评估剂量监测软件警报对计算机断层扫描辐射剂量降低的影响:一项系统综述
该系统综述检索至2023年12月的文献,纳入9篇使用剂量监测软件检测CT高剂量事件并发出警报的研究,发现高剂量警报最常出现在胸部、头部/脑部及腹盆部CT,主要由超重患者、重复扫描、患者未对准中心等技术因素触发,多数研究显示剂量监测软件有助于减少触发警报的高剂量事件。
要点
- 高剂量警报最常见于胸部CT、头部/脑部CT、胸腹盆联合CT及腹盆部CT,警报百分比分别为1.45%–10.21%、1.54%–4.18%、4.48%–6.60%及4.47%。
- 警报主要由超重患者、重复扫描、患者未对准中心、增加额外CT检查、扫描区域内有骨科硬件及扫描超出目标解剖范围触发。
- 多数纳入研究显示剂量监测软件在减少触发警报的高剂量事件数量方面发挥了作用。
核心结论
- 剂量监测软件能自动识别高剂量CT方案,便于快速调查和剂量优化;高剂量事件由患者因素和技术因素导致,可通过适当监测及对技术资源和人员培训的投入来减轻。
研究设计与样本
- 系统综述,采用PRISMA方法学,检索Medline、Scopus、CINAHL、Embase和Web of Science至2023年12月的文献并评估参考文献列表,纳入9篇使用剂量监测软件检测CT高剂量事件并发出警报的英文文章。
诊断标准与关键数值
- 胸部CT高剂量警报百分比范围为1.45%至10.21%。
- 头部/脑部CT高剂量警报百分比范围为1.54%至4.18%。
- 胸腹盆联合CT高剂量警报百分比范围为4.48%至6.60%。
- 腹盆部CT高剂量警报百分比为4.47%。
临床应用与阅片要点
- 剂量监测软件可自动识别高剂量CT方案并发出警报,便于快速调查和剂量优化。
- 超重患者、重复扫描、患者未对准中心、增加额外CT检查、扫描区域内有骨科硬件及扫描超出目标解剖范围是高剂量警报的主要触发因素。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
INTRODUCTION: Radiation Dose Monitoring Software (DMS) tools have been developed to monitor doses and alert computed tomography (CT) users of high radiation exposure. However, the causal factors for alerts and the impact of DMS in dose optimisation are poorly understood.
AIM: This review aims to identify high-dose CT examinations triggering alerts and their determinants, and to assess if the alerts from DMS help to reduce CT dose levels.
METHODS: To identify relevant articles published to December 2023, an electronic search of Medline, Scopus, CINAHL, Embase, and the Web of Science was undertaken. Reference lists of published articles were also assessed to identify further articles. The Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) methodology was employed to evaluate articles for relevance. Articles were included if they used the DMS tool to detect high-dose events that issued alerts in CT and were published in English.
RESULTS: The search resulted in 83 articles, of which, nine were included after a thorough screening process. High dose alerts most often occurred in CT chest, CT head/brain, CT Chest/abdomen/pelvis, and CT abdomen/pelvis with alert percentages ranging from 1.45% to 10.21%, 1.54% to 4.18%, 4.48% to 6.60%, and 4.47% respectively. Alerts were mainly triggered by overweight patients, scan repetition, miscentering of the patients, extra CT study added, orthopaedic hardware in the scanning area and scanning beyond the desire anatomy. Most of the studies reviewed show that DMS tools played a role in reducing the number of high-dose events that trigger alerts.
CONCLUSION: DMS tools are valuable in automatically identifying high-dose CT protocols, enabling quick investigation and dose optimisation. The high-dose events occurred due to patient and technical factors, which can be mitigated through proper monitoring and investment in both technological resources and staff training.
引用
- Alanazi M, Kench P, Taba S, et al. Evaluating the impact of dose monitoring software alerts on radiation dose reduction in computed tomography: A systematic review. Eur J Radiol. 2024. PMID: 39718305. doi:10.1016/j.ejrad.2024.111892
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。