小型3.0-T新生儿MRI系统的声学噪声评估

Acoustic noise in a small-format 3.0-T neonatal MRI system.
PubMed 文献 · Pediatr Radiol · 2024-10-15

本研究比较了小型3.0-T新生儿MRI扫描仪与常规成人尺寸1.5-T和3.0-T MRI扫描仪在相同新生儿头部成像协议下的声压级。结果显示,NICU专用3.0-T扫描仪在等中心处较1.5-T和3.0-T分别低5.98 dBA(P=0.007)和13.89 dBA(P<0.001),在扫描仪旁工作人员处分别低20.24 dBA和19.28 dBA(均P<0.001)。结论认为该系统显著更安静,提高了新生儿安全性。

要点

  • NICU专用3.0-T MRI在等中心处较1.5-T低5.98 dBA、较3.0-T低13.89 dBA(P<0.001)。
  • 扫描仪旁工作人员处噪声较1.5-T低20.24 dBA、较3.0-T低19.28 dBA(均P<0.001)。
  • 小型3.0-T系统噪声显著低于常规成人尺寸MRI,有利于新生儿安全。

核心结论

  • NICU专用3.0-T MRI系统显著比常规成人尺寸MRI系统安静,改善了新生儿患者的安全性;在扫描过程中临床医生可能所在的屏蔽室内也观察到声压级显著降低。

研究设计与样本

  • 研究论文;在小型3.0-T新生儿MRI扫描仪及成人尺寸1.5-T和3.0-T扫描仪上,采用相同的新生儿头部标准成像协议进行声压级测量;使用Brüel & Kjær 2250型声级计;采用单因素方差分析比较扫描仪间声压级差异的统计学显著性。

诊断标准与关键数值

  • 平均A加权声压级:1.5-T MRI为81.02 ± 0.28 dBA,3.0-T MRI为87.00 ± 0.85 dBA,NICU MRI为94.91 ± 0.65 dBA;NICU MRI等中心处声压级较1.5-T低5.98 dBA(P=0.007),较3.0-T低13.89 dBA(P<0.001);扫描仪旁工作人员处NICU MRI较1.5-T低20.24 dBA(P<0.001),较3.0-T低19.28 dBA(P<0.001)。

临床应用与阅片要点

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Today, magnetic resonance imaging (MRI) is rarely used in managing the care of premature neonates. This is in large part due to the medical and logistical challenges associated with moving neonates from the neonatal intensive care unit (NICU) to the radiology department. Furthermore, acoustic noise associated with MR scanning poses safety concerns for both practitioners and neonatal patients. A small-format 3.0-T neonatal scanner was recently developed and placed within the NICU to address these logistical and acoustic challenges.

OBJECTIVE: To compare acoustic noise measurements of a small-format 3.0-T neonatal MRI scanner with conventional adult-sized 1.5-T and 3.0-T MRI scanners using identical neonatal head imaging protocols.

MATERIALS AND METHODS: Sound pressure level (SPL) measurements of a standard imaging protocol were made in a small-format neonatal 3.0-T MRI scanner as well as in adult-sized 1.5-T and 3.0-T scanners. SPL measurements were made with a Brüel & Kjær sound level meter model 2250. The statistical significance of the differences in SPL between scanners was determined using one-way ANOVA.

RESULTS: Average sound pressure level values were measured in unweighted decibels (dB) and A-weighted decibels (dBA) for all imaging sequences in the protocol. The average A-weighted SPLs for the NICU from 1.5-T and 3.0-T MRI scanners were 81.02 ± 0.28 dBA, 87.00 ± 0.85 dBA, and 94.91 ± 0.65 dBA, respectively. SPLs at the isocenter of the NICU MRI scanner were 5.98 dBA quieter than in the 1.5-T scanner (P=0.007), and 13.89 dBA quieter than in the 3.0-T scanner (P<0.001). For staff standing next to the scanner, the NICU scanner was 20.24 dBA quieter than the 1.5-T scanner (P<0.001) and 19.28 dBA quieter than the 3.0-T scanner (P<0.001).

CONCLUSION: The NICU 3.0-T MRI system is significantly quieter than conventional adult-sized MRI systems, improving safety for neonatal patients. Significant reductions in SPL were also noted inside the screen room where clinicians may be present during scanning.