散斑追踪超声心动图中的近场杂乱回波抑制
本研究在有限元仿真和20例正常人心脏超声中评估基于随机奇异值分解(rSVD)的近场杂乱回波滤波对心肌运动估计的影响,并与形态学成分分析(MCA)对比。rSVD滤波后二维位移与真实值高度一致(R2=0.95),活体对比噪声比提升1.02 dB至17.68 dB,所有受试者平均均方根偏差低于5.0%,且rSVD比MCA更实用。
要点
- rSVD近场杂乱回波滤波后,仿真二维位移与真实值高度一致(R2=0.95)。
- 活体对比噪声比提升1.02 dB至17.68 dB,80%病例心尖段提升约4.9 dB。
- rSVD与MCA均能有效滤波,但rSVD显著更实用。
核心结论
- rSVD近场杂乱回波滤波在散斑追踪超声心动图中可靠、准确且高效,证实了以rSVD为代表的矩阵分解方法用于心肌运动估计近场滤波的可行性。
研究设计与样本
- 仿真研究在有限元模型中模拟单轴压缩(轴向应变0.5%至9%,步长0.5%)下均匀介质中材料点及位移,并导入k-Wave工具箱模拟有无真实相位畸变层的高对比度发散波复合超声图像;活体研究采集20例正常人心脏经胸心尖四腔切面编码发散波复合超声心动图,帧率3200帧/秒。
诊断标准与关键数值
- rSVD滤波后仿真二维位移与真实值高度一致,R2=0.95。
- 活体对比噪声比提升范围为1.02 dB至17.68 dB,所有受试者图像质量均持续改善。
- 80%病例心尖段对比噪声比提升约4.9 dB。
- 所有rSVD近场杂乱回波滤波数据的平均均方根偏差低于5.0%。
临床应用与阅片要点
- rSVD与MCA均能有效滤除近场杂乱回波,但rSVD计算效率显著更实用。
- rSVD滤波在散斑追踪前应用于波束合成射频数据,可改善心肌运动估计的图像质量。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Near-field (NF) clutter filters are critical for unveiling true myocardial structure and dynamics. Randomized singular value decomposition (rSVD) stands out for its proven computational efficiency and robustness. This study investigates the effect of rSVD-based NF clutter filtering on myocardial motion estimation.
METHODS: In silico, material points and their displacements in a homogeneous medium under uniaxial compressions (0.5% - 9% axial strains at 0.5% increments) were simulated in finite-element models. They were exported to the k-Wave toolbox for simulations of pre- and post-deformed ultrasound images with/ without a realistic phase aberrating layer in a high-contrast diverging wave compounding scheme. In vivo, echocardiograms of 20 normal human hearts were acquired using a coded diverging wave compounding imaging method at 3200 frames/second in the transthoracic apical four-chamber view. Morphological component analysis (MCA), which is also a sparse representation method but computationally intensive, was used for comparison with rSVD. Both rSVD- and MCA-based filters were applied to beamformed ultrasound radio-frequency (RF) data before cross-correlation-based speckle tracking. Contrast-to-noise ratios (CNRs) and root-mean-square deviations (RMSDs) were computed from regions of interest to evaluate NF clutter filtering performance of rSVD and MCA.
RESULTS: In silico, 2-D displacements estimated from rSVD-based clutter-reduced image data showed strong agreement with ground truth (R2 of 0.95). In vivo, CNR improvements ranged from 1.02 dB to 17.68 dB, consistently enhancing image quality across all subjects. An improvement of ∼4.9 dB in the apical segments was observed in 80% of cases. Mean RMSDs were below 5.0% for all rSVD-based NF clutter-reduced data. While both rSVD and MCA effectively filtered NF clutter, rSVD was significantly more practical.
CONCLUSION: Our findings confirm the reliability, accuracy, and efficiency of rSVD-based clutter filtering in speckle tracking echocardiography. This underscores the feasibility of matrix decomposition-based methods, exemplified by rSVD, in NF clutter filtering for myocardial motion estimation.