「T2环征」相关参数对高强度聚焦超声消融子宫肌瘤的影响
回顾性分析616例子宫肌瘤患者HIFU消融数据,T2环征的存在是影响消融效果的独立危险因素。207例有T2环征的肌瘤中,环征覆盖比与治疗效率及无灌注体积比(NPVR)负相关,平均厚度与治疗强度正相关、与NPVR负相关;等信号肌瘤中平均厚度与NPVR负相关更明显(r=-0.484)。
要点
- T2环征存在是HIFU消融效果的独立危险因素。
- T2环征覆盖比越高,治疗效率与NPVR越低,能量效率因子越高。
- T2环征平均厚度越大,治疗强度越高、NPVR越低。
核心结论
- T2环征的存在增加消融难度并降低消融效果;临床筛选HIFU适应证及制定治疗方案时应充分考虑T2环征的存在及相关参数。
研究设计与样本
- 回顾性分析616例子宫肌瘤患者HIFU消融数据,采用单因素及多因素logistic回归分析消融效果影响因素,并对207例有T2环征的肌瘤行Spearman相关分析。
关键影像表现/征象
- T2环征表现为T2加权MR图像上病灶周边高信号环。
- T2环征存在是影响HIFU消融效果的独立危险因素。
诊断标准与关键数值
- T2环征覆盖比与治疗效率负相关(r=-0.174,p=0.012),与NPVR负相关(r=-0.186,p=0.007),与能量效率因子(EEF)正相关(r=0.156,p=0.024)。
- T2环征平均厚度与治疗强度正相关(r=0.203,p=0.003),与NPVR负相关(r=-0.363,p<0.001)。
- 等信号肌瘤中T2环征平均厚度与NPVR负相关(r=-0.484,p<0.001)。
临床应用与阅片要点
- 作者建议临床筛选HIFU适应证及制定治疗方案时应充分考虑T2环征的存在及相关参数。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To investigate the effect of "high-signal-intensity peripheral rim on T2-weighted MR images (T2-rim sign)" related parameters on non-perfused volume ratio (NPVR) after high-intensity focused ultrasound (HIFU) ablation of uterine fibroids.
METHODS: Data from 616 patients with uterine fibroids treated with HIFU were retrospectively analyzed. Univariate and multivariate logistic regression was used to analyze the factors influencing the ablation effect. The effect of T2-rim sign on ablation parameters and results was also analyzed. Spearman correlation analysis was used to compare the correlation between coverage ratio, average thickness of T2-rim sign and NPVR in 207 cases of fibroids with T2-rim sign.
RESULTS: The presence of T2-rim sign was an independent risk factor affecting the ablation effect. The coverage ratio of T2-rim sign was negatively correlated with treatment efficiency (r = -0.174, p = 0.012) and NPVR (r = -0.186, p = 0.007), and positively correlated with energy efficiency factor (EEF) (r = 0.156, p = 0.024). The average thickness of T2-rim sign was positively correlated with treatment intensity (r = 0.203, p = 0.003) and negatively correlated with NPVR (r = -0.363, p < 0.001). There was a negative correlation between the average thickness of the T2-rim sign and NPVR in isointense fibroids (r = -0.484, p < 0.001).
CONCLUSION: The presence of T2-rim sign increases the difficulty of ablation and reduces the ablation effect. In clinical practice, the presence and related parameters of T2-rim sign should be fully considered when screening for HIFU indications and formulating treatment plans.
引用
- Wang L, Liu Y, Lin J, et al. Effect of "T2-rim sign" related parameters on high-intensity focused ultrasound ablation of uterine fibroids. Eur J Radiol. 2024. PMID: 39357287. doi:10.1016/j.ejrad.2024.111767
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。