零回波时间(ZTE)序列评估颅底与颅盖骨性病变的效用
本研究纳入35例患者,将ZTE MRI处理生成反转ZTE(iZTE)和伪CT(pCT)图像,与CT比较评估颅底及颅盖骨性病变。pCT与iZTE观察者间一致性极好(kappa分别为0.91和0.92),诊断准确率分别为95.56%和94.44%,但在磨玻璃样变、骨质增生和病灶内骨碎片方面存在挑战。
要点
- pCT与iZTE评估颅底及颅盖骨性病变的观察者间一致性极好,kappa分别为0.91和0.92。
- pCT诊断准确率95.56%,iZTE为94.44%,敏感度分别为84.3%–95.1%和82.6%–94.2%。
- 两者在磨玻璃样变、骨质增生和病灶内骨碎片评估中遇到挑战。
核心结论
- 这项先导研究提示ZTE成像在颅底和颅盖骨性病变评估中具有融入标准诊疗的较强潜力,但需更大规模研究全面评估其效能。
研究设计与样本
- 先导研究,纳入35例 eligible 患者,平均年龄42岁,男女比1:4,均行ZTE MRI并处理生成iZTE和pCT图像,与CT扫描采用评分量表比较。
关键影像表现/征象
- pCT和iZTE均能准确预测多种病变特征,敏感度分别为84.3%–95.1%和82.6%–94.2%(95% CI)。
- 两者在磨玻璃样变、骨质增生和病灶内骨碎片评估中遇到挑战。
- 两者在其他骨性病变评估中表现良好。
诊断标准与关键数值
- pCT诊断准确率95.56%,iZTE诊断准确率94.44%。
- pCT观察者间kappa系数0.91(P < 0.0001),iZTE为0.92(P < 0.0001)。
临床应用与阅片要点
- 作者认为ZTE成像在颅底和颅盖骨性病变评估中具有融入标准诊疗的较强潜力。
- 作者指出需更大规模研究全面评估其效能。
局限性
- 原文指出需更大规模研究全面评估其效能。
- 原文指出pCT和iZTE在磨玻璃样变、骨质增生和病灶内骨碎片评估中遇到挑战。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: The emergence of zero echo time (ZTE) imaging has transformed bone imaging, overcoming historical limitations in capturing detailed bone structures. By minimizing the time gap between radiofrequency excitation and data acquisition, ZTE generates CT-like images. While ZTE has shown promise in various applications, its potential in assessing skull base and calvarium lesions remains unexplored. Hence we aim to introduce a novel perspective by investigating the utility of inverted ZTE images (iZTE) and pseudoCT (pCT) images for studying lesions in the skull base and calvarium.
MATERIALS AND METHODS: A total of 35 eligible patients, with an average age of 42 years and a male/female ratio of 1:4, underwent ZTE MRI and images are processed to generate iZTE and pCT images were generated through a series of steps including intensity equalization, thresholding, and deep learning-based pCT generation. These images were then compared to CT scans using a rating scale; inter-rater kappa coefficient evaluated observer consensus while statistical metrics like sensitivity and specificity assessed their performance in capturing bone-related characteristics.
RESULTS: The study revealed excellent interobserver agreement for lesion assessment using both pCT and iZTE imaging modalities, with kappa coefficient of 0.91 (P < 0.0001) and 0.92 respectively (P < 0.0001). Also, pCT and iZTE accurately predicted various lesion characteristics with sensitivity ranging from 84.3% to 95.1% and 82.6%-94.2% (95% CI) with a diagnostic accuracy of 95.56% and 94.44% respectively. Although both of them encountered challenges with ground glassing, hyperostosis, and intralesional bony fragments, they showed good performance in other bony lesion assessments.
CONCLUSIONS: The pilot study suggests strong potential for integrating the ZTE imaging into standard care for skull base and calvarial bony lesions assessment. Additionally, larger-scale studies are needed for comprehensive assessment of its efficacy.
引用
- Chauhan V, Harikishore K, Girdhar S, et al. Utility of zero echo time (ZTE) sequence for assessing bony lesions of skull base and calvarium. Clin Radiol. 2024. PMID: 39322533. doi:10.1016/j.crad.2024.08.029