文献速递腹部MR

整合层面特异性动态匀场EPI与单次激发EPI肝脏扩散加权成像的比较

Comparisons of integrated slice-specific dynamic shimming EPI and single-shot EPI diffusion-weighted imaging of the liver.
PubMed 文献 · Eur J Radiol · 2025-01-19

纳入142例患者,在注射钆塞酸二钠前采集SS-EPI与原型iShim-EPI图像,由两名放射科医师评估图像质量及左肝叶信号丢失,并测量最大病灶的平均ADC值。iShim-EPI图像质量显著优于SS-EPI(ICC 0.843至0.991,p<0.05),左肝叶信号丢失大幅减少;两者ADC值高度相关(r=0.93,p<0.001)且一致性良好。iShim-EPI在鉴别良恶性肝脏病变方面表现更好,鉴别肝癌与良性病变的AUC显著更大(p<0.05),作者认为其是鉴别良恶性肝脏病变有前景的方法。

要点

  • iShim-EPI图像质量显著优于SS-EPI(ICC 0.843至0.991,p<0.05),左肝叶信号丢失大幅减少。
  • 两序列ADC值高度相关(r=0.93,p<0.001),一致性良好(CI -475.5~722.1×10-6 mm2/s)。
  • iShim-EPI鉴别肝癌与良性病变的AUC显著大于SS-EPI(p<0.05)。

核心结论

  • iShim-EPI DWI是鉴别良恶性肝脏病变有前景的方法,较SS-EPI图像质量更好、左肝叶信号丢失更少,并可提高疾病诊断信心。

研究设计与样本

  • 共纳入142例患者,于注射钆塞酸二钠增强肝脏MRI前采集SS-EPI与原型iShim-EPI图像,由两名放射科医师评估图像质量并测量最大病灶平均ADC值。

关键影像表现/征象

  • iShim-EPI的图像质量显著优于SS-EPI(ICC 0.843至0.991,p<0.05)。
  • iShim-EPI的左肝叶信号丢失较SS-EPI大幅减少。
  • 两序列ADC值高度相关(r=0.93,p<0.001)且一致性良好(CI -475.5~722.1×10-6 mm2/s)。

诊断标准与关键数值

  • iShim-EPI检测良性病变(肝血管瘤和囊肿)和恶性病变(原发性肝癌和肝转移瘤)的表现优于SS-EPI。
  • iShim-EPI鉴别肝癌与良性病变(肝血管瘤和囊肿)的AUC显著大于SS-EPI(p<0.05)。

临床应用与阅片要点

  • 与SS-EPI相比,iShim-EPI可提高图像质量、减少左肝叶信号丢失并增强疾病诊断信心。
  • 两序列ADC值高度相关且一致性良好,提示iShim-EPI ADC值可用于肝脏病灶评估。
  • 作者认为iShim-EPI DWI是鉴别良恶性肝脏病变有前景的方法。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To compare the quality of DWI images, signal loss of left hepatic lobe and diagnostic performance of apparent diffusion coefficient (ADC) values between SS-EPI and iShim-EPI in liver lesions.

METHODS: Totally 142 patients were involved, images using SS-EPI and the prototype iShim-EPI were acquired before injection of gadoxetic acid-enhanced liver MRI.Image quality of demarcation of liver capsule, resolution, lesion distortion, artifacts, lesion confidence score, and signal loss in left hepatic lobe was assessed by two radiologists. Mean ADC values of the largest lesions were measured, and the correlations, agreements, and diagnostic performances were compared between the two sequences.

RESULTS: Image quality of the iShim-EPI was significantly improved over that of SS-EPI (ICC 0.843 to 0.991, p < 0.05), the signal loss in the left hepatic lobe was greatly reduced. The ADC values were highly correlated (r = 0.93, p < 0.001) and had good agreement (CI -475.5 ∼ 722.1 × 10-6 mm2/s) between the two sequences. Compared with SS-EPI,iShim-EPI had better performance in detecting benign (hepatic haemangioma and cyst) and malignant (primary liver cancer and hepatic metastases) diseases. Furthermore, iShim-EPI had a significantly larger AUC in differentiating cancer from benign lesions (both hepatic haemangioma and cyst) (p < 0.05).

CONCLUSIONS: IShim-EPI DWI is a promising method for differentiating benign and malignant liver lesions with better image quality, less signal loss of left hepatic lobe and could enhance the confidence of diseases diagnosis compared with SS-EPI.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Liu S, Lin Y, Situ W, et al. Comparisons of integrated slice-specific dynamic shimming EPI and single-shot EPI diffusion-weighted imaging of the liver. Eur J Radiol. 2025. PMID: 39848125. doi:10.1016/j.ejrad.2025.111942

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。