文献速递技术与安全US

脉冲回波超声定量测量两种不相关弹性参数

Pulse-Echo Ultrasound for Quantitative Measurements of Two Uncorrelated Elastic Parameters.
PubMed 文献 · Ultrasound Med Biol · 2024-10-05

本文描述弹性组织特性与应变的关系,并在组织模拟体模中初步研究脉冲回波超声测量两种不相关弹性参数(剪切模量与非线性压缩性)。采用4.5 MHz超声辐射力诱导剪切波、12.5 MHz平面波成像跟踪剪切波,以及0.7 MHz与14 MHz双频脉冲复合测量非线性压缩性。超声弹性结果与机械压痕及文献结果吻合良好,作者计划开展甲状腺病变和肝脂肪变性患者研究。

要点

  • 脉冲回波超声可在体模中定量测量剪切模量和非线性压缩性两种不相关弹性参数。
  • 剪切模量超声测量与机械压痕验证结果吻合良好。
  • 非线性压缩性超声测量与文献推导结果吻合良好。

核心结论

  • 本初步研究结果 promising,超声测量的弹性参数与机械压痕及文献结果吻合良好,作者计划开展甲状腺病变和肝脂肪变性患者研究以探索形状变形和体积变形相关弹性参数的临床实用性。

研究设计与样本

  • 体模初步研究,制备含可变剪切模量和可变非线性压缩性病灶的组织模拟体模;使用4.5 MHz超声辐射力诱导剪切波、12.5 MHz平面波成像跟踪剪切波,0.7 MHz与14 MHz双频脉冲复合测量非线性压缩性;算法在Verasonics Vantage超声扫描仪上实现,使用定制多频超声换能器;以机械压痕测量验证剪切模量,非线性压缩性与文献推导结果比较。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: This paper describes the relationship between elastic tissue properties and strain and presents an initial investigation of pulse-echo ultrasound to measure two uncorrelated elastic parameters in tissue-mimicking phantoms. The two elastic parameters are the shear modulus, related to deformation of shape, and what we in the paper define as the nonlinear compressibility, related to deformation of volume.

METHODS: We prepared tissue-mimicking phantoms containing lesions of variable shear modulus and variable nonlinear compressibility. An in-house framework for shear wave imaging was developed using ultrasound radiation force at 4.5 MHz to induce shear waves and plane wave imaging with pulses in a frequency band centered around 12.5 MHz to track the shear waves. For measurements of nonlinear compressibility, co-propagating dual-frequency pulse complexes at 0.7 MHz and 14 MHz were applied. Algorithms were implemented on a Verasonics Vantage ultrasound scanner and a custom-made multi-frequency ultrasound transducer was used. Mechanical indentation measurements were performed to validate ultrasound measurements of the shear modulus. For the nonlinear compressibility, ultrasound measurements were compared to results derived from the literature.

RESULTS: We found good agreement in elasticity results from ultrasound measurements and mechanical indentation as well as when comparing with results derived from the literature.

CONCLUSION: Results of the current investigation were promising. We plan patient studies involving thyroid lesions and liver steatosis to explore whether measurements of elastic parameters related both to shape deformation and volume deformation are useful in clinical practice.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Hansen R, Ervik Ø, Amini N, et al. Pulse-Echo Ultrasound for Quantitative Measurements of Two Uncorrelated Elastic Parameters. Ultrasound Med Biol. 2024. PMID: 39370356. doi:10.1016/j.ultrasmedbio.2024.08.006

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。