肺结节计算机辅助检测系统在儿童标准剂量与低剂量CT扫描中的性能:个体内比较
本回顾性研究对73例儿童(平均14.7岁)在同次检查中行标准剂量与低剂量胸部CT,用两种基于成人数据训练的肺结节CAD系统检测3-30mm结节(n=247),在每扫描2个假阳性固定频率下比较敏感度;两种系统在低剂量扫描的敏感度均低于标准剂量扫描,对小于5mm结节的检出下降更明显。
要点
- FlyerScan在标准剂量与低剂量CT的结节检出敏感度分别为76.9%与66.8%,低剂量扫描敏感度下降。
- MONAI在标准剂量与低剂量CT的结节检出敏感度分别为67.6%与62.3%,低剂量扫描敏感度下降。
- 低剂量扫描时对小于5mm结节的检出下降总体更明显,使用低剂量CT联合CAD检测儿童小肺结节需谨慎。
核心结论
- 两种基于成人数据训练的肺结节CAD系统在儿童低剂量CT上的敏感度均低于同一患者的标准剂量CT,对小于5mm结节的检出下降总体更明显。
研究设计与样本
- 回顾性研究,纳入2018年11月30日至2020年8月31日期间同次检查行临床标准剂量与研究性低剂量胸部CT的73例患者(32例女性、41例男性;平均年龄14.7岁,年龄范围4-20岁),为先前前瞻性研究的一部分。
诊断标准与关键数值
- 以儿科放射科医师标注的肺结节为参考标准;两种CAD系统(FlyerScan与MONAI)均在每扫描2个假阳性的固定频率下计算3-30mm结节(n=247)的敏感度。
临床应用与阅片要点
- 在儿童低剂量CT方案中联合CAD系统辅助检测小肺结节时需谨慎。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND. When lung nodule computer-aided detection (CAD) systems are applied for pediatric CT, performance may be degraded on low-dose scans due to increased image noise. OBJECTIVE. The purpose of this study was to conduct an intraindividual comparison of the performance for lung nodule detection of two CAD systems trained using adult data between low-dose and standard-dose pediatric chest CT scans. METHODS. This retrospective study included 73 patients (32 female participants, 41 male participants; mean age, 14.7 years; age range, 4-20 years) who underwent both clinical standard-dose and investigational low-dose chest CT examinations during the same encounter from November 30, 2018, to August 31, 2020, as part of an earlier prospective study. Fellowship-trained pediatric radiologists annotated lung nodules to serve as the reference standard. Both CT scans were processed using two publicly available lung nodule CAD systems previously trained using adult data: FlyerScan (github.com/rhardie1/FlyerScanCT) and Medical Open Network for Artificial Intelligence (MONAI; github.com/Project-MONAI/model-zoo/releases). The sensitivities of the two CAD systems for nodules measuring 3-30 mm (n = 247) were calculated when operating at a fixed frequency of two false-positives per scan. RESULTS. FlyerScan exhibited detection sensitivities of 76.9% (190/247; 95% CI, 73.3-80.8%) on standard-dose scans and 66.8% (165/247; 95% CI, 62.6-71.5%) on low-dose scans. MONAI exhibited detection sensitivities of 67.6% (167/247; 95% CI, 61.5-72.1%) on standard-dose scans and 62.3% (154/247; 95% CI, 56.1-66.5%) on low-dose scans. The number of detected nodules for standard-dose versus low-dose scans for 3-mm nodules was 33 versus 24 (FlyerScan) and 16 versus 13 (MONAI), 4-mm nodules was 46 versus 42 (FlyerScan) and 39 versus 30 (MONAI), 5-mm nodules was 38 versus 33 (FlyerScan) and 32 versus 31 (MONAI), and 6-mm nodules was 27 versus 20 (FlyerScan) and 24 versus 24 (MONAI). For nodules measuring 7 mm or larger, detection did not show a consistent pattern between standard-dose and low-dose scans for either system. CONCLUSION. Two lung nodule CAD systems showed decreased sensitivity on low-dose versus standard-dose pediatric CT scans obtained in the same patients. The reduced detection at low dose was overall more pronounced for nodules measuring less than 5 mm. CLINICAL IMPACT. Caution is needed when using low-dose CT protocols in combination with CAD systems to help detect small lung nodules in pediatric patients.
引用
- Hardie RC, Trout AT, Dillman JR, et al. Performance of Lung Nodule Computer-Aided Detection Systems on Standard-Dose and Low-Dose Pediatric CT Scans: An Intraindividual Comparison. AJR Am J Roentgenol. 2024. PMID: 39382534. doi:10.2214/AJR.24.31972
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。