淀粉样蛋白PET成像中PET/MR衰减校正在软件与硬件升级中的准确性与纵向一致性
研究纳入329名参与者,其中38人约3年内接受两次三模态PET/MR与CT扫描,采用迁移学习在2种扫描仪软件版本(VB20P和VE11P)和2种头部线圈(16通道和32通道)上微调DL-Dixon模型。DL-Dixon衰减校正在不同软件版本和线圈下均表现出高准确性和纵向一致性,超过95.6%的脑体素PET相对绝对误差小于10%,结论认为其适用于AD的纵向PET/MR神经影像研究。
要点
- DL-Dixon衰减校正在不同扫描仪软件版本和头部线圈下均保持高准确性,超过95.6%脑体素PET相对绝对误差小于10%。
- 皮质汇总区PET平均相对绝对误差中位数[四分位距]:VB20P 16通道线圈1.10%[0.93%, 1.26%],VE11P 16通道线圈1.24%[1.03%, 1.54%],VE11P 32通道线圈0.99%[0.86%, 1.13%]。
- DL-Dixon与CT衰减校正的皮质区PET SUVR个体内变异系数和组内相关系数相当,检测相同水平PET变化所需样本量相似。
核心结论
- DL-Dixon在2种软件版本和头部线圈下表现出优异的准确性和纵向一致性,证明其对AD纵向PET/MR神经影像研究的稳健性。
研究设计与样本
- 研究纳入329名参与者获取三模态PET/MR和CT图像,其中38人约3年内接受两次三模态扫描;采用迁移学习在2种扫描仪软件版本(VB20P和VE11P)和2种头部线圈(16通道和32通道)上微调DL-Dixon模型,评估其准确性和纵向一致性。
诊断标准与关键数值
- 在所有参与者中,超过95.6%的脑体素PET相对绝对误差小于10%;皮质汇总区PET平均相对绝对误差中位数[四分位距]:VB20P 16通道线圈1.10%[0.93%, 1.26%],VE11P 16通道线圈1.24%[1.03%, 1.54%],VE11P 32通道线圈0.99%[0.86%, 1.13%]。
临床应用与阅片要点
- DL-Dixon衰减校正可用于PET/MR淀粉样蛋白PET成像的纵向定量分析,在软件和硬件升级时保持PET SUVR的个体内变异系数和组内相关系数与CT衰减校正相当,检测相同水平PET变化所需样本量相似。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Integrated PET/MR allows the simultaneous acquisition of PET biomarkers and structural and functional MRI to study Alzheimer disease (AD). Attenuation correction (AC), crucial for PET quantification, can be performed by using a deep learning approach, DL-Dixon, based on standard Dixon images. Longitudinal amyloid PET imaging, which provides important information about disease progression or treatment responses in AD, is usually acquired over several years. Hardware and software upgrades often occur during a multiple-year study period, resulting in data variability. This study aims to harmonize PET/MR DL-Dixon AC amid software and head coil updates and evaluate its accuracy and longitudinal consistency.
MATERIALS AND METHODS: Tri-modality PET/MR and CT images were obtained from 329 participants, with a subset of 38 undergoing tri-modality scans twice within approximately 3 years. Transfer learning was used to fine-tune DL-Dixon models on images from 2 scanner software versions (VB20P and VE11P) and 2 head coils (16-channel and 32-channel coils). The accuracy and longitudinal consistency of the DL-Dixon AC were evaluated. Power analyses were performed to estimate the sample size needed to detect various levels of longitudinal changes in the PET standardized uptake value ratio (SUVR).
RESULTS: The DL-Dixon method demonstrated high accuracy across all data, irrespective of scanner software versions and head coils. More than 95.6% of brain voxels showed less than 10% PET relative absolute error in all participants. The median [interquartile range] PET mean relative absolute error was 1.10% [0.93%, 1.26%], 1.24% [1.03%, 1.54%], 0.99% [0.86%, 1.13%] in the cortical summary region, and 1.04% [0.83%, 1.36%], 1.08% [0.84%, 1.34%], 1.05% [0.72%, 1.32%] in cerebellum by using the DL-Dixon models for the VB20P 16-channel coil, VE11P 16-channel coil, and VE11P 32-channel coil data, respectively. The within-subject coefficient of variation and intraclass correlation coefficient of PET SUVR in the cortical regions were comparable between the DL-Dixon and CT AC. Power analysis indicated that similar numbers of participants would be needed to detect the same level of PET changes by using DL-Dixon and CT AC.
CONCLUSIONS: DL-Dixon exhibited excellent accuracy and longitudinal consistency across the 2 software versions and head coils, demonstrating its robustness for longitudinal PET/MR neuroimaging studies in AD.