文献速递骨肌MR

新一代0.55T MRI在全髋关节置换术后患者中的应用:与1.5T MRI的比较

New-generation 0.55T MRI in patients with total hip arthroplasty: a comparison with 1.5T MRI.
PubMed 文献 · Clin Radiol · 2024-11-30

前瞻性纳入15例全髋关节置换术后患者,分别行0.55T与1.5T髋部MRI。两位肌骨放射科医师用5点Likert量表定性评估伪影,第三位医师定量测量信号缺失最大径及面积。0.55T伪影较1.5T轻,整体图像质量无显著差异。

要点

  • 0.55T上T1加权序列伪影评分3.7±0.7,1.5T为2.4±1.1,0.55T伪影更轻(p=0.004)。
  • 0.55T上流体敏感序列伪影评分3.1±0.5,1.5T为2.2±1.1,差异未达统计学显著(p=0.18)。
  • 0.55T与1.5T整体图像质量无统计学显著差异(各p≥0.18)。

核心结论

  • 全髋关节置换术后金属伪影在新一代0.55T MRI上主观感知较常规1.5T MRI轻,整体图像质量无差异。

研究设计与样本

  • 前瞻性研究,纳入15例全髋关节置换术后患者(年龄69±14岁),行0.55T与1.5T髋部MRI;两位肌骨放射科医师定性评分,第三位医师定量测量。

诊断标准与关键数值

  • 定性伪影评分采用5点Likert量表(0=严重伪影,5=无伪影)。
  • 0.55T上T1加权序列伪影评分3.7±0.7,1.5T为2.4±1.1(p=0.004)。
  • 0.55T上流体敏感序列伪影评分3.1±0.5,1.5T为2.2±1.1(p=0.18)。
  • 定量测量全髋关节置换所致信号缺失在髋臼、近端、中段及远端柄水平的前后径d(ap)和内外侧径d(ml)及面积,0.55T所有序列均小于1.5T(各p>0.016)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To quantitatively and qualitatively compare the magnitude of metal total hip arthroplasty-induced imaging artifacts in vivo between 1.5T and 0.55T MRI.

MATERIALS AND METHODS: 15 patients (69 ± 14 years) with THA, prospectively underwent 0.55T and 1.5T MRI of the hip. Two fellowship-trained musculoskeletal radiologists qualitatively rated artifact magnitude on T1-weighted and fluid-sensitive sequences using a 5-point Likert scale (0 = severe - 5 = no artifact). Quantitative artifact magnitude was assessed by a third fellowship-trained musculoskeletal radiologist, by measuring the maximum diameter of the THA-induced signal void (mm) anteroposteriorly d(ap) and mediolaterally d(ml) at the level of the acetabulum, proximal, mid and distal stem, including the prosthesis. Additionally, the area of signal void was noted (mm2). Statistical differences between Likert scores were evaluated using the t-test or the Wilcoxon signed-rank test (P-values <.05 = significant).

RESULTS: Qualitative artifact magnitude was on average rated as moderate to small on 0.55T and as large to moderate on 1.5T by both readers. Artifacts were rated less severe on 0.55T compared with 1.5T on T1-weighted (3.7 ± 0.7 vs. 2.4 ± 1.1, p=0.004) and fluid-sensitive sequences (3.1 ± 0.5 vs. 2.2 ± 1.1, p=0.18). Overall image quality did not show any statistically significant differences between 0.55T and 1.5T MRI (each p≥0.18). Metal artifacts' areas and diameters were smaller on 0.55T when compared with 1.5T MRI for all sequences (each p>0.016).

CONCLUSIONS: Total hip arthroplasty-induced metal artifacts are perceived as less severe at new-generation 0.55T when compared with conventional 1.5T MRI with no difference in overall image quality.

来源:Clin Radiol,查看原文。版权归原作者所有。

引用

  1. Breit HC, Obmann M, Schlicht F, et al. New-generation 0.55T MRI in patients with total hip arthroplasty: a comparison with 1.5T MRI. Clin Radiol. 2024. PMID: 39729689. doi:10.1016/j.crad.2024.106758