基于超高清光子计数探测器CT的冠状动脉CT血管造影衍生血流储备分数:与能量积分探测器CT的个体内比较
前瞻性纳入34例在EID-CT上检出钙化斑块的患者,30天内行PCD-CT超高清CCTA,用机器学习算法计算CT-FFR。PCD-CT与EID-CT的CT-FFR一致性优秀(血管水平ICC 0.93,患者水平ICC 0.94),但PCD-CT测值更高(血管水平0.58±0.23 vs 0.55±0.23,P<0.001;患者水平0.73±0.23 vs 0.70±0.22,P<0.001),2例(5.9%)在EID-CT为血流动力学显著者被PCD-CT重新分类为非显著。
要点
- PCD-CT与EID-CT的CT-FFR一致性优秀:血管水平ICC 0.93(0.90-0.95),患者水平ICC 0.94(0.88-0.97)。
- PCD-CT测得的CT-FFR值高于EID-CT:血管水平0.58±0.23 vs 0.55±0.23(P<0.001),患者水平0.73±0.23 vs 0.70±0.22(P<0.001)。
- 2例(5.9%)EID-CT判为血流动力学显著(CT-FFR≤0.80)的病变在PCD-CT上被重新分类为非显著,其余患者分类一致。
核心结论
- UHR PCD-CT的CT-FFR与EID-CT测量一致性优秀,但PCD-CT产生更高的CT-FFR值,可能导致血流动力学显著性的重新分类。
研究设计与样本
- 前瞻性研究,纳入2023年4月1日至2024年1月31日在EID-CT上检出钙化斑块的患者34例(年龄67.3±6.6岁,女性7例[20.6%]),30天内行PCD-CT超高清CCTA,采用相同或更低的CT剂量指数和等量对比剂,使用现场机器学习算法计算CT-FFR,以CT-FFR≤0.80为血流动力学显著。
诊断标准与关键数值
- CT-FFR≤0.80判定为血流动力学显著;血管水平ICC 0.93(0.90-0.95),患者水平ICC 0.94(0.88-0.97);PCD-CT血管水平CT-FFR 0.58±0.23 vs EID-CT 0.55±0.23(P<0.001);PCD-CT患者水平0.73±0.23 vs EID-CT 0.70±0.22(P<0.001);2例(5.9%)EID-CT显著病变在PCD-CT重新分类为非显著。
临床应用与阅片要点
- 在解读PCD-CT衍生的CT-FFR时,需注意其测值可能高于EID-CT,以CT-FFR≤0.80为界值时存在将EID-CT判定的显著病变重新分类为非显著的可能。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To evaluate the feasibility of CT angiography-derived fractional flow reserve (CT-FFR) calculations on ultrahigh-resolution (UHR) photon-counting detector (PCD)-CT series and to intra-individually compare the results with energy-integrating (EID)-CT measurements.
METHOD: Prospective patients with calcified plaques detected on EID-CT between April 1st, 2023 and January 31st, 2024 were recruited for a UHR CCTA on PCD-CT within 30 days. PCD-CT was performed using the same or a lower CT dose index and an equivalent volume of contrast media. An on-site machine learning algorithm was used to obtain CT-FFR values on a per-vessel and per-patient basis. For all analyses, CT-FFR values ≤ 0.80 were deemed to be hemodynamically significant.
RESULTS: A total of 34 patients (age: 67.3 ± 6.6 years, 7 women [20.6 %]) were included. Excellent inter-scanner agreement was noted for CT-FFR values in the per-vessel (ICC: 0.93 [0.90-0.95]) and per-patient (ICC: 0.94 [0.88-0.97]) analysis. PCD-CT-derived CT-FFR values proved to be higher compared to EID-CT values on both vessel (0.58 ± 0.23 vs. 0.55 ± 0.23, p < 0.001) and patient levels (0.73 ± 0.23 vs. 0.70 ± 0.22, p < 0.001). Two patients (5.9 %) with hemodynamically significant lesions on EID-CT were reclassified as non-significant on PCD-CT. All remaining participants were classified into the same category with both scanner systems.
CONCLUSIONS: While UHR CT-FFR values demonstrate excellent agreement with EID-CT measurements, PCD-CT produces higher CT-FFR values that could contribute to a reclassification of hemodynamic significance.
引用
- Vecsey-Nagy M, Tremamunno G, Schoepf UJ, et al. Coronary CT angiography-based FFR with ultrahigh-resolution photon-counting detector CT: Intra-individual comparison to energy-integrating detector CT. Eur J Radiol. 2024. PMID: 39454427. doi:10.1016/j.ejrad.2024.111797
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。