儿童光子计数CT的最佳光谱性能:基于体模的体部光谱成像kV选择研究
该体模研究用4种体型(新生儿、5岁、10岁、成人)模拟儿童腹部,在光子计数CT上比较70、90、120、140 kV四种kV选择对虚拟单能图像和碘图质量的影响。结果显示高kV(120或140 kV)在碘CT值准确度、碘定量误差和偏差方面总体优于低kV(70或90 kV),作者据此推荐在光子计数CT体部常规成像中采用高kV。
要点
- 高kV(120或140 kV)在虚拟单能图像碘CT值准确度和碘图定量误差方面总体优于低kV(70或90 kV)。
- 新生儿体型50 keV虚拟单能图像上,低kV噪声(70 kV:10.5 HU,90 kV:10.4 HU)低于高kV(120 kV:13.8 HU,140 kV:13.9 HU)。
- 成人体型70 kV的碘CT值绝对百分比误差中位数18.0%,显著高于120 kV的1.4%和140 kV的0.8%(P<0.0001)。
核心结论
- 光子计数CT上高kV选项(120或140 kV)在虚拟单能图像和碘图质量方面总体优于低kV选项(70或90 kV)。
- 作者推荐在光子计数CT体部常规成像中使用高kV。
研究设计与样本
- 体模研究,使用4种体型体模模拟新生儿、5岁、10岁和成人儿童患者腹部。
- 每个体模内置1个固体水插入物和4个已知浓度(2、5、10、15 mg I/mL)的固体碘插入物。
- 在光子计数CT系统(Siemens Alpha)上以4种kV选项(Quantum模式下70和90 kV,QuantumPlus模式下120和140 kV)及临床双源(3.0螺距)协议扫描。
- 所有kV采集的辐射剂量(CTDI vol)由临床剂量设置决定并匹配,另采集60%临床剂量图像。
- 重建统一使用Qr40内核和QIR level 3,生成40至80 keV、间隔10 keV的虚拟单能图像,并用基于图像的双材料分解法生成碘图。
关键影像表现/征象
- 新生儿体型50 keV虚拟单能图像上,70 kV噪声10.5 HU、90 kV噪声10.4 HU,低于120 kV的13.8 HU和140 kV的13.9 HU。
- 新生儿体型70 keV虚拟单能图像上,不同kV选项噪声相近(70 kV:9.4 HU,90 kV:8.9 HU,120 kV:9.7 HU,140 kV:10.2 HU)。
- 高kV(120或140 kV)在虚拟单能图像碘CT值绝对百分比误差方面总体优于低kV(70或90 kV)。
- 5岁体型90 kV的绝对百分比误差中位数3.6%,显著高于140 kV的1.6%(P<0.001,Kruskal-Wallis秩和检验加Bonferroni校正)。
- 成人体型70 kV的绝对百分比误差中位数18.0%,显著高于120 kV的1.4%和140 kV的0.8%(P<0.0001,Kruskal-Wallis秩和检验加Bonferroni校正)。
诊断标准与关键数值
- 虚拟单能图像碘CT值准确度以绝对百分比误差评估,5岁体型90 kV中位数3.6%显著高于140 kV中位数1.6%(P<0.001)。
- 成人体型70 kV绝对百分比误差中位数18.0%显著高于120 kV中位数1.4%和140 kV中位数0.8%(P<0.0001)。
- 碘图碘定量以均方根误差和偏差评估,10岁体型120 kV的RMSE为1.4、偏差为0.2 mg I/mL,70 kV的RMSE为1.9、偏差为0.8 mg I/mL。
- 高kV在所有受控条件下均显示比低kV更低的RMSE和偏差。
临床应用与阅片要点
- 作者推荐在光子计数CT体部常规成像中使用高kV(120或140 kV),因其虚拟单能图像和碘图质量总体更优。
- 新生儿体型在50 keV虚拟单能图像上低kV噪声较低,但该条件未改变作者对体部常规成像使用高kV的推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: The comprehensive evaluation of kV selection on photon-counting computed tomography (PCCT) has yet to be performed. The aim of the study is to evaluate and determine the optimal kV options for variable pediatric body sizes on the PCCT unit.
MATERIALS AND METHODS: In this study, 4 phantoms of variable sizes were utilized to represent abdomens of newborn, 5-year-old, 10-year-old, and adult-sized pediatric patients. One solid water and 4 solid iodine inserts with known concentrations (2, 5, 10, and 15 mg I/mL) were inserted into phantoms. Each phantom setting was scanned on a PCCT system (Siemens Alpha) with 4 kV options (70 and 90 kV under Quantum Mode, 120 and 140 kV under QuantumPlus Mode) and clinical dual-source (3.0 pitch) protocol. For each phantom setting, radiation dose (CTDI vol ) was determined by clinical dose settings and matched for all kV acquisitions. Sixty percent clinical dose images were also acquired. Reconstruction was matched across all acquisitions using Qr40 kernel and QIR level 3. Virtual monoenergetic images (VMIs) between 40 and 80 keV with 10 keV interval were generated on the scanner. Low-energy and high-energy images were reconstructed from each scan and subsequently used to generate an iodine map (IM) using an image-based 2-material decomposition method. Image noise of VMIs from each kV acquisition was calculated and compared between kV options. Absolute percent error (APE) of iodine CT number accuracy in VMIs was calculated and compared. Root mean square error (RMSE) and bias of iodine quantification from IMs were compared across kV options.
RESULTS: At the newborn size and 50 keV VMI, noise is lower at low kV acquisitions (70 kV: 10.5 HU, 90 kV: 10.4 HU), compared with high kV acquisitions (120 kV: 13.8 HU, 140 kV: 13.9 HU). At the newborn size and 70 keV VMI, the image noise from different kV options is comparable (9.4 HU for 70 kV, 8.9 HU for 90 kV, 9.7 HU for 120 kV, 10.2 HU for 140 kV). For APE of VMI, high kV (120 or 140 kV) performed overall better than low kV (70 or 90 kV). At the 5-year-old size, APE of 90 kV (median: 3.6%) is significantly higher ( P < 0.001, Kruskal-Wallis rank sum test with Bonferroni correction) than 140 kV (median: 1.6%). At adult size, APE of 70 kV (median: 18.0%) is significantly higher ( P < 0.0001, Kruskal-Wallis rank sum test with Bonferroni correction) than 120 kV (median: 1.4%) or 140 kV (median: 0.8%). The high kV also demonstrated lower RMSE and bias than the low kV across all controlled conditions. At 10-year-old size, RMSE and bias of 120 kV are 1.4 and 0.2 mg I/mL, whereas those from 70 kV are 1.9 and 0.8 mg I/mL.
CONCLUSIONS: The high kV options (120 or 140 kV) on the PCCT unit demonstrated overall better performance than the low kV options (70 or 90 kV), in terms of image quality of VMIs and IMs. Our results recommend the use of high kV for general body imaging on the PCCT.
引用
- Zhou W, Ataei A, Huo D, et al. Optimal Spectral Performance on Pediatric Photon-Counting CT: Investigating Phantom-Based Size-Dependent kV Selection for Spectral Body Imaging. Invest Radiol. 2024. PMID: 39159359. doi:10.1097/RLI.0000000000001119
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。