文献速递骨肌CT

利用光子计数CT在桡骨远端定量测定骨质量的指标

Quantitative metrics of bone quality determined at the distal radius using photon-counting CT.
PubMed 文献 · Skeletal Radiol · 2024-08-09

本研究在10具尸体腕部比较光子计数探测器CT(PCD-CT)与高分辨率外周定量CT(HRpQCT)的骨形态计量指标相关性及一致性,并在2022年2月至2023年8月间40例患者(31例成人、9例儿童)的活体桡骨远端评估临床效用;尸体中皮质厚度、小梁间距、小梁骨体积分数相关性好且一致性良好,小梁厚度一致性差;活体骨质疏松患者小梁间距更大、小梁体积骨密度更低。

要点

  • PCD-CT与HRpQCT在尸体皮质厚度、小梁间距、小梁骨体积分数相关性好且一致性良好。
  • 小梁厚度相关性中等且一致性差,临床解读需谨慎。
  • 活体骨质疏松患者小梁间距更大、小梁体积骨密度更低。

核心结论

  • PCD-CT在尸体中多数皮质与骨形态计量指标与HRpQCT一致,并能从常规临床PCD-CT图像提供活体桡骨远端骨微结构定量指标。

研究设计与样本

  • 研究扫描10具尸体腕部,使用PCD-CT与HRpQCT比较骨形态计量指标;另评估2022年2月至2023年8月间患者的腕部PCD-CT临床效用,共40例参与者(31例成人、9例儿童)。

诊断标准与关键数值

  • 尸体中PCD-CT与HRpQCT的Spearman相关系数:皮质厚度ρ=0.85,小梁间距ρ=0.98,小梁骨体积分数ρ=0.68,小梁厚度ρ=-0.49。
  • PCD-CT与HRpQCT的Lin一致性相关系数:皮质厚度0.80,小梁间距0.94,小梁骨体积分数0.86,小梁厚度-0.1。
  • 40例活体桡骨远端指标:皮质厚度0.99 mm(IQR 0.89-1.06),小梁厚度0.38 mm(IQR 0.25-0.40),小梁间距0.82 mm(IQR 0.72-1.05),小梁骨体积分数0.28(IQR 0.25-0.33)。
  • 骨质疏松患者小梁间距更大(p=0.025),小梁体积骨密度更低(p=0.042)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To determine the accuracy of photon-counting-detector CT (PCD-CT) at deriving bone morphometric indices and demonstrate utility in vivo in the distal radius.

METHODS: Ten cadaver wrists were scanned using PCD-CT and high-resolution peripheral quantitative CT (HRpQCT). Correlation between PCD-CT and HRpQCT morphometric indices was determined. Agreement was assessed by Lin's concordance correlation coefficient (Lin's CCC). Wrist PCD-CTs of patients between 02/2022 and 08/2023 were also evaluated for clinical utility. Morphometric indices of the in vivo distal radii were extracted and compared between patients with or without osteoporosis.

RESULTS: In cadavers, strong correlation between PCD-CT and HRpQCT was observed for cortical thickness (Spearman correlation, ρ, 0.85), trabecular spacing (ρ = 0.98), and trabecular bone volume fraction (ρ = 0.68). Moderate negative correlation (ρ = - 0.49) was observed for trabecular thickness. PCD-CT shows good agreement to HRpQCT for cortical thickness, trabecular spacing, and trabecular bone volume fraction (Lin's CCC = 0.80, 0.94, and 0.86, respectively) but poor agreement (Lin's CCC = - 0.1) for trabecular thickness. In forty participants (31 adults and 9 pediatric), bone morphometrics indices for cortical thickness, trabecular thickness, trabecular spacing, and trabecular bone volume fraction were 0.99 mm (IQR, 0.89-1.06), 0.38 mm (IQR, 0.25-0.40), 0.82 mm (IQR, 0.72-1.05), and 0.28 (IQR, 0.25-0.33), respectively. Patients with osteoporosis had statistically significantly larger trabecular spacing (p = 0.025) and lower trabecular volumetric bone mineral density (p = 0.042).

CONCLUSION: This study demonstrates the agreement of PCD-CT to HRpQCT in cadavers of most cortical and bone morphometrics examined and provide in vivo quantitative metrics of bone microarchitecture from routine clinical PCD-CT images of the distal radius.

来源:Skeletal Radiol,查看原文。版权归原作者所有。

引用

  1. Dutta A, Ferrero A, Rajendran K, et al. Quantitative metrics of bone quality determined at the distal radius using photon-counting CT. Skeletal Radiol. 2024. PMID: 39120685. doi:10.1007/s00256-024-04770-9

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。