经颅MRI引导的组织粉碎术靶向:利用MR测温与MR声辐射力成像

Transcranial MRI-guided Histotripsy Targeting Using MR-thermometry and MR-ARFI.
PubMed 文献 · Ultrasound Med Biol · 2024-11-25

在体外牛脑经人颅骨条件下,评估MR测温与MR声辐射力成像(MR-ARFI)为组织粉碎术做治疗前靶向的可行性;12例牛脑经35 MPa、75 MPa及经颅36 MPa组织粉碎术,两种方法均能估计病灶位置,经颅时靶向精度未受明显影响。

要点

  • MR-ARFI与MR测温均可用于组织粉碎术治疗前靶向定位。
  • 经颅条件下焦点平均绝对误差:MR-ARFI横向/纵向1.10/2.91 mm,MR测温1.29/2.91 mm。
  • 颅骨降低焦点压力,但未损害两种方法的靶向精度。

核心结论

  • 经颅组织粉碎术利用MR测温与MR-ARFI进行治疗前靶向是可行的。

研究设计与样本

  • 体外研究:12例离体牛脑,经人颅骨,使用700 kHz、128阵元MR兼容组织粉碎术阵列,以35 MPa、75 MPa及经颅36 MPa行组织粉碎术,比较病灶位置与MR测温、MR-ARFI估计的焦点。

关键影像表现/征象

  • MR-ARFI与MR测温成功完成治疗前靶向,病灶位置与估计焦点相符。
  • 无颅骨时MR-ARFI焦点估计平均绝对误差横向/纵向为2.06/2.95 mm,MR测温为2.13/2.51 mm。
  • 经颅时MR-ARFI横向/纵向平均绝对误差为1.10/2.91 mm,MR测温为1.29/2.91 mm。
  • 人颅骨降低焦点区域压力,但未损害两种方法的靶向精度。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Transcranial magnetic resonance imaging (MRI)-guided histotripsy has been demonstrated to treat various locations in in vivo swine brain through a human skull. To ensure that the histotripsy treatment is delivered to the intended target location, accurate pre-treatment targeting is necessary. In this work, we investigate the feasibility of MR-thermometry and MR-acoustic radiation force imaging (MR-ARFI) to perform pre-treatment targeting of histotripsy in ex vivo bovine brain through a human skull.

METHODS: A 700 kHz, 128-element MR-compatible histotripsy array was used to generate histotripsy and tone-burst sonications. The array's electronic drivers were modified to also generate low-amplitude tone-burst sonications to perform MR-thermometry and MR-ARFI-based targeting. Twelve ex vivo bovine brains were treated with histotripsy at 35 MPa, 75 MPa and through a skull at 36 MPa. Before treating the tissue, both MR-ARFI and MR-thermometry were used to estimate the lesion location. Finally, the location of the histotripsy lesion was compared with the focus estimated by MR-thermometry and MR-ARFI.

RESULTS: MR-thermometry and MR-ARFI were able to successfully perform pre-treatment targeting of histotripsy using the modified histotripsy array driver. Histotripsy focus was estimated with mean absolute errors along the transverse/longitudinal axis of 2.06/2.95 mm and 2.13/2.51 mm for MR-ARFI and MR-thermometry, respectively. The presence of the human skull reduced the pressure at the focal region, but it did not compromise the targeting accuracy of either of the two methods with a mean absolute error of 1.10/2.91 mm and 1.29/2.91 mm for MR-ARFI and MR-thermometry, respectively.

CONCLUSION: This study demonstrated that transcranial histotripsy pre-treatment targeting is feasible with MR-thermometry and MR-ARFI.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。