文献速递外周血管介入US

用于动脉超声消融的超声引导高强度聚焦超声探头的实验评估

Experimental Evaluation of an Ultrasound-Guided High-Intensity-Focused Ultrasound Probe for Sonication of Artery.
PubMed 文献 · J Ultrasound Med · 2024-09-06

本研究开发了一种超声引导高强度聚焦超声(USgHIFU)探头用于动脉消融并评估血管收缩。在血管径向方向聚焦时,管壁显著变形(n=4),而轴向聚焦仅影响前壁收缩(n=4)。245 W、980 J、25 mL/min条件下收缩指数为56.2±9.7%(n=12);519 W、1038 J条件下为56.5±7.8%(n=17);总能量519 J条件下为18.3±5.1%(n=12)。声学能量通过管壁温度升高影响血管收缩;高声学功率时,即使声学能量相当,增加声学功率并未导致更大的血管收缩。

要点

  • 径向聚焦可致管壁显著变形(n=4),轴向聚焦仅影响前壁收缩(n=4)。
  • 245 W、980 J、25 mL/min条件下收缩指数56.2±9.7%(n=12)。
  • 519 W、1038 J条件下收缩指数56.5±7.8%(n=17);519 J条件下为18.3±5.1%(n=12)。

核心结论

  • 所开发的USgHIFU探头通过在径向方向覆盖更大范围的血管壁诱导更大的血管收缩;声学能量通过管壁温度升高影响血管收缩;高声学功率时,即使声学能量相当,增加声学功率并未导致更大的血管收缩。

研究设计与样本

  • 实验研究;使用USgHIFU探头(由两个共聚焦球形换能器用于消融和彩色多普勒血流成像探头用于引导组成);在体模中沿两个方向消融;离体兔主动脉嵌入离体猪肝,在不同声学功率(245和519 W)、流速(25、30、50 mL/min)和消融能量(519、980、1038 J)下消融;通过超声成像、显微镜观察和组织病理学分析评估消融后血管变化。

关键影像表现/征象

  • 沿血管径向方向聚焦导致管壁显著变形(n=4)。
  • 沿轴向方向聚焦仅影响前壁收缩(n=4)。

诊断标准与关键数值

  • 245 W、980 J、25 mL/min条件下收缩指数(Dc)为56.2±9.7%(n=12)。
  • 519 W、1038 J条件下Dc为56.5±7.8%(n=17)。
  • 总能量519 J条件下Dc为18.3±5.1%(n=12)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: This study aimed to develop an ultrasound-guided high-intensity-focused ultrasound (USgHIFU) probe for arterial sonication and to evaluate vascular contraction.

METHODS: The USgHIFU probe comprised two confocal spherical transducers for sonication and a US color Doppler flow imaging probe for guidance. A vessel-mimicking phantom was sonicated in two directions. In the vascular radial direction, an isolated rabbit aorta embedded in ex vivo pork liver was sonicated at different acoustic powers (245 and 519 W), flow rates (25, 30, and 50 mL/minute), and sonication energies (519, 980, and 1038 J). Changes in the postsonication vessels were evaluated using US imaging, microscopic observation, and histopathological analysis.

RESULTS: Beam focusing along the vascular radial direction caused significant deformation of both tube walls (n = 4), whereas focusing along the axial direction only affected the contraction of the anterior wall (n = 4). The contraction index (Dc) of the vessel sonicated at 245 W and 980 J was 56.2 ± 9.7% (n = 12) with 25 mL/minute. The Dc of the vessel sonicated at 519 W and 1038 J was 56.5 ± 7.8% (n = 17). The Dc of the vessel sonicated at 519 J total energy was 18.3 ± 5.1% (n = 12).

CONCLUSION: The developed USgHIFU probe induced greater vascular contractions by covering a larger area of the vessel wall in the radial direction. Sonication energy affects vascular contraction through temperature elevation of the vessel wall. When the acoustic power was high, an increase in acoustic power, even with comparable sonication energy, did not result in greater vessel contraction.