多厂商双能CT有效原子序数量化准确性的体模比较研究

Multivendor comparison of quantification accuracy of effective atomic number by Dual-Energy CT: A phantom study.
PubMed 文献 · Eur J Radiol · 2024-08-23

本研究使用TomoTherapy奶酪体模,比较三个厂商五代双能CT在十二种管电压配置下有效原子序数(Zeff)的准确性。所有配置偏差均小于1,双层探测器双能CT -140 kV的MAPE最低(1.79%±1.93%);第三代双源和二代快速切换双能CT的MAPE高于其前代。Zeff准确性随厂商和双能CT代际而异,双层探测器双能CT准确性最高,升级代际未必提高准确性。

要点

  • 所有管电压配置的Zeff偏差均小于1。
  • 双层探测器双能CT -140 kV的MAPE最低(1.79%±1.93%)。
  • 第三代双源和二代快速切换双能CT的MAPE高于其前代。

核心结论

  • 扫描不同插入物时,Zeff准确性随厂商和双能CT代际而异;所有扫描仪中双层探测器双能CT的Zeff准确性最高;升级双能CT代际并不带来更高准确性,甚至更低。

研究设计与样本

  • 体模研究,使用TomoTherapy奶酪体模评估三个厂商五台双能CT在十二种管电压配置下的Zeff准确性,并测试三种辐射剂量(5、15、25 mGy)及不同插入物位置深度的影响。

诊断标准与关键数值

  • 所有管电压配置的偏差均小于1;双层探测器双能CT -140 kV的MAPE最低(1.79%±1.93%);线性混合模型显示管电压配置(F=16.92,p < 0.001)和插入物类型(F=53.26,p < 0.001)显著影响MAPE;辐射剂量仅对快速切换双能CT的MAPE有显著影响;插入物位置不影响最终MAPE。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: Our study aimed to compare the accuracy of the effective atomic number (Zeff) of five dual-energy CT (DECT) from three vendors and different generations under different scanning parameters.

METHODS: Zeff accuracy of five DECT scanners with twelve tube voltage configurations was evaluated by using the TomoTherapy cheese phantom. The potential dose dependence of the Zeff was investigated using three radiation dose (5, 15, and 25 mGy), and the robustness of Zeff was simulated for different organs of the body by placing the inserts at different positional depths. Bias and mean absolute percentage error (MAPE) were used to characterize the accuracy of Zeff. Data underwent analysis using one-way ANOVA, followed by the Turky and LSD post hoc tests, simple linear regression, and linear mixed models.

RESULTS: All tube voltage configurations had a bias of less than 1. Dual layer detector DECT (dl-DECT) -140 kV has the lowest MAPE (1.79 %±1.93 %). The third generation dual source DECT (ds-DECT) and the second generation rapid switch DECT (rs-DECT) have higher MAPE than their predecessor DECT. The results of the linear mixed model showed that tube voltage configuration (F=16.92, p < 0.001) and insert type (F=53.26, p < 0.001) significantly affect the MAPE. In contrast, radiation dose only has a significant effect on the MAPE of rs-DECT. The inserts position does not affect the final MAPE.

CONCLUSION: When scanning different inserts, Zeff accuracy varies by vendor and DECT generation. Of all the scanners, dl-DECT had the highest Zeff accuracy. Upgrading DECT generation doesn't lead to higher accuracy, or even lower.

来源:Eur J Radiol,查看原文。版权归原作者所有。