文献速递腹部技术与安全MR

5 T、3 T与1.5 T磁共振肝脏脂肪定量比较分析:一致性与可行性研究

Comparative analysis of hepatic fat quantification across 5 T, 3 T and 1.5 T: A study on consistency and feasibility.
PubMed 文献 · Eur J Radiol · 2024-08-30

研究用含脂0%–30%体模及35名参与者,在5 T、3 T、1.5 T下行化学位移编码MRI测量质子密度脂肪分数(PDFF),并以5 T和1.5 T磁共振波谱(MRS)PDFF为参考标准,分析5 T测量的一致性与可重复性。结果显示5 T CSE-PDFF与3 T、1.5 T一致性高(ICC分别为0.988、0.980,均p<0.001),5 T与MRS-PDFF一致性也高(ICC 0.988,p<0.001);部分肝叶内ROI间存在统计学显著差异(p=0.004和0.012)。作者结论认为5 T PDFF测量兼具准确性与可重复性,对肝脏脂肪含量定量有临床应用潜力。

要点

  • 5 T CSE-PDFF与3 T、1.5 T一致性高,ICC分别为0.988和0.980(均p<0.001)。
  • 5 T与MRS-PDFF一致性高,ICC为0.988(p<0.001)。
  • 肝叶内部分ROI间存在统计学显著差异(p=0.004和0.012)。

核心结论

  • 5 T MRI的PDFF测量表现出准确性和可重复性,提示5 T成像可提供可靠的肝脏脂肪含量定量,并显示出较大的临床诊断应用潜力。

研究设计与样本

  • 研究纳入含脂量0%–30%的体模及35名参与者(女21名、男14名;平均年龄30.17±13.98岁,体重指数25.84±4.76,腰臀比0.84±0.09),在5 T、3 T、1.5 T下行化学位移编码(CSE)MRI PDFF测量,并在5 T和1.5 T下对肝脏和体模行磁共振波谱(MRS),用jMRUI软件分析,以MRS所得PDFF值为参考标准。

诊断标准与关键数值

  • 5 T CSE-PDFF与3 T及1.5 T CSE-PDFF的ICC分别为0.988和0.980(均p<0.001)。
  • 肝叶内ROI间一致性ICC为0.975–0.986(均p<0.001)。
  • 体模与肝脏的5 T与1.5 T MRS-PDFF一致性ICC分别为0.996和0.980(均p<0.001),肝脏ROI的ICC为0.963–0.990(均p<0.001)。
  • 5 T CSE与5 T MRS-PDFF的ICC为0.988(p<0.001);5 T CSE与1.5 T MRS-PDFF的ICC为0.978(p<0.001)。
  • 同一5 T平台上肝叶内ROI一致性ICC为0.986–0.991(均p<0.001)。

临床应用与阅片要点

  • 原文结论认为5 T MRI的PDFF测量兼具准确性与可重复性,提示5 T成像可提供可靠的肝脏脂肪含量定量并显示出较大的临床诊断应用潜力。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Magnetic resonance imaging (MRI) is a critical noninvasive technique for evaluating liver steatosis, with efficient and precise fat quantification being essential for diagnosing liver diseases. This study leverages 5 T ultra-high-field MRI to demonstrate the clinical significance of liver fat quantification, and explores the consistency and accuracy of the Proton Density Fat Fraction (PDFF) in the liver across different magnetic field strengths and measurement methodologies.

METHODS: The study involved phantoms with lipid contents ranging from 0 % to 30 % and 35 participants (21 females, 14 males; average age 30.17 ± 13.98 years, body mass index 25.84 ± 4.76, waist-hip ratio 0.84 ± 0.09). PDFF measurements were conducted using chemical shift encoded (CSE) MRI at 5 T, 3 T, and 1.5 T, alongside magnetic resonance spectroscopy (MRS) at 5 T and 1.5 T for both liver and phantoms, analyzed using jMRUI software. The MRS-derived PDFF values served as the reference standard. Repeatability of 5 T MRI measurements was assessed through correlation analysis, while accuracy was evaluated using linear regression analysis against the reference standards.

RESULTS: The CSE-PDFF measurements at 5 T demonstrated strong consistency with those at 3 T and 1.5 T, showing high intraclass correlation coefficients (ICC) of 0.988 and 0.980, respectively (all p < 0.001). There was also significant consistency across ROIs within liver lobes, with ICC values ranging from 0.975 to 0.986 (all p < 0.001). MRS-PDFF measurements for both phantoms and liver at 5 T and 1.5 T exhibited substantial agreement, with ICC values of 0.996 and 0.980, respectively (all p < 0.001). Particularly, ICC values for ROIs in the liver ranged from 0.963 to 0.990 (all p < 0.001). Despite overall agreement, statistically significant differences were noted in specific ROIs within the liver lobes (p = 0.004 and 0.012). The CSE and MRS PDFF measurements at 5 T displayed strong consistency, with an ICC of 0.988 (p < 0.001), and significant agreement was also found between 5 T CSE and 1.5 T MRS PDFF measurements, with an ICC of 0.978 (p < 0.001). Agreement was significant within the ROIs of the liver lobes on the same platform at 5 T, with ICC values ranging from 0.986 to 0.991 (all p < 0.001).

CONCLUSION: PDFF measurements at 5 T MR imaging exhibited both accuracy and repeatability, indicating that 5 T imaging provides reliable quantification of liver fat content and shows substantial potential for clinical diagnostic applications.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Liu J, Wang Z, Yu D, et al. Comparative analysis of hepatic fat quantification across 5 T, 3 T and 1.5 T: A study on consistency and feasibility. Eur J Radiol. 2024. PMID: 39222564. doi:10.1016/j.ejrad.2024.111709

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。