文献速递外周血管心血管CT

60 kVp下深度学习重建实现主动脉CTA超低辐射与对比剂剂量的可行性:图像质量评估

Feasibility of Ultra-low Radiation and Contrast Medium Dosage in Aortic CTA Using Deep Learning Reconstruction at 60 kVp: An Image Quality Assessment.
PubMed 文献 · Acad Radiol · 2024-11-13

前瞻性纳入非肥胖行主动脉CTA受试者,比较100kVp常规组(210 mg碘/kg,混合迭代重建)与60kVp实验组(105 mg碘/kg,分别用混合迭代重建和深度学习重建)。实验组辐射剂量和对比剂剂量分别降低45%和50%,60kVp深度学习重建图像质量主观评分与常规组无差异,且客观指标更优。

要点

  • 非肥胖患者主动脉CTA采用60kVp联合深度学习重建,可在辐射剂量降低45%、对比剂剂量降低50%的同时保持诊断质量。
  • 60kVp深度学习重建的图像SNR和CNR高于60kVp混合迭代重建(均P<0.001),与100kVp常规组相当或更优。
  • 60kVp深度学习重建在各血管节段的主观图像质量与100kVp常规组无差异(均P>0.05),均满足诊断要求。

核心结论

  • 对非肥胖患者,60kVp联合ClearInfinity深度学习重建算法可在保持图像质量的同时,较100kVp联合ClearView混合迭代重建降低主动脉CTA的辐射剂量和对比剂用量。

研究设计与样本

  • 前瞻性研究,非肥胖行主动脉CTA受试者前瞻性分配为两组并重建为三组:100kVp-CV组(100kVp,210 mg碘/kg,ClearView混合迭代重建)、60kVp-CV组(60kVp,105 mg碘/kg,ClearView混合迭代重建)和60kVp-CI组(60kVp,105 mg碘/kg,ClearInfinity深度学习重建);体模研究碘衰减曲线用于指导对比剂方案。

关键影像表现/征象

  • 60kVp-CI组的CT衰减、SNR和CNR优于或等于100kVp-CV组。
  • 与60kVp-CV组相比,60kVp-CI组图像SNR和CNR更高(均P<0.001)。
  • 60kVp-CI组与100kVp-CV组在主动脉各部位的主观图像质量无差异(均P>0.05),且60kVp-CI图像在所有血管节段均被认为诊断充分。

诊断标准与关键数值

  • 实验组辐射剂量1.12 ± 0.23 mSv vs 常规组2.03 ± 0.82 mSv,实验组降低45%。
  • 实验组对比剂剂量19.04 ± 3.03 mL vs 常规组38.11 ± 6.47 mL,实验组降低50%。
  • 60kVp-CI组SNR和CNR高于60kVp-CV组(均P<0.001)。
  • 60kVp-CI组与100kVp-CV组主观图像质量评分无差异(均P>0.05)。

临床应用与阅片要点

  • 对非肥胖患者行主动脉CTA时,可考虑采用60kVp管电压联合深度学习重建算法,以在保持图像质量的同时降低辐射剂量和对比剂用量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To assess the viability of using ultra-low radiation and contrast medium (CM) dosage in aortic computed tomography angiography (CTA) through the application of low tube voltage (60kVp) and a novel deep learning image reconstruction algorithm (ClearInfinity, DLIR-CI).

METHODS: Iodine attenuation curves obtained from a phantom study informed the administration of CM protocols. Non-obese participants undergoing aortic CTA were prospectively allocated into two groups and then obtained three reconstruction groups. The conventional group (100kVp-CV group) underwent imaging at 100kVp and received 210 mg iodine/kg in combination with a hybrid iterative reconstruction algorithm (ClearView, HIR-CV). The experimental group was imaged at 60kVp with 105 mg iodine/kg, while images were reconstructed with HIR-CV (60kVp-CV group) and with DLIR-CI (60kVp-CI group). Student's t-test was used to compare differences in CM protocol and radiation dose. One-way ANOVA compared CT attenuation, image noise, SNR, and CNR among the three reconstruction groups, while the Kruskal-Wallis H test assessed subjective image quality scores. Post hoc analysis was performed with Bonferroni correction for multiple comparisons, and consistency analysis conducted in subjective image quality assessment was measured using Cohen's kappa.

RESULTS: The radiation dose (1.12 ± 0.23mSv vs. 2.03 ± 0.82mSv) and CM dosage (19.04 ± 3.03mL vs. 38.11 ± 6.47mL) provided the reduction of 45% and 50% in the experimental group compared to the conventional group. The CT attenuation, SNR, and CNR of 60kVp-CI were superior to or equal to those of 100kVp-CV. Compared to the 60kVp-CV group, images in 60kVp-CI showed higher SNR and CNR (all P < 0.001). There was no difference between the 60kVp-CI and 100kVp-CV group in terms of the subjective image quality of the aorta in various locations (all P > 0.05), with 60kVp-CI images were deemed diagnostically sufficient across all vascular segments.

CONCLUSION: For non-obese patients, the combined use of 60kVp and DLIR-CI algorithm can be preserving image quality while enabling radiation dose and contrast medium savings for aortic CTA compared to 100kVp using HIR-CV.

来源:Acad Radiol,查看原文。版权归原作者所有。