文献速递神经CT

自动化算法在大中型血管闭塞检测中的真实世界表现

Performance of Automated Algorithm in Large and Medium Vessel Occlusion Detection: A Real-World Experience.
PubMed 文献 · AJNR Am J Neuroradiol · 2025-03-04

单中心回顾性队列纳入2020年3月至2023年2月连续3590例卒中代码患者,3576例可分析,其中616例(17.2%)存在血管闭塞。以12名放射科医师报告为参考标准,Viz.ai检测前循环大血管闭塞的敏感度91%、特异度93%,对后循环大血管闭塞及中等血管闭塞检测效能较低。

要点

  • Viz.ai检测前循环大血管闭塞敏感度91%(87-94%)、特异度93%(92-94%)。
  • 检测所有大血管闭塞敏感度降至73%(68-77%),特异度92%(91-93%)。
  • 检测前循环大血管闭塞合并中等血管闭塞敏感度65%(61-69%),特异度93%(92-94%)。

核心结论

  • 自动化算法识别前循环大血管闭塞准确性高,但对后循环大血管闭塞和中等血管闭塞检测效能较低。

研究设计与样本

  • 单中心回顾性登记研究,纳入2020年3月至2023年2月连续3590例卒中代码患者,3576例经CTA及Viz.ai自动化处理可分析,以12名委员会认证放射科医师的报告为参考标准。

关键影像表现/征象

  • 大血管闭塞定义为颅内ICA或M1-MCA闭塞(前循环)或基底动脉或V4-椎动脉闭塞(后循环)。
  • 中等血管闭塞定义为M2-MCA、A1/A2-大脑前动脉或P1/P2-大脑后动脉闭塞。

临床应用与阅片要点

  • 急性卒中团队应知晓自动化算法结果与实际大血管闭塞及中等血管闭塞发生率之间存在不一致,以利于及时诊断和分诊。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: Fast, accurate detection of large (LVO) and medium vessel occlusion (MeVO) is critical for triage and management of acute ischemic stroke. Multiple artificial intelligence (AI)-based software programs are available commercially for automated detection and rapid prioritization of LVO. However, their ability, strengths, and limitations for detection of acute vessel occlusion in the context of expanding indications for mechanical thrombectomy are not entirely understood. We aimed to investigate the performance of a fully automated commercial detection algorithm to detect large and medium vessel occlusions in code stroke patients.

MATERIALS AND METHODS: We utilized a single-center, institutional, retrospective registry of all consecutive code stroke patients with CTA and automated processing by using Viz.ai presenting at a large comprehensive stroke center between March 2020 and February 2023. LVO was categorized as anterior LVO (aLVO), defined as occlusion of the intracranial ICA or M1-MCA, and posterior LVO (pLVO), defined as occlusion of the basilar artery or V4-vertebral artery. MeVO was defined as occlusion of the M2-MCA, A1/A2-anterior cerebral artery, or P1/P2-posterior cerebral artery. We compared the accuracy of Viz.ai by using STARD guidelines. Radiology reports from 12 board-certified radiologists were considered the reference standard. Our primary outcome was assessing accuracy of the automated software for aLVO. Our secondary outcome was assessing accuracy for 3 additional categories: all LVO (aLVO and pLVO), aLVO with M2-MCA, and aLVO with MeVO.

RESULTS: Of 3590 code stroke patients, 3576 were technically sufficient for analysis by the automated software (median age 67 years; 51% women; 68% white), of which 616 (17.2%) had vessel occlusions. The respective sensitivity and specificity for our prespecified categories were: aLVO: 91% (87-94%), 93% (92-94%); all LVO: 73% (68-77%), 92% (91-93%); aLVO + M2-MCA occlusion: 74% (70-78%), 93% (92-94%); and aLVO + all MeVO: 65% (61-69%), 93% (92-94%).

CONCLUSIONS: The automated algorithm demonstrated high accuracy in identifying anterior LVO with lower performance for pLVO and MeVO. It is crucial for acute stroke teams to be aware of the discordance between automated algorithm results and true rates of LVO and MeVO for timely diagnosis and triage.

来源:AJNR Am J Neuroradiol,查看原文。版权归原作者所有。

引用

  1. Sriwastwa A, Aziz YN, Weiss K, et al. Performance of Automated Algorithm in Large and Medium Vessel Occlusion Detection: A Real-World Experience. AJNR Am J Neuroradiol. 2025. PMID: 39326885. doi:10.3174/ajnr.A8509

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。