文献速递心血管MR

非缺血性心肌病5.0 T与3.0 T延迟钆增强对比:交叉设计研究

Late Gadolinium Enhancement of Nonischemic Cardiomyopathy at 5.0 T versus 3.0 T: A Crossover Design Study.
PubMed 文献 · Radiol Cardiothorac Imaging · 2024-01-01

前瞻性交叉研究纳入49例疑似或确诊非缺血性心肌病受试者,随机分两组比较5.0 T与3.0 T心脏MRI的PSIR序列。5.0 T在1.2 mm分辨率下图像质量和LGE显示优于3.0 T,各分辨率下信噪比和对比噪声比更高,但采集时间和LGE质量测量无差异。

要点

  • 5.0 T PSIR在1.2 mm分辨率下图像质量和LGE显示评分均优于3.0 T。
  • 5.0 T各分辨率下信噪比和对比噪声比均高于3.0 T。
  • 5.0 T与3.0 T单层PSIR采集时间和LGE质量测量无差异。

核心结论

  • 5.0 T PSIR成像在非缺血性心肌病患者中比3.0 T提供更好的图像质量和LGE显示,尤其在1.2 mm分辨率下。

研究设计与样本

  • 前瞻性交叉设计研究,2023年4月至2024年3月纳入49例疑似或确诊非缺血性心肌病受试者(平均年龄43.7岁±13.1,39例男性)。
  • 受试者随机分入组1(先5.0 T后3.0 T LGE心脏MRI)或组2(先3.0 T后5.0 T LGE心脏MRI)。
  • PSIR图像采集空间分辨率分别为1.2、0.9和1.6 mm。
  • 由两名阅片者用五点Likert量表定性评估图像质量和LGE,并定量评估信噪比、对比噪声比和LGE质量;用Bland-Altman图评估阅片者间一致性。

关键影像表现/征象

  • 5.0 T与3.0 T获取单层PSIR图像的采集时间无差异证据(所有P > .05),不受1.2、0.9和1.6 mm分辨率影响。
  • 1.2 mm分辨率下5.0 T PSIR图像质量评分中位数5(IQR 5-5)优于3.0 T中位数5(IQR 4-5),P = .004。
  • 1.2 mm分辨率下5.0 T PSIR LGE评分中位数5(IQR 5-5)优于3.0 T中位数4.25(IQR 4-5),P < .001。
  • 1.6 mm分辨率下5.0 T与3.0 T心脏MRI的图像质量和LGE评分无差异证据。
  • 5.0 T PSIR各分辨率下信噪比和对比噪声比均高于3.0 T(所有P < .001),但LGE质量测量无差异证据。

诊断标准与关键数值

  • 图像质量和LGE采用五点Likert量表定性评估,由两名阅片者完成。
  • 1.2 mm分辨率下5.0 T图像质量中位数5(IQR 5-5)vs 3.0 T中位数5(IQR 4-5),P = .004。
  • 1.2 mm分辨率下5.0 T LGE评分中位数5(IQR 5-5)vs 3.0 T中位数4.25(IQR 4-5),P < .001。
  • 5.0 T各分辨率信噪比和对比噪声比均高于3.0 T(所有P < .001)。

临床应用与阅片要点

  • 在非缺血性心肌病LGE成像中,5.0 T PSIR在1.2 mm分辨率下可提供优于3.0 T的图像质量和LGE显示。
  • 5.0 T与3.0 T的单层PSIR采集时间无差异证据,LGE质量测量亦无差异证据。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Purpose To compare the acquisition time, image quality, and late gadolinium enhancement (LGE) visualization and quantification on phase-sensitive inversion recovery (PSIR) images using 5.0-T versus 3.0-T cardiac MRI. Materials and Methods In this prospective crossover study, 49 participants (mean ± SD age, 43.7 years ± 13.1; 39 men) suspected or diagnosed with nonischemic cardiomyopathy were enrolled from April 2023 to March 2024 and randomly assigned to group 1 (5.0-T followed by 3.0-T LGE cardiac MRI) or group 2 (3.0-T followed by 5.0-T LGE cardiac MRI). PSIR images were acquired at spatial resolutions of 1.2, 0.9, and 1.6 mm. Image quality and LGE were qualitatively evaluated using a five-point Likert scale by two readers, and signal-to-noise ratio, contrast-to-noise ratio, and LGE mass were quantitatively assessed. Bland-Altman plots were used to evaluate interreader agreement. Results There was no evidence of a difference in the acquisition time for obtaining a single-layer PSIR image at 5.0 T compared with 3.0 T (P > .05 for all), irrespective of resolutions at 1.2, 0.9, and 1.6 mm. The 5.0-T PSIR images demonstrated better image quality and LGE visualization compared with 3.0-T images, particularly at 1.2 mm (image quality: median 5 [IQR, 5-5] vs median 5 [IQR, 4-5]; P = .004; LGE score: median 5 [IQR, 5-5] vs median 4.25 [IQR, 4-5]; P < .001). No evidence of differences in image quality or LGE scores was found between 5.0-T and 3.0-T cardiac MRI at 1.6-mm resolution. Signal-to-noise ratio and contrast-to-noise ratio were higher on 5.0-T PSIR images across all resolutions compared with 3.0-T images (P < .001 for all), but no evidence of a difference was found in LGE mass measurements. Conclusion The study demonstrates that 5.0-T PSIR imaging offers better image quality and LGE visualization than 3.0-T PSIR, particularly at a 1.2-mm resolution, in individuals with nonischemic cardiomyopathy. Keywords: MRI, Cardiac, Heart, Comparative Studies, Nonischemic Cardiomyopathy, Late Gadolinium Enhancement, Phase-Sensitive Inversion Recovery Supplemental material is available for this article. ©RSNA, 2024.

来源:Radiol Cardiothorac Imaging,查看原文。版权归原作者所有。

引用

  1. Qian X, Wang S, Wu Y, et al. Late Gadolinium Enhancement of Nonischemic Cardiomyopathy at 5.0 T versus 3.0 T: A Crossover Design Study. Radiol Cardiothorac Imaging. 2024. PMID: 39699303. doi:10.1148/ryct.240035

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。