文献速递儿科影像心血管CT

前瞻性心电触发CT在先天性心脏病儿童心脏功能评估中的应用:心率依赖性单扩展扫描与心率无关性双焦点扫描的辐射剂量与图像质量比较

Cardiac functional assessment using prospectively electrocardiography-triggered computed tomography in children with congenital heart disease: Comparison of radiation dose and image quality between heart rate-dependent single-extended and heart rate-independent dual-focused scans.
PubMed 文献 · Eur J Radiol · 2024-11-19

本研究纳入1252例前瞻性心电触发儿童心胸CT检查,采用倾向性评分匹配比较单焦点、单扩展与双焦点扫描的辐射剂量与图像质量。单扩展与双焦点扫描的有效剂量分别约为年龄性别匹配单焦点扫描的2.0-3.2倍(取决于心率)和约1.8倍(与心率无关)。双焦点扫描图像噪声与信噪比与单焦点扫描相似,但在较大年龄匹配组中对比噪声比略低。

要点

  • 双焦点前瞻性心电触发扫描有效剂量约1.1±0.4 mSv,约为匹配单焦点扫描的1.8倍。
  • 单扩展扫描有效剂量约1.4±0.5 mSv,约为匹配单焦点扫描的2.0-3.2倍(取决于心率)。
  • 双焦点与单焦点扫描图像噪声和信噪比相似,但较大年龄组双焦点对比噪声比略低。

核心结论

  • 对于先天性心脏病儿童的心脏功能评估,与心率依赖性单扩展前瞻性心电触发扫描相比,心率无关性双焦点前瞻性心电触发扫描可降低CT辐射剂量(尤其在较低心率时),且图像质量相当。

研究设计与样本

  • 研究纳入1252例前瞻性心电触发儿童心胸CT检查,包括单焦点(匹配比较中的参照)、单扩展(较年轻患者)和双焦点(较年长患者)扫描;采用倾向性评分匹配减少年龄和性别混杂,形成两个匹配组(较年轻匹配组:单焦点 vs. 单扩展;较年长匹配组:单焦点 vs. 双焦点)。

关键影像表现/征象

  • 单扩展扫描图像噪声和信噪比与年龄性别匹配的单焦点扫描相似(p值>0.05)。
  • 双焦点扫描图像噪声和信噪比与年龄性别匹配的单焦点扫描相似(p值>0.05)。
  • 较年轻匹配组中单焦点与单扩展扫描的对比噪声比相当(p>0.05)。
  • 较年长匹配组中双焦点扫描对比噪声比略低于单焦点扫描(p<0.02)。

诊断标准与关键数值

  • 单扩展扫描有效剂量1.4±0.5 mSv,约为年龄性别匹配单焦点扫描(0.6±0.2 mSv)的2.0-3.2倍(取决于心率)。
  • 双焦点扫描有效剂量1.1±0.4 mSv,约为年龄性别匹配单焦点扫描(0.6±0.2 mSv)的1.8倍(与心率无关)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To evaluate radiation dose (RD) reduction potential of heart rate-independent dual-focused scan of prospectively electrocardiography (ECG)-triggered computed tomography (CT) for cardiac functional assessment in children with congenital heart disease (CHD), RD and image quality of the scan mode were compared to those of heart rate-dependent single-extended scan.

METHODS: This study encompassed 1,252 prospectively ECG-triggered pediatric cardiothoracic CT examinations, including single-focused (a reference in matched comparisons), single-extended (younger patients), and dual-focused (older patients) scans. Propensity score matching was used to reduce the confounding effect of age and sex in two matched groups (MPs) (younger MP: single-focused vs. single-extended; older MP: single-focused vs. dual-focused). CT RD, image noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) of the MPs were compared.

RESULTS: The effective dose of single-extended (1.4 ± 0.5 mSv) and dual-focused (1.1 ± 0.4 mSv) scans were approximately 2.0-3.2 times higher than (depending on heart rate) and approximately 1.8 times (irrespective of heart rate) that of the age- and sex-matched single-focused scans (0.6 ± 0.2 mSv), respectively. Image noise and SNR of single-extended and dual-focused scans were similar to those of the age- and sex-matched single-focused scans (p values > 0.05). The CNR was also comparable between single-focused and single-extended scans (younger MP) (p > 0.05), but a slightly lower CNR of the dual-focused scans compared to single-focused scans was observed in the older MP (p < 0.02).

CONCLUSION: For cardiac functional assessment in children with CHD, heart rate-independent dual-focused prospectively ECG-triggered scan can reduce CT RD, especially at lower heart rates, with comparable image quality, compared to heart rate-dependent single-extended scan.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Goo HW, Goo SY. Cardiac functional assessment using prospectively electrocardiography-triggered computed tomography in children with congenital heart disease: Comparison of radiation dose and image quality between heart rate-dependent single-extended and heart rate-independent dual-focused scans. Eur J Radiol. 2024. PMID: 39579580. doi:10.1016/j.ejrad.2024.111838