评估方向性脑深部电刺激电极导线方向时降低辐射剂量的潜力

Potential for reduction of radiation dose in the assessment of the lead orientation in directional deep brain stimulation electrodes.
PubMed 文献 · Eur J Radiol · 2025-01-10

回顾性分析88例方向性脑深部电刺激术后同时行CBCT和MSCT的患者,CBCT中位有效剂量0.20±0.07 mSv,MSCT为2.11±0.33 mSv(p<0.001)。CBCT优化准直可进一步降低有效剂量76%,MSCT较CBCT有效剂量高90.3%,优化准直后高97.7%。

要点

  • MSCT有效剂量2.11±0.33 mSv,CBCT为0.20±0.07 mSv,CBCT降低90.3%。
  • CBCT视野可在头尾方向准直33.6%、内外侧方向准直60.8%,有效剂量进一步降低76%。
  • 作者提出该方法可能适用于其他神经影像操作,如治疗性脑动脉瘤的3D可视化。

核心结论

  • 评估方向性脑深部电刺激电极导线方向时,CBCT辐射剂量显著低于MSCT,CBCT优化准直可进一步大幅降低剂量。

研究设计与样本

  • 回顾性研究,纳入88例方向性脑深部电刺激术后同时行CBCT(C臂锥形束CT,3D旋转透视)和MSCT(多层螺旋CT)的患者,比较剂量面积乘积和剂量长度乘积并转换为有效剂量。

诊断标准与关键数值

  • CBCT中位有效剂量0.20±0.07 mSv vs MSCT 2.11±0.33 mSv(p<0.001),MSCT较CBCT有效剂量高90.3%(p<0.001)。
  • CBCT视野在头尾方向可准直33.6%、内外侧方向可准直60.8%,优化准直后有效剂量降低76%(p<0.001)。
  • 优化准直后MSCT较CBCT有效剂量高97.7%(p<0.001)。

临床应用与阅片要点

  • 作者提出CBCT优化准直方法可能适用于其他神经影像操作,如治疗性脑动脉瘤的3D可视化。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: Directional deep brain stimulation (dDBS) relies on electrodes steering the stimulation field in a specific direction. Post implantation, however, the intended and real orientation of the lead frequently deviates e.g. due to torque of the electrode. The orientation is assessed by postoperative imaging such as C-arm cone-beam CT (CBCT) using 3D rotational fluoroscopy and multi-slice CT (MSCT). The objective is to determine the effect of different X-ray systems such as CBCT and MSCT, and scan acquisition parameters on effective radiation doses.

MATERIAL AND METHODS: We retrospectively investigated the dose area product (DAP) and dose length product (DLP) of patients who underwent both CBCT and MSCT after dDBS surgery. These metrics were then converted into the effective dose by established conversion coefficients. For CBCT, the feasible dose optimization by collimation was virtually assessed. Univariate analysis was performed to determine differences.

RESULTS: Among 88 patients median effective dose was 0.20 ± 0.07 mSv in CBCT and 2.11 ± 0.33 mSv in MSCT (p < 0.001). The field of view (FOV) in CBCT can be collimated in craniocaudal orientation by 33.6 % and mediolaterally by 60.8 %. This optimized collimation implies an effective dose reduction of 76 % (p < 0.001). Comparison of the effective dose of MSCT vs. CBCT revealed a significant dose reduction by 90.3 % (p < 0.001), whereas optimal collimation by 97.7 % (p < 0.001).

CONCLUSIONS: For assessment of dDBS orientation, CBCT had significantly lower radiation doses compared to MSCT. Optimized collimation in CBCT allows for a further substantial dose reduction. Moreover, the proposed approach is applicable on other neuroimaging procedures such as 3D visualization of treated intracranial aneurysm.