Nicolab的StrokeViewer在发展中卒中取栓服务体系中的引入与准确性评估:一项服务开发/改进项目

Introduction and accuracy assessment of Nicolab's StrokeViewer in a developing stroke thrombectomy UK service. a service development/improvement project.
PubMed 文献 · Clin Radiol · 2024-11-08

本研究在一个四级卒中中心评估人工智能软件StrokeViewer的落地情况,比较其检测大血管闭塞的准确性与低年资住院医师的表现。两个三个月阶段分别纳入37例和47例,技术失败率从25%降至17%,呈现统计学显著趋势;StrokeViewer敏感度91%和93%、特异度93%和94%,准确率92%和93%,与低年资住院医师的95%和98%相比无统计学显著差异。

要点

  • StrokeViewer检测大血管闭塞的敏感度91%~93%、特异度93%~94%,与低年资住院医师准确率相近。
  • 技术失败率第一阶段25%、第二阶段17%,呈统计学显著趋势。
  • 研究期间通过与厂商沟通和多学科协作改进了技术失败情况。

核心结论

  • StrokeViewer在真实临床环境中检测大血管闭塞的敏感度和特异度接近其宣传值,准确率与低年资住院医师相当,技术失败率经改进后呈下降趋势。

研究设计与样本

  • 服务开发/改进项目,在两个独立的三个月期间收集数据,第一阶段37例、第二阶段47例,比较StrokeViewer与低年资住院医师检测大血管闭塞的准确性,使用SPSS软件进行统计分析。

诊断标准与关键数值

  • StrokeViewer检测大血管闭塞:第一阶段敏感度91%、特异度93%;第二阶段敏感度93%、特异度94%。准确率:StrokeViewer第一阶段92%、第二阶段93%;低年资住院医师第一阶段95%、第二阶段98%,两者差异无统计学显著意义。

临床应用与阅片要点

  • 原文未据此给出推荐。

局限性

  • 第一阶段技术失败率高,第二阶段研究正是为改进技术失败而设计。
  • 原文未报告其他局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

AIM: The aim of this study was to evaluate the implementation of artificial intelligence (AI) software in a quaternary stroke centre as well as assess the accuracy and efficacy of StrokeViewer software in large vessel occlusion detection and its potential impact on radiological workflow.

MATERIALS AND METHODS: Data were collected during two separate three-month periods comparing the accuracy rate of StrokeViewer in detection of large vessel occlusion to that of a junior registrar. During the first three months, 37 cases were identified and during the second leg, 47. The second leg of the study was performed due to a high number of technical failures during the first one and in an attempt to improve those via communication with the manufacturer and co-operation between allied healthcare professionals. Statistical analysis was performed using SPSS software.

RESULTS: Technical failure rate was 25% in the first leg and reduced to 17% in the second leg, showing a trend to statistical significance. Specificity and sensitivity of StrokeViewer were similar in the two legs of the study, measuring 91% and 93% initially and 94% and 93% finally, respectively. Efficacy was comparable to that of the junior registrar with StrokeViewer, demonstrating 92% accuracy during the first leg vs 95% by the junior registrar and 93% in the second leg vs 98% by the junior registrar. These did not show statistical significance.

CONCLUSION: This is a real-life analysis of StrokeViewer efficacy and its potential failures, showing a reduction in failure rate, accuracy rate of a junior registrar, and sensitivity and specificity values close to the advertised ones.

来源:Clin Radiol,查看原文。版权归原作者所有。