文献速递儿科影像心血管胸部CT

高螺距螺旋CT对接受心胸病变评估婴儿的辐射暴露与图像质量的真实世界影响

Real-world impact of high-pitch helical CT on radiation exposure and image quality in infants being evaluated for cardiothoracic pathologies.
PubMed 文献 · Pediatr Radiol · 2024-12-04

本回顾性研究纳入2016至2022年间98例婴儿心胸CT检查,比较非门控常规螺旋CT(对照组47例)与非门控高螺距螺旋CT(高螺距组51例)的辐射剂量与图像质量。高螺距组辐射暴露约降低75%(0.54 mSv vs 2.46 mSv,P < .001),主观与客观图像质量相当或更优。

要点

  • 高螺距组辐射剂量约降低75%(0.54 mSv vs 2.46 mSv,P < .001)。
  • 高螺距组主动脉主观显示评分更优(P = .04),其余指标相当或更优。
  • 高螺距组左心房、左心室及主动脉SNR更优,左心房及主动脉CNR更优。

核心结论

  • 婴儿心胸病变评估中采用高螺距螺旋CT方案可降低辐射暴露,同时获得与常规螺旋CT相似或稍好的图像质量。

研究设计与样本

  • 回顾性研究,纳入2016至2022年间98例连续婴儿心脏方案CT检查(男56例、女42例,平均年龄3.3 ± 2.8个月),管电压均为80 kV。

诊断标准与关键数值

  • 对照组47例采用非门控常规螺旋方案在多排探测器CT上采集,高螺距组51例采用非门控高螺距螺旋方案在双源CT上采集。
  • 辐射暴露高螺距组0.54 mSv vs 对照组2.46 mSv,高螺距组约降低75%(P < .001)。
  • 主观分析中高螺距组主动脉显示评分更优(P = .04),其余指标相当或更优。
  • 客观分析中高螺距组左心房SNR(P < .001)、左心室SNR(P = .04)、主动脉SNR(P = .003)更优,左心房CNR(P = .003)和主动脉CNR(P = .009)更优。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Evaluation of cardiothoracic pathologies is a common indication for computed tomography (CT) in infants. However, CT is fraught with challenges specific to the patient population, such as increased sensitivity to radiation and inability to remain stationary during imaging.

OBJECTIVE: This study investigates potential advantages of a high-pitch helical CT protocol for infants with cardiothoracic pathologies. Namely, we evaluate whether a high-pitch helical CT protocol can minimize radiation exposure without compromising image quality.

MATERIALS AND METHODS: This retrospective study included 98 consecutive cardiac protocol CT examinations of infants (56 males, 42 females; mean age 3.3 ± 2.8 months) performed at a tube voltage of 80 kV between 2016 and 2022. Forty-seven examinations were acquired with a non-gated conventional helical protocol on a multi-detector CT scanner (control group) and 51 were acquired with a non-gated high-pitch helical protocol on a dual-source CT scanner (high-pitch (HP) group). Patient characteristics, radiation exposure parameters, and imaging datasets were extracted from the picture archiving and communication system (PACS). Image quality was assessed subjectively by two radiologists who independently assigned ratings, and objectively through attenuation measurements.

RESULTS: Radiation exposure was approximately 75% lower in the HP group (0.54 mSv vs. 2.46 mSv, P < .001). HP examinations demonstrated comparable, or better, image quality across all metrics in both subjective and objective analyses. In the subjective analysis, the HP group achieved superior ratings for visualization of the aorta (P = .04). In the objective analysis, the HP group achieved superior signal-to-noise ratio (SNR) in the left atrium (P < .001), left ventricle (P = .04), and aorta (P = .003), and superior contrast-to-noise ratio (CNR) in the left atrium (P = .003) and aorta (P = .009).

CONCLUSION: Our findings suggest that employing high-pitch helical CT protocols for evaluation of cardiothoracic pathologies in infants decreases radiation exposure while achieving similar to slightly better image quality compared to conventional helical CT protocols.

来源:Pediatr Radiol,查看原文。版权归原作者所有。

引用

  1. Hunter JG, Bera K, Ciancibello L, et al. Real-world impact of high-pitch helical CT on radiation exposure and image quality in infants being evaluated for cardiothoracic pathologies. Pediatr Radiol. 2024. PMID: 39630241. doi:10.1007/s00247-024-06115-z

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。