降低管电压和管电流并使用深度学习图像重建的超低剂量冠状动脉钙化积分CT
前瞻性研究纳入75例(A组)行常规剂量CSCT(120kVp/30mAs)后加做低剂量(120kVp/20mAs)扫描,另81例(B组)加做超低剂量(80kVp/20mAs)扫描;常规剂量用混合迭代重建,降低剂量用深度学习重建(DLR)。两组钙化积分和风险分级均无显著差异(ICC均为0.99),DLR图像冠状动脉与心周脂肪的CNR更高。
要点
- 超低剂量(80kVp/20mAs)联合DLR的CSCT钙化积分与常规剂量无显著差异(p=0.14,ICC=0.99)。
- 低剂量(120kVp/20mAs)联合DLR的钙化积分与常规剂量无显著差异(p=0.10,ICC=0.99)。
- DLR在两组均提高冠状动脉与心周脂肪的CNR(A组p<0.001,B组p=0.001)。
核心结论
- DLR不仅能在降低管电流的低剂量CSCT中可靠计算钙化积分,也能在同时降低管电压和管电流的超低剂量CSCT中可靠计算,显示其有潜力用于常规应用。
研究设计与样本
- 前瞻性研究,A组75例行常规剂量CSCT(120kVp/30mAs)后加做低剂量扫描(120kVp/20mAs),B组81例行常规剂量CSCT后加做超低剂量扫描(80kVp/20mAs);常规剂量数据用混合迭代重建,降低剂量数据用DLR重建。
诊断标准与关键数值
- A组钙化积分常规剂量与低剂量比较p=0.10、ICC=0.99,风险分级比较p=0.32、ICC=0.99。
- B组钙化积分常规剂量与超低剂量比较p=0.14、ICC=0.99,风险分级比较p=0.16、ICC=0.99。
- A组有效剂量常规剂量0.48±0.05 mSv vs 低剂量0.32±0.03 mSv。
- B组有效剂量常规剂量0.49±0.05 mSv vs 超低剂量0.09±0.01 mSv。
临床应用与阅片要点
- 作者认为DLR使低剂量和超低剂量CSCT的钙化积分和风险分级与常规剂量无显著差异,显示可用于常规应用的可行性。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To explore the potential of the deep learning reconstruction (DLR) for ultralow dose calcium scoring CT (CSCT) with simultaneously reduced tube voltage and current.
METHODS: In this prospective study, seventy-five patients (group A) undergoing routine dose CSCT (120kVp/30mAs) were followed by a low dose (120kVp/20mAs) scan and another 81 (group B) were followed by an ultralow dose (80kVp/20mAs) scan. The hybrid iterative reconstruction was used for the routine dose data while the DLR for data of reduced dose. The calcium score and risk categorization were compared, where the correlation was evaluated using the intraclass correlation coefficient (ICC). The noise suppression performance of DLR was characterized by the contrast-to-noise ratio (CNR) between coronary arteries and pericoronary fat.
RESULTS: The effective dose was 0.32 ± 0.03 vs. 0.48 ± 0.05 mSv for the two scans in group A and 0.09 ± 0.01 vs. 0.49 ± 0.05 mSv in group B. No significant difference was found on CACSs within either group (A: p = 0.10, ICC=0.99; B: p = 0.14, ICC=0.99), nor was it different on risk categorization (A: p = 0.32, ICC=0.99; B: p = 0.16, ICC=0.99). The DLR images exhibited higher CNR in both groups (A: p < 0.001; B: p = 0.001).
CONCLUSIONS: The DLR allowed reliable calcium scoring in not only low dose CSCT with reduced tube current but ultralow dose CSCT with simultaneously reduced tube voltage and current, showing feasibility to be adopted in routine applications.