围术期应用高频多普勒超声检测来自骨骼的脑脂肪栓子
研究假设猪髓内钉固定和婴儿正中胸骨切开术可导致骨髓释放并产生可检测的脑栓子,采用新型脑多普勒超声监测系统(NeoDoppler)在猪和婴儿中检测高强度瞬态信号(HITS),并以组织学确认脑栓子;结果显示猪股骨手术后5/12(45%)出现HITS而组织学显示11/12(92%)存在骨髓栓子,婴儿胸骨切开术与栓子负荷显著增加相关但HITS强度低于猪,结论认为高频脑多普勒超声可实时检测栓子但敏感度和特异度可能有限且依赖栓子大小。
要点
- 猪股骨手术后仅45%(5/12)监测到HITS,而组织学显示92%(11/12)存在骨髓栓子,提示该方法敏感度有限。
- 婴儿正中胸骨切开术与栓子负荷显著增加相关,但HITS强度低于猪。
- 高频脑多普勒超声可在围术期实时检测栓子,但敏感度和特异度可能有限且依赖栓子大小。
核心结论
- 高频脑多普勒超声是围术期有价值的监测工具,可实时检测栓子,但对骨髓栓子的敏感度和特异度可能有限且依赖栓子大小。
研究设计与样本
- 动物实验:12只无特定病原体挪威长白猪分配至双侧股骨髓内钉固定组或自体骨髓注射阳性对照组,持续多普勒监测并以组织学确认脑栓子;另对8例心脏手术前接受胸骨切开术的婴儿进行NeoDoppler数据比较分析。
关键影像表现/征象
- 猪股骨手术中12只中11只接受脑多普勒监测,5只(45%)出现HITS。
- 猪脑活检显示12只中11只(92%)存在骨髓栓子。
- 4只阳性对照猪接受动脉内注射骨髓、生理盐水或造影剂,全部猪(100%)检测到强HITS。
- 8例婴儿正中胸骨切开术与栓子负荷显著增加相关,HITS强度低于猪的HITS。
局限性
- 结论指出高频脑多普勒超声对骨髓栓子的敏感度和特异度可能有限且依赖栓子大小。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Fat embolism syndrome and cerebral fat emboli are rare yet serious conditions arising from systemic distribution of bone marrow emboli. Emboli are known to produce high-intensity transient signals (HITS) in a Doppler signal. We hypothesized that both intramedullary nailing in pigs and median sternotomy in human infants cause bone marrow release, that some of these cause cerebral emboli, and that these were detectable by a new cerebral doppler ultrasound monitoring system (NeoDoppler). We also aimed to describe the intensity of HITS generated during these procedures.
METHODS: Specific pathogen-free Norwegian landrace pigs were allocated to either bilateral femoral nailing or injection of autologous bone marrow (positive controls). Testing was carried out under continuous Doppler monitoring. Presence of cerebral emboli was confirmed with histology. NeoDoppler data from infants undergoing sternotomy prior to cardiac surgery were investigated for comparison.
RESULTS: Eleven of twelve pigs were monitored with cerebral Doppler ultrasound during femoral surgery. HITS were seen in five (45%). Brain biopsies demonstrated bone marrow emboli in 11 of the 12 (92%). Four positive control pigs received intraarterial injections of bone marrow, saline, or contrast, and strong HITS were detected in all pigs (100%). Median sternotomy in eight human infants was associated with a significant increase in embolic burden; the HITS intensity was lower than HITS in pigs.
CONCLUSION: High-frequency cerebral Doppler ultrasound is a valuable tool for perioperative monitoring that can detect emboli in real-time, but sensitivity and specificity for bone marrow emboli may be limited and size-dependent.
引用
- Jarmund AH, Kristiansen S, Leth-Olsen M, et al. Perioperative Detection of Cerebral Fat Emboli From Bone Using High-Frequency Doppler Ultrasound. Ultrasound Med Biol. 2024. PMID: 39438224. doi:10.1016/j.ultrasmedbio.2024.09.017
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。