手持床旁超声定量评估肝脏脂肪:技术实现与成人初步研究
这项前瞻性初步研究纳入18名超重或肥胖成人,用化学位移编码磁共振成像测质子密度脂肪分数(PDFF),并在1个月内由专家或新手操作手持床旁超声采集射频数据分析衰减系数和背向散射系数。两个参数区分PDFF≥5%的肝脏脂肪变性的AUC在专家为0.96-0.97、新手为0.88-0.89,与PDFF显著相关。
要点
- 手持床旁超声定量参数衰减系数和背向散射系数可区分PDFF≥5%的肝脏脂肪变性。
- 专家操作AUC为0.96-0.97,新手(无超声扫描经验)经有限培训后AUC为0.88-0.89。
- 两个参数与PDFF的Spearman相关系数在专家为0.65-0.69、新手为0.58-0.65。
核心结论
- 定量超声可在床旁超声设备上实现,专家或经有限培训的新手操作者在超重或肥胖成人中均获得有前景的初步结果。
研究设计与样本
- 前瞻性、经IRB批准、符合HIPAA的初步研究,纳入18名超重或肥胖成人(平均年龄43岁±14;17名女性),以化学位移编码磁共振成像估计的PDFF为参考标准。
诊断标准与关键数值
- 肝脏脂肪变性定义为PDFF≥5%;衰减系数和背向散射系数区分肝脏脂肪变性的AUC在专家操作者为0.96-0.97、新手操作者为0.88-0.89(所有p<0.01);与PDFF的Spearman相关系数在专家为0.65-0.69、新手为0.58-0.65(所有p<0.02)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To implement, examine the feasibility of, and evaluate the performance of quantitative ultrasound (QUS) with a handheld point-of-care US (POCUS) device for assessing liver fat in adults.
MATERIALS AND METHODS: This prospective IRB-approved, HIPAA-compliant pilot study enrolled adults with overweight or obesity. Participants underwent chemical-shift-encoded magnetic resonance imaging to estimate proton density fat fraction (PDFF) and, within 1 mo, QUS with a POCUS device by expert sonographers and novice operators (no prior US scanning experience). Radiofrequency data from the liver collected with the POCUS device were analyzed offline using probe-specific calibrations to estimate two QUS parameters: attenuation coefficient (AC) and backscatter coefficient (BSC). Area under the receiver operating characteristic curve (AUC) of each parameter was estimated for classifying presence/absence of hepatic steatosis (defined as PDFF ≥ 5%). Spearman rank correlation between each parameter and PDFF was estimated and its significance assessed.
RESULTS: Of 18 participants (mean age, 43 y ± 14; 17 women), 8 had hepatic steatosis (PDFF ≥ 5%). Both AC and BSC classified hepatic steatosis accurately with AUCs of 0.96-0.97 for expert and 0.88-0.89 for novice operators (p < 0.01 for all) and correlated significantly with PDFF with rho's of 0.65-0.69 for expert and 0.58-0.65 for novice operators (p < 0.02 for all).
CONCLUSION: QUS can be implemented on a POCUS device and can be performed by expert or novice operators after limited training in adults with overweight or obesity with promising initial results.
引用
- Tamayo-Murillo D, Weeks JT, Keller CA, et al. Quantitative Liver Fat Assessment by Handheld Point-of-Care Ultrasound: A Technical Implementation and Pilot Study in Adults. Ultrasound Med Biol. 2024. PMID: 39690042. doi:10.1016/j.ultrasmedbio.2024.11.005
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。