重建ADC图。
- 由3名亚专科放射科医生对b=800 s/mm²图像进行盲法配对比较,比较5项指标:胰腺边界勾画、运动伪影、信号均匀性、感知噪声和诊断偏好。
- 勾画胰头、体、尾感兴趣区并计算平均ADC值,采用重复测量方差分析和Bonferroni校正的事后配对t检验比较平均ADC值,Bland-Altman分析比较平均ADC值,读者偏好采用Wilcoxon符号秩检验(Bonferroni校正)和Fleiss κ比较。
关键影像表现/征象
- M1在感知运动伪影和信号均匀性上显著优于M0(P<0.001)。
- M0在感知噪声上显著优于M1(P<0.001),但基于深度学习的降噪(M1+DL)逆转了这一趋势,且M1+DL显著优于M0(P<0.001)。
诊断标准与关键数值
- M0的ADC测量值在胰腺不同区域间存在差异(P=0.001),而M1和M1+DL运动校正后ADC值均匀(P=0.24),与文献报道的运动校正ssDWI值相似。
- M0的ADC值在胰头(偏倚16.6%;P<0.0001)、胰体(偏倚11.0%;P<0.0001)和胰尾(偏倚8.6%;P=0.001)均显著高于M1。
- M1与M1+DL的ADC值之间存在小但显著的偏倚(2.6%)。
临床应用与阅片要点
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Pancreatic diffusion-weighted imaging (DWI) has numerous clinical applications, but conventional single-shot methods suffer from off resonance-induced artifacts like distortion and blurring while cardiovascular motion-induced phase inconsistency leads to quantitative errors and signal loss, limiting its utility. Multishot DWI (msDWI) offers reduced image distortion and blurring relative to single-shot methods but increases sensitivity to motion artifacts. Motion-compensated diffusion-encoding gradients (MCGs) reduce motion artifacts and could improve motion robustness of msDWI but come with the cost of extended echo time, further reducing signal. Thus, a method that combines msDWI with MCGs while minimizing the echo time penalty and maximizing signal would improve pancreatic DWI. In this work, we combine MCGs generated via convex-optimized diffusion encoding (CODE), which reduces the echo time penalty of motion compensation, with deep learning (DL)-based denoising to address residual signal loss. We hypothesize this method will qualitatively and quantitatively improve msDWI of the pancreas.
MATERIALS AND METHODS: This prospective institutional review board-approved study included 22 patients who underwent abdominal MR examinations from August 22, 2022 and May 17, 2023 on 3.0 T scanners. Following informed consent, 2-shot spin-echo echo-planar DWI (b = 0, 800 s/mm 2 ) without (M0) and with (M1) CODE-generated first-order gradient moment nulling was added to their clinical examinations. DL-based denoising was applied to the M1 images (M1 + DL) off-line. ADC maps were reconstructed for all 3 methods. Blinded pair-wise comparisons of b = 800 s/mm 2 images were done by 3 subspecialist radiologists. Five metrics were compared: pancreatic boundary delineation, motion artifacts, signal homogeneity, perceived noise, and diagnostic preference. Regions of interest of the pancreatic head, body, and tail were drawn, and mean ADC values were computed. Repeated analysis of variance and post hoc pairwise t test with Bonferroni correction were used for comparing mean ADC values. Bland-Altman analysis compared mean ADC values. Reader preferences were tabulated and compared using Wilcoxon signed rank test with Bonferroni correction and Fleiss κ.
RESULTS: M1 was significantly preferred over M0 for perceived motion artifacts and signal homogeneity ( P < 0.001). M0 was significantly preferred over M1 for perceived noise ( P < 0.001), but DL-based denoising (M1 + DL) reversed this trend and was significantly favored over M0 ( P < 0.001). ADC measurements from M0 varied between different regions of the pancreas ( P = 0.001), whereas motion correction with M1 and M1 + DL resulted in homogeneous ADC values ( P = 0.24), with values similar to those reported for ssDWI with motion correction. ADC values from M0 were significantly higher than M1 in the head (bias 16.6%; P < 0.0001), body (bias 11.0%; P < 0.0001), and tail (bias 8.6%; P = 0.001). A small but significant bias (2.6%) existed between ADC values from M1 and M1 + DL.
CONCLUSIONS: CODE-generated motion compensating gradients improves multishot pancreatic DWI as interpreted by expert readers and eliminated ADC variation throughout the pancreas. DL-based denoising mitigated signal losses from motion compensation while maintaining ADC consistency. Integrating both techniques could improve the accuracy and reliability of multishot pancreatic DWI.
引用
- Wang K, Middione MJ, Loening AM, et al. Motion-Compensated Multishot Pancreatic Diffusion-Weighted Imaging With Deep Learning-Based Denoising. Invest Radiol. 2025. PMID: 39823511. doi:10.1097/RLI.0000000000001148
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。