不同厂商软件评估二尖瓣的观察者间与观察者内变异性:对经导管二尖瓣修复术的意义

Interobserver and intraobserver variability among different vendors for mitral valve assessment: implications for transcatheter mitral valve repair.
PubMed 文献 · Radiol Med · 2025-01-15

研究将10例患者的多期相心脏CT血管造影数据载入原型二尖瓣评估工具(Siemens)与商用工具(CVI42),由3名专家读者分别用两种工具在收缩期和舒张期图像上测量二尖瓣环。两种工具观察者间一致性良好至优秀(ICC 0.60–1.00),但瓣环面积参数存在系统性差异,作者认为临床与科研流程中应考虑使用同一软件。

要点

  • 两种软件观察者间一致性良好至优秀(ICC 0.60–1.00)。
  • 瓣环面积参数在鞍形与D形瓣环测量中均存在系统性差异。
  • 临床与科研流程中应考虑使用同一软件进行测量。

核心结论

  • 两种工具均可有效评估二尖瓣环,观察者间一致性良好至优秀;由于瓣环面积和距离存在系统性差异,临床与科研流程中应考虑使用同一软件。

研究设计与样本

  • 研究纳入10例患者(平均年龄61.9±9.9岁,70%为男性),其中二尖瓣正常5例、原发性二尖瓣反流5例;将多期相心脏CT血管造影数据分别载入原型二尖瓣评估工具(Siemens)和商用工具(CVI42),由3名专家读者独立在收缩期和舒张期图像上测量瓣环。

诊断标准与关键数值

  • 采用t检验评估两种工具间测量一致性,p≤0.05视为有统计学显著差异。
  • 观察者间一致性采用组内相关系数(ICC)评估,良好为ICC 0.60–0.74,优秀为ICC 0.75–1.00。
  • 采用Bland-Altman图评估系统性差异。

临床应用与阅片要点

  • 作者认为,由于瓣环面积和距离存在系统性差异,临床与科研流程中应考虑使用同一软件。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: Pre-procedural imaging is critical for transcatheter mitral valve repair planning in patients with mitral valve disease. As differences among various measurement techniques for valve evaluation are still poorly understood, we sought to assess the intra- and interobserver agreement of complex measurements derived from a prototype mitral evaluation tool (Siemens) and a commercially available tool (CVI42) using both saddle- and D-shaped mitral annulus techniques.

MATERIALS AND METHODS: Multiphasic cardiac computed tomography angiography data were loaded into each software. Three expert readers independently measured the annuli on systolic- and diastolic-phase images using both tools. Measurement agreement between the tools was assessed with t tests, with p ≤ 0.05 considered statistically significant. Intraclass correlation coefficient (ICC) was used for interobserver agreement. Bland-Altman plots were used to assess for systematic differences.

RESULTS: Ten patients (mean age: 61.9 ± 9.9 years, 70%males) were included, with either normal mitral valve (n = 5) or with primary mitral regurgitation (n = 5). The intraobserver comparison showed statistically significantly different circumference and planar surface areas only for one reader in each group. However, there was significant difference (p ≤ 0.05) between measurements for intercommissural distance, septolateral distance, and intertrigone distance. The interobserver agreement was good (ICC, 0.60-0.74) to excellent (ICC, 0.75-1.00). There were systematic differences in mitral annulus area parameters for both saddle- and D-shaped annulus.

CONCLUSION: Both tools effectively assess mitral annulus with good-to-excellent interobserver agreement. As there were systematic differences in the annular area and distance, use of the same software should be considered in clinical and research workflows.