文献速递胸部儿科影像技术与安全CT

儿童光子计数探测器CT对比增强胸部CT:辐射剂量与图像质量对比能量积分探测器CT

Pediatric contrast-enhanced chest CT on a photon-counting detector CT: radiation dose and image quality compared to energy-integrated detector CT.
PubMed 文献 · Pediatr Radiol · 2024-10-28

该回顾性观察研究纳入20对儿童对比增强胸部CT扫描,比较光子计数探测器(PCD)CT与能量积分探测器(EID)CT的辐射剂量和图像质量。PCD CT的容积CT剂量指数和体型特异性剂量估计均显著低于EID CT(P<0.01),肺实质图像噪声更低、信噪比和对比噪声比更优,运动伪影更少,而整体图像质量和小结构显示无显著差异。

要点

  • 儿童对比增强胸部CT中,PCD CT的辐射剂量显著低于EID CT(容积CT剂量指数中位数1.7 mGy vs 3.8 mGy,P<0.01)。
  • PCD CT肺实质图像噪声更低(119.5 HU vs 143.1 HU,P<0.01),信噪比和对比噪声比更优。
  • PCD CT运动伪影评分优于EID CT(P<0.01),整体图像质量和小结构显示无显著差异。

核心结论

  • 与EID CT相比,PCD CT儿童胸部对比增强CT辐射暴露显著降低、图像质量改善且运动伪影减少。

研究设计与样本

  • 经机构审查委员会批准的回顾性观察研究,纳入20对扫描(20例PCD CT;20例EID CT),患儿在12个月内分别接受PCD CT(Siemens NAEOTOM Alpha)和EID CT(Siemens SOMATOM Definition Edge或Force)对比增强胸部CT检查。

关键影像表现/征象

  • PCD CT肺实质图像噪声中位数119.5 HU(IQR 95.4-135.7 HU)低于EID CT的143.1 HU(IQR 125.4-169.8 HU)(P<0.01)。
  • PCD CT信噪比中位数-6.1(IQR -8.4至-4.8)优于EID CT的-4.9(IQR -5.6至-3.8)(P=0.01)。
  • PCD CT对比噪声比中位数-63.9(IQR -84.1至-57.5)优于EID CT的-60.5(IQR -76.3至-52.5)(P=0.01)。
  • PCD CT运动伪影评分较EID CT改善(P<0.01)。
  • 整体图像质量和小结构显示在PCD CT与EID CT间无显著差异(P=0.06和P=0.31)。

诊断标准与关键数值

  • 主观图像质量采用5点Likert量表评估整体质量、小结构显示和运动伪影,由三名儿科放射科医师独立评分,加权kappa分析观察者间一致性。
  • 客观图像质量通过感兴趣区测量CT值(HU)和标准差评估图像噪声、信噪比和对比噪声比,由一名放射科医师在胸部多个位置完成。
  • 辐射暴露参数包括容积CT剂量指数:PCD CT中位数1.7 mGy(IQR 1.1-2.4 mGy)vs EID CT 3.8 mGy(IQR 2.0-4.7 mGy)(P<0.01)。
  • 体型特异性剂量估计:PCD CT中位数2.6 mGy(IQR 1.8-3.1 mGy)vs EID CT 5.0 mGy(IQR 3.3-6.2 mGy)(P<0.01)。

临床应用与阅片要点

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Photon counting detector (PCD) CT benefits from reduced noise compared with conventional energy-integrating detector (EID) CT, which should translate to improved image quality and reduced radiation exposure for pediatric patients undergoing chest CT with IV contrast.

OBJECTIVE: To determine the differences in radiation exposure and image quality of PCD CT and EID CT in pediatric chest CT with intravenous (IV) contrast.

MATERIALS AND METHODS: In this institutional review board-approved retrospective observational study, 20 scan pairs (20 PCD CT; 20 EID CT) for children who underwent chest CT with IV contrast on both a PCD CT (Siemens NAEOTOM Alpha) and an EID CT (Siemens SOMATOM Definition Edge or Force) within 12 months were reviewed independently by three pediatric radiologists for three subjective quality features on 5-point Likert scales: overall quality, small structure delineation, and motion artifact. Objective measures of image quality (image noise, signal-to-noise ratio, and contrast-to-noise ratio) were assessed by a single radiologist in several locations in the chest through region of interest measurement of Hounsfield units (HU) and standard deviation. Patient-related and radiation exposure parameters were collected for each scan and summarized with median and interquartile range (IQR). The Wilcoxon rank-sum test was utilized to compare groups. A P < 0.05 indicated statistical significance. Inter-observer agreement of subjective image quality metrics was analyzed using weighted kappa.

RESULTS: Age (14.2 years vs 13.8 years, P= 0.15), height (P= 0.13), weight (P= 0.21), and BMI (P = 0.24) did not significantly differ between groups. There were 10 male and 3 female patients. Compared to EID CT, PCD CT showed lower radiation exposure parameters including volumetric CT dose index, 1.7 mGy (IQR 1.1-2.4 mGy) vs 3.8 mGy (IQR 2.0-4.7 mGy) (P< 0.01), and size-specific dose estimate, 2.6 mGy (IQR 1.8-3.1 mGy) vs 5.0 mGy (IQR 3.3-6.2 mGy) (P< 0.01). Objective image quality of lung parenchyma was improved on the PCD CT scanner, including image noise 119.5 HU (IQR 95.4-135.7 HU) vs 143.1 HU (IQR 125.4-169.8 HU) (P < 0.01), signal-to-noise ratio (SNR) -6.1 (IQR -8.4 to -4.8) vs -4.9 (IQR -5.6 to -3.8) (P= 0.01), and contrast-to-noise ratio -63.9 (-84.1 to -57.5) vs -60.5 (-76.3 to -52.5) (P = 0.01). Motion artifact was improved on the PCD CT scanner (P< 0.01). No significant differences in overall image quality or small structure delineation were identified (P= 0.06 and P= 0.31).

CONCLUSION: PCD CT pediatric chest CT had significantly reduced radiation exposure, improved image quality, and reduced motion artifact compared with EID CT.

来源:Pediatr Radiol,查看原文。版权归原作者所有。

引用

  1. El-Ali AM, Strubel N, Pinkney L, et al. Pediatric contrast-enhanced chest CT on a photon-counting detector CT: radiation dose and image quality compared to energy-integrated detector CT. Pediatr Radiol. 2024. PMID: 39466387. doi:10.1007/s00247-024-06078-1

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。