基于深度学习的剪切波检测与分割工具用于慢性肝病床旁评估
本研究测试了Velacur设备中基于深度学习的剪切波检测与分割工具在MASLD患者中的性能。算法在103例患者的15045张专家分割图像上训练,在36名志愿者和患者的4429张图像上测试,图像水平敏感度81%、特异度84%,Dice系数0.74(图像水平)和0.75(患者水平)。将该工具叠加于B型超声可改善操作者采集的检查质量。
要点
- 剪切波检测算法图像水平敏感度81%、特异度84%,Dice系数0.74。
- 患者水平平均Dice系数为0.75。
- 软件工具叠加于B型超声可改善床旁检查质量。
核心结论
- 剪切波算法在志愿者和MASLD患者测试集上表现良好,在Velacur系统上实现该工具后改善了真实世界床旁肝脏评估的质量。
研究设计与样本
- 深度学习算法采用U-Net架构,在来自103例患者的15045张专家分割图像上训练,在来自36名志愿者和MASLD患者的4429张图像上测试,受试者在不同诊所由不同Velacur操作者扫描;参考标准为每张图像内剪切波的专家分割。
诊断标准与关键数值
- 图像水平正确波检测敏感度81%、特异度84%。
- 含波图像的Dice系数图像水平为0.74,患者水平平均为0.75。
临床应用与阅片要点
- 该软件工具作为B型超声上的叠加层实现,可改善操作者采集的检查质量。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: As metabolic dysfunction-associated steatotic liver disease (MASLD) becomes more prevalent worldwide, it is imperative to create more accurate technologies that make it easy to assess the liver in a point-of-care setting. The aim of this study is to test the performance of a new software tool implemented in Velacur (Sonic Incytes), a liver stiffness and ultrasound attenuation measurement device, on patients with MASLD. This tool employs a deep learning-based method to detect and segment shear waves in the liver tissue for subsequent analysis to improve tissue characterization for patient diagnosis.
METHODS: This new tool consists of a deep learning based algorithm, which was trained on 15,045 expert-segmented images from 103 patients, using a U-Net architecture. The algorithm was then tested on 4429 images from 36 volunteers and patients with MASLD. Test subjects were scanned at different clinics with different Velacur operators. Evaluation was performed on both individual images (image based) and averaged across all images collected from a patient (patient based). Ground truth was defined by expert segmentation of the shear waves within each image. For evaluation, sensitivity and specificity for correct wave detection in the image were calculated. For those images containing waves, the Dice coefficient was calculated. A prototype of the software tool was also implemented on Velacur and assessed by operators in real world settings.
RESULTS: The wave detection algorithm had a sensitivity of 81% and a specificity of 84%, with a Dice coefficient of 0.74 and 0.75 for image based and patient-based averages respectively. The implementation of this software tool as an overlay on the B-Mode ultrasound resulted in improved exam quality collected by operators.
CONCLUSION: The shear wave algorithm performed well on a test set of volunteers and patients with metabolic dysfunction-associated steatotic liver disease. The addition of this software tool, implemented on the Velacur system, improved the quality of the liver assessments performed in a real world, point of care setting.
引用
- Honarvar M, Lobo J, Schneider C, et al. Deep Learning Based Shear Wave Detection and Segmentation Tool for Use in Point-of-Care for Chronic Liver Disease Assessments. Ultrasound Med Biol. 2024. PMID: 39244483. doi:10.1016/j.ultrasmedbio.2024.08.002
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。