成人与儿童体模中双能CT与单能CT在病灶对比噪声比等效时的辐射剂量比较

Comparison of radiation exposure from dual- and single-energy CT imaging protocols resulting in equivalent contrast-to-noise ratio of lesions for adults and children: a phantom study.
PubMed 文献 · Eur Radiol · 2024-12-18

研究用成人和10岁儿童体模,比较快速kV切换双能CT与单能CT在病灶对比噪声比等效时的辐射剂量。成人胸部DECT器官剂量较SECT低5-11%,腹部低44-45%;儿童胸部DECT剂量较SECT高2.8-3.4倍,腹部高1.5-1.6倍。作者认为成人DECT剂量相似或更低,儿童则明显更高。

要点

  • 成人腹部DECT较SECT器官剂量降低44-45%,胸部降低5-11%。
  • 儿童胸部DECT较SECT器官剂量高2.8-3.4倍,腹部高1.5-1.6倍。
  • 儿童仅在额外信息对有效诊断至关重要时,才优先选择快速kV切换DECT而非SECT。

核心结论

  • 快速kV切换DECT替代SECT在成人中可能伴随相似或更低的剂量负担,但在儿童中剂量负担明显更高。

研究设计与样本

  • 体模研究:使用成人和10岁儿童体模,以不同浓度碘溶液模拟病灶,行胸部及腹部SECT和快速kV切换DECT扫描。
  • 在SECT图像和DECT虚拟单能图像上测量病灶对比噪声比,筛选与DECT最大CNR等效的SECT扫描进行剂量比较。
  • 通过蒙特卡洛模拟测定胸部扫描的肺、乳腺、食管剂量和腹部扫描的肝、肾、胃剂量。

诊断标准与关键数值

  • 以病灶对比噪声比等效作为SECT与DECT病灶识别能力等同的条件。
  • 成人胸部DECT较SECT器官剂量低5-11%,腹部低44-45%。
  • 10岁儿童胸部DECT较SECT器官剂量高2.8-3.4倍,腹部高1.5-1.6倍。

临床应用与阅片要点

  • 作者认为儿童仅在DECT额外信息对有效诊断至关重要时,才优先选择快速kV切换DECT而非SECT。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To compare the radiation exposure from single-energy CT (SECT) against rapid kV-switching dual-energy CT (DECT) imaging in both adults and children when resulting image data offer equivalent lesion identification power.

MATERIALS AND METHODS: Lesions in an adult and a 10-year-old-child body phantom were imitated using iodine solutions of different concentrations. Phantoms were subjected to several SECT and DECT thoracic and abdominal scans using a rapid kV-switching DECT scanner. The contrast-to-noise ratio (CNR) of each lesion was measured on resulting SECT images and virtual monoenergetic images (VMI) available from DECT. The SECT scans that resulted in CNR values similar to the maximum CNR observed in VMIs derived from corresponding DECT scans were identified. SECT and DECT scans with equivalent lesion-discriminating power were compared regarding the associated radiation dose burden. Doses to the lung, breast, and esophagus from thoracic imaging and doses to the liver, kidneys, and stomach from abdominal imaging were determined through Monte Carlo simulations of SECT and DECT exposures.

RESULTS: Compared to SECT imaging of the adult body phantom, organ doses from DECT were found to be 5-11% lower in thoracic imaging and 44-45% lower in abdominal imaging. Compared to SECT imaging of the 10-year-old body phantom, organ doses from DECT were found to be 2.8-3.4 times higher in thoracic imaging and 1.5-1.6 times higher in abdominal imaging.

CONCLUSION: The use of rapid kV-switching DECT instead of SECT imaging may be associated with a similar or lower dose burden in adults but a noticeably higher dose burden in children.

KEY POINTS: Question How does the radiation exposure from single-energy and dual-energy CT imaging compare when both techniques provide equivalent lesion identification power? Findings Rapid kV-switching dual-energy CT compared to single-energy CT may result in a similar or lower radiation dose in adults, but higher radiation dose in children. Clinical relevance Rapid kV-switching dual-energy CT imaging in children should be preferred over single-energy CT imaging only in cases where the additional information provided is crucial for an effective diagnosis.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Ntoufas N, Raissaki M, Damilakis J, et al. Comparison of radiation exposure from dual- and single-energy CT imaging protocols resulting in equivalent contrast-to-noise ratio of lesions for adults and children: a phantom study. Eur Radiol. 2024. PMID: 39694888. doi:10.1007/s00330-024-11273-7

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。