在不同辐射剂量下使用光子计数CT定量骨微架构:与微型CT的比较
研究对同一人体的双侧远端桡骨、远端胫骨、股骨头及两节椎体进行离体扫描,比较不同辐射剂量(2.5–120 mGy)下光子计数探测器CT与微型CT定量骨小梁微架构参数的一致性。最低剂量下即呈高相关性(R²=0.55–0.95),随剂量增加相关性提高(20 mGy时R²=0.76–0.96),结论认为PCD-CT可在临床相关辐射剂量下定量骨小梁微架构,准确性堪比微型CT。
要点
- PCD-CT可在临床相关辐射剂量下定量骨小梁微架构,准确性堪比微型CT。
- 最低剂量下PCD-CT与微型CT骨小梁参数即呈高相关(R²=0.55–0.95),随剂量增加相关性提高。
- 随剂量增加,PCD-CT图像的对比噪声比和骨小梁锐度值均升高。
核心结论
- PCD-CT可用于定量骨小梁微架构,准确性堪比微型CT,且可在临床相关辐射剂量下实现。
研究设计与样本
- 离体研究,标本取自同一人体的双侧远端桡骨、远端胫骨、股骨头及两节椎体。
- PCD-CT扫描参数:层厚增量0.1 mm,像素尺寸0.1042–0.127 mm;辐射剂量基于容积CT剂量指数(CTDIvol32)在2.5至120 mGy间变化。
- 微型CT扫描参数:各向同性体素尺寸49–68.4 µm;以微型CT为金标准。
关键影像表现/征象
- 随辐射剂量增加,PCD-CT图像的对比噪声比和骨小梁锐度值均升高。
- PCD-CT与微型CT扫描配准后,将骨小梁分割为边长6 mm的立方体,逐立方体计算骨体积/总体积、骨小梁厚度、骨小梁数目和骨小梁异质性并进行比较。
诊断标准与关键数值
- PCD-CT与微型CT骨小梁微架构参数在最低剂量下即呈高相关(R²=0.55–0.95)。
- 相关性随辐射剂量增加而提高,20 mGy时R²=0.76–0.96。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: Accurate measurements of trabecular bone microarchitecture are required for a proper assessment of bone fragility. Photon-counting detector CT (PCD-CT) has different technical properties than conventional CT, resulting in higher resolution and thereby potentially enabling in-vivo measurement of trabecular microarchitecture. The purpose of this study was to quantify trabecular bone microarchitectural parameters with PCD-CT at varying radiation doses and compare this to µCT as gold standard.
METHOD: Both distal radii, distal tibiae, femoral heads, and two vertebrae were dissected from one human. All specimens were scanned ex-vivo on a PCD-CT system (slice increment 0.1 mm; pixel size 0.1042-0.127 mm) and a µCT system (isotropic voxel size 49-68.4 µm). The radiation doses of the PCD-CT scans were varied from 2.5 to 120 mGy based on the volume CT dose index (CTDIvol32). For the PCD-CT scans, contrast-to-noise ratio and trabecular sharpness were calculated and compared between radiation doses. µCT and PCD-CT scans were registered. The trabecular bone was then segmented from all PCD-CT and µCT scans and split into cubes with 6-mm edge length. For each cube, bone volume over total volume, trabecular thickness, trabecular number, and trabecular heterogeneity were calculated and compared between corresponding PCD-CT and µCT cubes.
RESULTS: With increasing dose, contrast-to-noise ratio and trabecular sharpness values increased for the PCD-CT images. Already at the lowest dose, high correlations between the trabecular microarchitectural parameters between µCT and PCD-CT were found (R2 = 0.55-0.95), which improved with increasing radiation dose (R2 = 0.76-0.96 at 20 mGy).
CONCLUSIONS: PCD-CT can be used to quantify trabecular bone microarchitecture, with accuracy comparable to µCT and at clinically relevant radiation doses.