机器人辅助远程超声引导介入系统的初步应用
本研究在体模、猪肝模型及临床患者中比较机器人辅助远程超声引导介入系统(RTIS)与常规超声引导(CUG)经皮穿刺的可行性与准确性。结果显示两组穿刺距离误差无显著差异,但RTIS组操作时间显著更长;所有患者均穿刺成功且无并发症,提示RTIS安全可行且准确性相当。
要点
- RTIS与CUG穿刺距离误差在体模和猪肝模型中均无显著差异。
- RTIS组操作时间在体模、猪肝及临床应用中均显著长于CUG组。
- RTIS组6例和CUG组10例患者均穿刺成功且无并发症。
核心结论
- RTIS用于经皮穿刺安全可行,准确性可与常规方法相媲美。
研究设计与样本
- 经动物伦理委员会和人体研究审查委员会批准的前瞻性研究,由两名有经验的介入超声医师在体模、猪肝模型及临床患者中分别使用RTIS和CUG进行经皮穿刺,比较穿刺距离误差和操作时间;临床记录操作时间、成功率和并发症。
诊断标准与关键数值
- 体模研究穿刺距离误差:RTIS组 2.85 ± 2.07 mm vs CUG组 1.79 ± 1.93 mm,p = 0.158。
- 猪肝研究穿刺距离误差:RTIS组 3.28 ± 1.20 mm vs CUG组 2.56 ± 0.98 mm,p = 0.148。
- 体模研究操作时间:RTIS组 50 ± 19 s vs CUG组 19 ± 7 s,p < 0.001。
- 猪肝研究操作时间:RTIS组 106 ± 19 s vs CUG组 61 ± 32 s,p = 0.001。
- 临床应用操作时间:RTIS组 200 ± 27.02 s vs CUG组 104.8 ± 33.92 s,p < 0.001。
临床应用与阅片要点
- RTIS在体模和猪肝模型中穿刺距离误差与CUG无显著差异,但操作时间显著更长。
- 临床应用中RTIS组6例、CUG组10例患者均穿刺成功且无并发症。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Teleultrasound has gained significant traction in clinical practice in recent years. However, studies focusing on remote interventional ultrasound remain limited.
OBJECTIVES: To evaluate the feasibility and accuracy of percutaneous puncture using a robot-assisted teleultrasound-guided interventional system (RTIS).
MATERIALS AND METHODS: This study was approved by the institutional animal ethics committee and human research review board. Written informed consent was obtained from all patients. Two experienced interventional ultrasound physicians performed percutaneous punctures using both RTIS and conventional ultrasound guidance (CUG) in phantom and swine liver models, as well as in clinical settings. Puncture distance errors and operation durations were compared between the RTIS and CUG groups in the experimental models. For clinical applications, operation duration, success rates, and complications were recorded.
RESULTS: No significant differences were observed in puncture distance errors between the RTIS and CUG groups in the phantom study (2.85 ± 2.07 mm vs. 1.79 ± 1.93 mm; p = 0.158) or the swine liver study (3.28 ± 1.20 mm vs. 2.56 ± 0.98 mm; p = 0.148). However, puncture operation durations were significantly longer in the RTIS group compared to the CUG group across all scenarios: phantom study (50 ± 19 s vs. 19 ± 7 s; p < 0.001), swine liver study (106 ± 19 s vs. 61 ± 32 s; p = 0.001), and clinical application (200 ± 27.02 s vs. 104.8 ± 33.92 s; p < 0.001). All six patients in the RTIS group and ten patients in the CUG group successfully underwent percutaneous puncture without complications.
CONCLUSION: The RTIS demonstrated safety and feasibility for percutaneous puncture, providing comparable accuracy to conventional methods.
CLINICAL RELEVANCE STATEMENT: The RTIS offers a safe and effective solution for percutaneous puncture, with the potential to address the scarcity of medical resources in remote and underserved regions.