自动化方法测量早发性脊柱侧凸儿童超声图像中的生长棒长度
本研究开发并验证了两种基于Mask RCNN的机器学习模型(Boundary模型和Rod模型),用于从超声图像中自动提取磁控生长棒长度。共采集23例早发性脊柱侧凸患儿的90张超声图像,其中70张用于模型开发(训练和验证),20张用于测试,比较AI与手动测量。模型平均精度分别为88.5%和60.2%,组间相关系数0.98,平均绝对差±标准差0.86±1.0 mm,Bland-Altman分析显示无偏倚,90%数据在95%置信区间内,输入图像后4.6秒显示测量结果。
要点
- AI模型可在输入超声图像后4.6秒内自动显示磁控生长棒长度测量结果。
- AI与手动测量的组间相关系数为0.98,平均绝对差为0.86±1.0 mm。
- 两个模型平均精度分别为88.5%(Boundary模型)和60.2%(Rod模型)。
核心结论
- 本研究首次提出基于AI的方法在超声图像上自动测量磁控生长棒长度,该方法可靠、准确且快速。
研究设计与样本
- 试点研究,采用Mask RCNN开发Boundary模型和Rod模型两个机器学习模型;从23例接受早发性脊柱侧凸手术的参与者采集90张超声图像,其中70张用于模型开发(训练和验证),20张用于测试;以手动测量为参考比较AI测量。
诊断标准与关键数值
- Boundary模型平均精度(AP)为88.5%,Rod模型平均精度为60.2%。
- AI与手动测量的组间相关系数(ICC)为0.98。
- AI与手动测量的平均绝对差±标准差(MAD±SD)为0.86±1.0 mm。
- Bland-Altman分析显示无偏倚,90%的数据在95%置信区间内。
- 输入超声图像后4.6秒显示测量结果。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To develop and validate machine learning algorithms to automatically extract the rod length of the magnetically controlled growing rod from ultrasound images (US) in a pilot study.
METHODS: Two machine-learning (ML) models, called the "Boundary model" and "Rod model," were developed to identify specific rod segments on ultrasound images. The models were developed utilizing Mask Regional Convolutional Neural Networks (Mask RCNN). Ninety US images were acquired from 23 participants who had early onset scoliosis (EOS) surgeries; among those, 70 were used for model development, including training and validation, and 20 were used for testing by comparing the AI-based vs. manual measurements.
RESULTS: The average precision (AP) of the ML models was 88.5% and 60.2%, respectively. The inter-method correlation coefficient (ICC) was 0.98, and the mean absolute difference ± standard deviation (MAD ± SD) between AI and manual measurements was 0.86 ± 1.0 mm. The Bland-Altman analysis showed no bias, and 90% of the data were within the 95% confidence interval. The automated method was reliable, accurate, and fast. Measurements were displayed in 4.6 seconds after the US image was inputted.
CONCLUSION: This was the first AI-based method to measure the MCGR rod length on US images automatically.