文献速递神经头颈CT

超高清光子计数探测器CT用于臂丛神经可视化

Ultra-high-resolution photon-counting detector CT for visualization of the brachial plexus.
PubMed 文献 · Eur J Radiol · 2024-10-30

这项回顾性研究纳入2023年3月至12月行颈部CT的患者,评估光子计数探测器CT(PCD-CT)对臂丛神经显示的效果。研究先比较不同定量常规(Qr)核与量子迭代重建(QIR)强度下的信噪比(SNR),再比较PCD-CT超高清模式(PCD-UHR)、PCD-CT标准模式(PCD-STD)与能量积分探测器CT标准模式(EID-STD)各20例患者的臂丛神经显示。结论为PCD-CT对臂丛神经的描绘优于EID-CT。

要点

  • PCD-UHR的整体图像质量评分4[4-5]、锐利度评分4[3-5],均高于PCD-STD和EID-STD(均P<0.05)。
  • PCD-UHR与PCD-STD的SNR、CNR均高于EID-STD(组间P<0.05),噪声评分低于EID-STD。
  • Qr40与QIR4的SNR分别为3.6±1.1和4.1±1.2,显著高于其他重建条件(P<0.05)。

核心结论

  • PCD-CT对臂丛神经的显示优于EID-CT。

研究设计与样本

  • 回顾性研究,纳入2023年3月至12月行颈部CT的患者共60例(平均年龄63岁±18,男性30例);比较PCD-UHR、PCD-STD与EID-STD各20例患者的臂丛神经显示,并以5分Likert量表进行主观评价、定量评价SD、SNR与CNR。

诊断标准与关键数值

  • Qr40与QIR4的SNR分别为3.6±1.1和4.1±1.2,显著高于其他重建条件(P<0.05)。
  • PCD-UHR、PCD-STD与EID-STD的整体图像质量评分分别为4[4-5]、3[3-4]和2[2-3](均P<0.05)。
  • PCD-UHR、PCD-STD与EID-STD的锐利度评分分别为4[3-5]、3[3-3]和2[1-3](均P<0.05)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To investigate the use of photon-counting detector CT (PCD-CT) to improve brachial plexus depiction.

MATERIALS AND METHODS: This retrospective study included patients who underwent neck CT from March to December 2023. To assess the optimal reconstruction condition in PCD-CT, the signal-to-noise ratios (SNRs) on images using various quantitative regular (Qr) kernels and strengths of quantum iterative reconstruction (QIR) were evaluated. Next, images obtained by ultra-high-resolution mode in PCD-CT (PCD-UHR), standard mode in PCD-CT (PCD-STD), and standard mode in energy-integrating detector CT (EID-STD) of 20 patients each were compared regarding brachial plexus depiction. A qualitative evaluation was performed using a 5-point Likert scale regarding sharpness, noise, and overall image quality. The standard deviations (SDs), SNRs, and contrast-to noise ratios (CNRs) were quantitatively evaluated.

RESULTS: Overall, 60 patients (mean age, 63 years ± 18; 30 males) were included. The SNRs for the Qr40 and QIR4 (means ± SDs) were 3.6 ± 1.1 and 4.1 ± 1.2, respectively, significantly higher than others (P < 0.05). The scores for overall image quality were 4 [4-5], 3 [3-4], and 2 [2-3], and those for sharpness were 4 [3-5], 3 [3-3], and 2 [1-3] for PCD-UHR, PCD-STD, and EID-STD, respectively (all, P < 0.05). Those for noise were 3 [3-4], 4 [3-4], and 2 [2-2], the SDs were 6.6 ± 1.6, 5.4 ± 0.8, and 8.8 ± 1.7, SNRs were 5.0 ± 1.4, 6,1 ± 1.2, and 3.5 ± 1.6, and CNRs were 5.6 ± 1.9, 7.9 ± 1.7, and 4.4 ± 1.8, respectively (between either of the PCD groups and EID-STD, P < 0.05).

CONCLUSION: PCD-CT showed superior delineation for brachial plexus to EID-CT.